Abstract
We present an in-situ azimuthal rotation stage for small-angle neutron scattering measurements. The stage provides an azimuthal rotation range of 315∘ about the stage normal. At the reference sample-stick orientation, the stage normal is parallel to the incident neutron beam, so that the sample rotates in the plane perpendicular to the beam. Combined with a sample stick rotation about a second, orthogonal axis, the stage provides control of the magnetic-field direction relative to the crystal axes, while the field remains fixed in the laboratory frame. For a 5 mm diameter incident beam, sample-stick rotations of approximately ±315∘ are accessible with negligible additional attenuation from the holder. The commercial rotator is specified for operation between 1 and 373 K and in magnetic fields up to 35 T, making the stage nominally compatible with the full temperature (1.8 to 320 K) and field (0 to 6.8 T) ranges of the used cryomagnet. As a working example, we characterize its performance on a spherical single crystal of MnSi between 20 and 35 K and in fields up to 183 mT, observing local sample heating during operation of ∼20 mK per 5∘ step. We illustrate the design considerations behind the holder construction, with specific notes for mounting spherical samples, and discuss the resulting performance and limitations.
Related
- Scattering-robust Imaging of Azimuthal Features with Enhanced Resolution
- Torsion and Nonmetricity Constraints From Neutron Spin Rotation in Polarized Matter
- Reduced vortex descriptors linking polycrystallinity in magnetic nanoparticles with polarized magnetic small-angle neutron scattering
- Fabrication and characterization of multi-shell nested fully annular quasi-ellipsoidal focusing mirrors with m=3 supermirrors for small-angle neutron scattering
- A search for non-Newtonian force in nm range from slow neutron scattering by noble gases and neutron diffraction on atom pairs
BibTeX
@misc{littlehales2026insitu,
title = {{In-situ azimuthal sample rotation within a horizontal-field cryomagnet for small-angle neutron scattering}},
author = {Matthew T. Littlehales and Andreas Bauer and Marek Bartkowiak and Markus Zolliker and Jonathan S. White and Christian Pfleiderer},
howpublished = {arXiv},
year = {2026},
month = nov,
doi = {10.48550/arxiv.2611.05159},
eprint = {2611.05159},
archivePrefix = {arXiv},
url = {https://arxiv.org/abs/2611.05159},
note = {CC BY 4.0},
}
RIS
TY - UNPB
TI - In-situ azimuthal sample rotation within a horizontal-field cryomagnet for small-angle neutron scattering
AU - Matthew T. Littlehales
AU - Andreas Bauer
AU - Marek Bartkowiak
AU - Markus Zolliker
AU - Jonathan S. White
AU - Christian Pfleiderer
PY - 2026
DA - 2026/11/03
PB - arXiv
DO - 10.48550/arxiv.2611.05159
UR - https://arxiv.org/abs/2611.05159
AB - We present an in-situ azimuthal rotation stage for small-angle neutron scattering measurements. The stage provides an azimuthal rotation range of 315^∘ about the stage normal. At the reference sample-stick orientation, the stage normal is parallel to the incident neutron beam, so that the sample rotates in the plane perpendicular to the beam. Combined with a sample stick rotation about a second, orthogonal axis, the stage provides control of the magnetic-field direction relative to the crystal axes, while the field remains fixed in the laboratory frame. For a 5 mm diameter incident beam, sample-stick rotations of approximately ±315^∘ are accessible with negligible additional attenuation from the holder. The commercial rotator is specified for operation between 1 and 373 K and in magnetic fields up to 35 T, making the stage nominally compatible with the full temperature (1.8 to 320 K) and field (0 to 6.8 T) ranges of the used cryomagnet. As a working example, we characterize its performance on a spherical single crystal of MnSi between 20 and 35 K and in fields up to 183 mT, observing local sample heating during operation of ∼20 mK per 5^∘ step. We illustrate the design considerations behind the holder construction, with specific notes for mounting spherical samples, and discuss the resulting performance and limitations.
ER -