Visual discrimination down to the analytical limit of detection in colorimetric lateral flow biosensors via streamlined image processing.
- 1Analytical/Bioanalytical & Nanotechnology Research Group, Department of Chemistry, University of Patras Rio Patras 26504 Greece [email protected] [email protected].
Abstract
For point-of-care diagnostics, the visual limit of detection (LOD) is a critical parameter directly affecting diagnostic accuracy, as clinical decisions rely heavily on naked-eye classification. Improving visual detection while maintaining operational simplicity is highly advantageous for clinical testing. Here, colorimetric lateral flow biosensors for SARS-CoV-2 nucleocapsid protein (N-protein) and Streptococcus pyogenes antigen (Strep-A) were employed as models. The effect of N-protein concentration was studied in the range of 0-300 pM. The analytical LOD, determined both by longitudinal intensity profiling with subsequent test zone peak area integration, and by fixed region-of-interest analysis, was approximately 2 pM, whereas the visual LOD was 10-15 times higher (about 20-30 pM). Similar results were obtained for Strep-A. Within this concentration gap, human visual perception fails to resolve the colorimetric signal, leading to false-negative readouts despite the analyte level exceeding the analytical LOD. We demonstrate that a streamlined workflow based on histogram equalization of sensing membrane digital images reliably resolves subvisual, true analytical signals, effectively bridging the gap between the visual and analytical LOD. By redistributing pixel intensities via a cumulative distribution function, this process amplifies subtle reflectance intensity differences that remain imperceptible to the human eye. Robustness studies confirm that this resolution enhancement remains effective across both uncompressed TIFF and compressed JPEG formats, with compressed blank controls (0 pM) remaining free from non-specific signals. Because it operates exclusively on digital records without modifying the physical sensor architecture, this accessible, data-validated strategy can be combined with chemical signal-amplification methods to achieve synergistic performance improvement effects.
Related
- Streamlined Ultrafast Chromatography for High-Throughput Discovery and Development Chemistry via Predictive Analytical Modeling
- Developing an Aptamer-Based Colorimetric Method for Highly Specific Discrimination of Hg2+ and MeHg
- Chemometrics reshapes the role of biosensors: from signal transducers to intelligent analytical systems
- Multi-analytical discrimination of ceylon and cassia cinnamon using elemental and volatile fingerprinting
- Molecular Determinants of IgE-Mediated Food Allergenicity: Structural Dynamics, Processing-Induced Remodeling, and Analytical Interpretation
BibTeX
@article{christopoulou2026visual,
title = {{Visual discrimination down to the analytical limit of detection in colorimetric lateral flow biosensors via streamlined image processing.}},
author = {Natalia-Maria Christopoulou and Panagiotis K Tsikas and Theodore K Christopoulos},
journal = {RSC advances},
year = {2026},
month = nov,
doi = {10.1039/d6ra08921f},
url = {https://europepmc.org/article/MED/43011389},
note = {CC BY},
}
RIS
TY - JOUR
TI - Visual discrimination down to the analytical limit of detection in colorimetric lateral flow biosensors via streamlined image processing.
AU - Natalia-Maria Christopoulou
AU - Panagiotis K Tsikas
AU - Theodore K Christopoulos
PY - 2026
DA - 2026/11/10
JO - RSC advances
DO - 10.1039/d6ra08921f
UR - https://europepmc.org/article/MED/43011389
AB - For point-of-care diagnostics, the visual limit of detection (LOD) is a critical parameter directly affecting diagnostic accuracy, as clinical decisions rely heavily on naked-eye classification. Improving visual detection while maintaining operational simplicity is highly advantageous for clinical testing. Here, colorimetric lateral flow biosensors for SARS-CoV-2 nucleocapsid protein (N-protein) and Streptococcus pyogenes antigen (Strep-A) were employed as models. The effect of N-protein concentration was studied in the range of 0-300 pM. The analytical LOD, determined both by longitudinal intensity profiling with subsequent test zone peak area integration, and by fixed region-of-interest analysis, was approximately 2 pM, whereas the visual LOD was 10-15 times higher (about 20-30 pM). Similar results were obtained for Strep-A. Within this concentration gap, human visual perception fails to resolve the colorimetric signal, leading to false-negative readouts despite the analyte level exceeding the analytical LOD. We demonstrate that a streamlined workflow based on histogram equalization of sensing membrane digital images reliably resolves subvisual, true analytical signals, effectively bridging the gap between the visual and analytical LOD. By redistributing pixel intensities via a cumulative distribution function, this process amplifies subtle reflectance intensity differences that remain imperceptible to the human eye. Robustness studies confirm that this resolution enhancement remains effective across both uncompressed TIFF and compressed JPEG formats, with compressed blank controls (0 pM) remaining free from non-specific signals. Because it operates exclusively on digital records without modifying the physical sensor architecture, this accessible, data-validated strategy can be combined with chemical signal-amplification methods to achieve synergistic performance improvement effects.
SN - 2046-2069
ER -