Skip to content
DraconiticResearch
Search Sources API
203,234 studies31 sourcesUpdated 39 min ago
PreprintCC BY 4.0arXiv

An optical cryostat for automated testing of silicon photomultipliers with vacuum-ultraviolet light

Abstract

We present the design and performance of an optical cryostat for testing silicon photomultipliers (SiPMs) with vacuum-ultraviolet (VUV) light. The cryostat operates within a temperature range of 120 to 340 K, achieving a temperature stability of 1.2 mK, measured in a rolling 10-minute window once settled within a liquid nitrogen fill-cycle, and 13.8 mK over a complete fill-cycle including the refill transient. The cryostat includes a VUV-transparent CaF2 viewport with a 142 mm diameter clear aperture, which limits the optically accessible sample area to 158 cm2. The viewport is integrated into a sealed dark box and optical table to facilitate light delivery. VUV illumination is achieved using a custom xenon scintillation lamp or a xenon flash lamp, with a motorized two-axis linear rail system ensuring precise light positioning over the sample. This paper details the cryostat and optical system design, as well as commissioning current-voltage (IV) measurements using Hamamatsu VUV-sensitive SiPMs from 165 to 295 K and a measurement of the photodetection efficiency at 165 K.

Related

  1. Regulation Mechanism of Well-barrier Thickness on Optical Gain and Loss for High-Power Ultraviolet Semiconductor Lasers
    Jing Yang, yuheng zhang, De Zhao +1
    Optics ExpressPeer-reviewedLicence not statedOptics & photonicsCrossref
  2. Ultraviolet to infrared thermally driven SSPM and spatially asymmetric optical propagation in vdW NbOCl₂ nanosheets
    Rui Liu, Jianlong Kang, Jinlai Zhao +3
    Optics ExpressPeer-reviewedLicence not statedOptics & photonicsCrossref
  3. Towards noble gas quantum optical magnetometry using direct ultraviolet detection
    James Maldaner, Gil Porat
    arXivPreprintarXiv licenceFull textAtomic, molecular & optical physicsApplied physics & instrumentationarXiv
  4. Cryogenic performance of Silicon Photo-Multipliers at LN2 temperature and at 3 K in the RES-NOVA cryostat
    Riccardo Rossini, Massimo Rossella, Massimiliano Clemenza +14
    arXivPreprintCC BY 4.0Full textApplied physics & instrumentationHigh-energy & particle physicsarXiv
  5. An extensible software platform for automated operational and characterization testing of scientific cameras
    Baolong Chen, Hongfei Zhang, Wei Liu +5
    arXivPreprintarXiv licenceFull textApplied physics & instrumentationarXiv
BibTeX
@misc{darroch2026optical,
  title = {{An optical cryostat for automated testing of silicon photomultipliers with vacuum-ultraviolet light}},
  author = {Lucas Darroch and Riya Rai and Soud Al Kharusi and David Gallacher and Megan Marquis and Sebastian Eppelt and Iskender Isikbay and Xiao Shang and Tsvetelin Totev and Naman Walia and Thomas Brunner},
  howpublished = {arXiv},
  year = {2026},
  month = sep,
  doi = {10.48550/arxiv.2609.05840},
  eprint = {2609.05840},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2609.05840},
  note = {CC BY 4.0},
}
RIS
TY  - UNPB
TI  - An optical cryostat for automated testing of silicon photomultipliers with vacuum-ultraviolet light
AU  - Lucas Darroch
AU  - Riya Rai
AU  - Soud Al Kharusi
AU  - David Gallacher
AU  - Megan Marquis
AU  - Sebastian Eppelt
AU  - Iskender Isikbay
AU  - Xiao Shang
AU  - Tsvetelin Totev
AU  - Naman Walia
AU  - Thomas Brunner
PY  - 2026
DA  - 2026/09/12
PB  - arXiv
DO  - 10.48550/arxiv.2609.05840
UR  - https://arxiv.org/abs/2609.05840
AB  - We present the design and performance of an optical cryostat for testing silicon photomultipliers (SiPMs) with vacuum-ultraviolet (VUV) light. The cryostat operates within a temperature range of 120 to 340 K, achieving a temperature stability of 1.2 mK, measured in a rolling 10-minute window once settled within a liquid nitrogen fill-cycle, and 13.8 mK over a complete fill-cycle including the refill transient. The cryostat includes a VUV-transparent CaF_2 viewport with a 142 mm diameter clear aperture, which limits the optically accessible sample area to 158 cm^2. The viewport is integrated into a sealed dark box and optical table to facilitate light delivery. VUV illumination is achieved using a custom xenon scintillation lamp or a xenon flash lamp, with a motorized two-axis linear rail system ensuring precise light positioning over the sample. This paper details the cryostat and optical system design, as well as commissioning current-voltage (IV) measurements using Hamamatsu VUV-sensitive SiPMs from 165 to 295 K and a measurement of the photodetection efficiency at 165 K.
ER  -