Abstract

Background: Stable fixation is essential to prevent relapse after Le Fort I osteotomy, particularly during the early postoperative period when the fixation system bears the full masticatory load. Patient-specific implants (PSI) have been increasingly used, but their biomechanical behavior under postoperative surgical conditions remains insufficiently investigated. The aim of this study was to evaluate PSI fixation using subject-specific finite element modelling based on real surgical plans. Methods: Two patient-specific models were developed from cone-beam computed tomography (CBCT) data of patients undergoing bimaxillary orthognathic surgery. For each case, two fixation strategies were simulated: a multiarm patient-specific plate and conventional L-shaped miniplates. Von Mises stress distribution and relative displacement were analyzed for fixation devices and bone segments. Results: PSI fixation reduced stress levels in plates and screws while maintaining comparable global stability. However, its effect on bone stress differed between the two models. In Patient 2, whose surgical plan included greater maxillary advancement and posterior impaction, PSI fixation was associated with higher bone stress and a micrometric increase in displacement. Conclusions: These findings highlight the influence of surgical movement on fixation biomechanics and support the integration of biomechanical personalization alongside anatomical planning in orthognathic surgery.

Related

  1. Piezoelectric Versus Conventional Osteotomy Instruments in Bimaxillary Orthognathic Surgery: A Narrative Review of Surgical Efficiency, Safety, and Clinical Outcomes
    Renato Isufi, Aurora Isufi, Ramazan Isufi +1
    Journal of Clinical MedicinePeer-reviewedCC BY 4.0Full textGeneral medicineCrossrefEurope PMC
  2. A biomechanical comparison of su5ture with a lateral plate and cerclage wire and suture constructs for lesser tuberosity osteotomy repair in stemless anatomic total shoulder arthroplasty
    Kyle B. Christy, Randall A. Arroyo, Mitchell S. Kirkham +5
    Journal of Shoulder and Elbow ArthroplastyPeer-reviewedCC BY 4.0Full textSurgery & anaesthesiaPMC
  3. Study Protocol for a Randomized Controlled Trial Comparing Swiss Ball Balance Training and Conventional Balance Training to Improve Trunk Control and Mobility in Individuals with Sub acute Stroke.
    B. Nkhoma, M. Mapulanga
    medRxivPreprintCC BYFull textGeneral medicinemedRxiv
  4. Real-world patient-reported outcome comparing antibody-drug conjugates and conventional chemotherapy in metastatic breast cancer.
    Manuela Milani, Francesco Riccetelli, Carla Strina +8
    The oncologistPeer-reviewedCC BYFull textOncologyEurope PMC
  5. Same patient, same total knee arthroplasty, different outcomes ? A pilot study comparing conventional vs. OMNIBotics Robotic-Assisted HLS Kneetec TM (Corin ® )
    Galdéric Roblot, Marie Moret, Quentin Seguret +6
    SICOT-JPeer-reviewedCC BYFull textGeneral biologySurgery & anaesthesiaHAL
BibTeX
@article{rios2026subjectspecific,
  title = {{Subject-Specific Finite Element Modelling of Le Fort I Osteotomy Fixation: Biomechanical Insights Comparing Patient-Specific Implants and Conventional Miniplates}},
  author = {Olina Rios and Sarah Latreche and Yannick Tillier and Charles Savoldelli},
  journal = {Journal of Clinical Medicine},
  year = {2026},
  month = oct,
  doi = {10.3390/jcm15207994},
  url = {https://www.mdpi.com/2077-0383/15/20/7994},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Subject-Specific Finite Element Modelling of Le Fort I Osteotomy Fixation: Biomechanical Insights Comparing Patient-Specific Implants and Conventional Miniplates
AU  - Olina Rios
AU  - Sarah Latreche
AU  - Yannick Tillier
AU  - Charles Savoldelli
PY  - 2026
DA  - 2026/10/01
JO  - Journal of Clinical Medicine
DO  - 10.3390/jcm15207994
UR  - https://www.mdpi.com/2077-0383/15/20/7994
AB  - Background: Stable fixation is essential to prevent relapse after Le Fort I osteotomy, particularly during the early postoperative period when the fixation system bears the full masticatory load. Patient-specific implants (PSI) have been increasingly used, but their biomechanical behavior under postoperative surgical conditions remains insufficiently investigated. The aim of this study was to evaluate PSI fixation using subject-specific finite element modelling based on real surgical plans. Methods: Two patient-specific models were developed from cone-beam computed tomography (CBCT) data of patients undergoing bimaxillary orthognathic surgery. For each case, two fixation strategies were simulated: a multiarm patient-specific plate and conventional L-shaped miniplates. Von Mises stress distribution and relative displacement were analyzed for fixation devices and bone segments. Results: PSI fixation reduced stress levels in plates and screws while maintaining comparable global stability. However, its effect on bone stress differed between the two models. In Patient 2, whose surgical plan included greater maxillary advancement and posterior impaction, PSI fixation was associated with higher bone stress and a micrometric increase in displacement. Conclusions: These findings highlight the influence of surgical movement on fixation biomechanics and support the integration of biomechanical personalization alongside anatomical planning in orthognathic surgery.