Skip to content
DraconiticResearch
Search Sources API
201,818 studies31 sourcesUpdated 26 min ago
Peer-reviewedCC BY 4.0JACC: Case Reports

Catheter-Based Ventricular Assist Device–Induced Severe Hyperthermia

  1. 1Department of Medicine, University of California-Irvine Medical Center, Orange, California, USA
  2. 2Department of Medicine, University of California-Irvine Medical Center, Orange, California, USA; Mary & Steve Wen Cardiovascular Division, University of California-Irvine Health, Irvine, California, USA
  3. 3Mary & Steve Wen Cardiovascular Division, University of California-Irvine Health, Irvine, California, USA; Jeffrey M. Carlton Heart & Vascular Institute, Hoag Memorial Hospital Presbyterian, Newport Beach, California, USA

Abstract

Background: Hemolysis is a potential complication of catheter-based ventricular assist devices (CVADs), but severe hyperthermia is a rare occurrence that significantly affects further management and outcomes. Case Summary: A young man with toxic methamphetamine-associated cardiomyopathy presented with cardiogenic shock, prompting placement of a CVAD. He soon developed severe hyperthermia unresponsive to cooling measures. The hyperthermia ultimately subsided with removal of the device. Discussion: Hemolysis and renal failure are known complications of CVAD placement. The hyperthermia in this patient was likely related to erythrocytosis, highly viscous blood, and resulting hemolysis. The patient returned to normothermia only after removal of the device, supporting the idea that the CVAD was a cause of hyperthermic episode. Take-Home Messages: The use of CVADs in cardiogenic shock offers favorable outcomes but has potential complications. This unique case of severe hyperthermia resulting from massive hemolysis that required prompt removal of the device supports a rare, high-morbidity potential complication.

Related

  1. Novel placement technique of a left ventricular assist device for advanced heart failure with extensive left ventricular thrombus
    JTCVS techniquesPeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMC
  2. Transcatheter left ventricular assist device decommissioning
    Raziye E. Akdogan, Rohan Dod, Eric Batista +2
    Journal of cardiology casesPeer-reviewedCC BY-NC-NDFull textCardiovascular medicineEurope PMC
  3. Non-Cardiac Surgery After Left Ventricular Assist Device Implantation: Perioperative Management and Outcomes from a Single Center
    Journal of cardiovascular development and diseasePeer-reviewedCC BYFull textCardiovascular medicineEurope PMC
  4. ST-Segment Elevation Myocardial Infarction Secondary to Aortic Root Thrombosis in a Patient With a Left Ventricular Assist Device
    Umama Gorsi, Julia Kumar, Nathan Kim +2
    CureusPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. First-in-human totally robotic left ventricular assist device implantation through a right minithoracotomy
    Yi-Chun Kuo, Chu-Lin Hsiao, Ling-Yi Wei +4
    JTCVS techniquesPeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMC
BibTeX
@article{sarahnoorani2026catheterbased,
  title = {{Catheter-Based Ventricular Assist Device–Induced Severe Hyperthermia}},
  author = {Sarah Noorani, MD and Tyson Burnham, MD and Eric Zuniga, MD, MBA and Antonio H. Frangieh, MD, MPH and Andy Y. Lee, MD},
  journal = {JACC: Case Reports},
  year = {2026},
  month = sep,
  doi = {10.1016/j.jaccas.2026.110201},
  url = {https://doi.org/10.1016/j.jaccas.2026.110201},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Catheter-Based Ventricular Assist Device–Induced Severe Hyperthermia
AU  - Sarah Noorani, MD
AU  - Tyson Burnham, MD
AU  - Eric Zuniga, MD, MBA
AU  - Antonio H. Frangieh, MD, MPH
AU  - Andy Y. Lee, MD
PY  - 2026
DA  - 2026/09/02
JO  - JACC: Case Reports
DO  - 10.1016/j.jaccas.2026.110201
UR  - https://doi.org/10.1016/j.jaccas.2026.110201
AB  - Background: Hemolysis is a potential complication of catheter-based ventricular assist devices (CVADs), but severe hyperthermia is a rare occurrence that significantly affects further management and outcomes. Case Summary: A young man with toxic methamphetamine-associated cardiomyopathy presented with cardiogenic shock, prompting placement of a CVAD. He soon developed severe hyperthermia unresponsive to cooling measures. The hyperthermia ultimately subsided with removal of the device. Discussion: Hemolysis and renal failure are known complications of CVAD placement. The hyperthermia in this patient was likely related to erythrocytosis, highly viscous blood, and resulting hemolysis. The patient returned to normothermia only after removal of the device, supporting the idea that the CVAD was a cause of hyperthermic episode. Take-Home Messages: The use of CVADs in cardiogenic shock offers favorable outcomes but has potential complications. This unique case of severe hyperthermia resulting from massive hemolysis that required prompt removal of the device supports a rare, high-morbidity potential complication.
SN  - 2666-0849
ER  -