A multiway approach to overcome the peak resolution challenges in gas-chromatography coupled to ion mobility spectrometry

Abstract

Background: Gas-chromatography coupled to ion mobility spectrometry (GC-IMS) is a powerful platform for untargeted analysis of volatile compounds, combining high sensitivity and short analysis times. However, its practical implementation still presents several analytical and computational challenges. From the instrumental perspective, the high sensitivity of isothermal GC–IMS makes it prone to signal saturation and memory effects, which can compromise the reproducibility of measurements. On the data-analysis side, the nature of GC–IMS outputs produce large and collinear datasets, in which baseline drift, shift in ion mobility and retention time modes, and peak overlap hinder the reliable extraction of relevant chemical information. Results: This work proposes a comprehensive workflow from the analytical strategy to the data analysis using as benchmark a data set of honey samples of different origin. Analysing the GC-IMS data, obtained with an instrument operating in isothermal condition, a prominent and marked memory effect was observed; it could be contained only through the systematic use of frequent blanks. Notably, most papers in the literature do not adopt blanks at such frequency. Attention was then devoted to the preprocessing, to improve data quality prior to modelling while limiting the number of preprocessing steps. A multiway approach, i.e. Parallel Factor Analysis 2 (PARAFAC2) applied interval wise, was adopted, for the first time, to resolve overlapping peaks and extract chemically meaningful components from the GC-IMS landscape. The results obtained were also compared with Multivariate Curve Resolution–Alternating Least Squares (MCR–ALS) and recent improved PARAFAC2 algorithms. Significance: A complete pipeline for the GC-IMS analysis was carefully developed and validated. PARAFAC2, was successfully applied for the deconvolution of GC-IMS data. This approach provided comparable results with previous methods and indicated that the same grade of automation reached in the analysis of GC-MS data with the same approach is feasible.

Related

  1. Ultra-High-Performance Liquid Chromatography-High Resolution Mass Spectrometry and Gas Chromatography-Mass Spectrometry Screening of Classical Psychoactive Substances, Novel Psychoactive Substances, and Metabolites in Postmortem Urine Samples
    Eleazar Vargas-Mena, Milton Rosero-Moreano, Gonzalo Taborda-Ocampo +3
    Molecules (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral chemistryEurope PMC
  2. Analysis of rubber-derived contaminants in surface water and snow by liquid chromatography coupled with a hybrid linear ion-trap-Orbitrap high-resolution mass spectrometry
    Julie Anquetin, Mathieu Babin, Rachid Amara +1
    Journal of Chromatography APeer-reviewedCC BY-NC-ND 4.0Full text5 citedGeneral chemistryEnvironmental science & pollutionHAL
  3. Efficient dual-mode liquid chromatography–tandem mass spectrometry methods for quantifying major cyanotoxins in natural water
    Dani Khoury, Christelle Pallez, Sophie Lardy-Fontan +1
    Journal of Chromatography APeer-reviewedPublisher termsFull textAnalytical chemistryEnvironmental science & pollutionHAL
  4. The Development of a Simple Liquid Chromatography‐Precursor Ion Scan‐Mass Spectrometry Screening Method for the Detection of Phytocannabinoids in Plant Extracts
    Radwa Mahmoud, Deborah Michel, Fatma Elessawy +3
    Journal of mass spectrometry : JMSPeer-reviewedCC BY-NCFull textPhysical chemistryEurope PMC
  5. Preliminary pharmacokinetic investigation of two discoidin domain receptor 2 inhibitors in murine plasma and brain using liquid chromatography–tandem mass spectrometry analysis
    Paul Gueroue, Audrey Burban, Ahmad Sharanek +6
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Full textAnalytical chemistryHAL
BibTeX
@article{pellacani2026multiway,
  title = {{A multiway approach to overcome the peak resolution challenges in gas-chromatography coupled to ion mobility spectrometry}},
  author = {Samuele Pellacani and Lorenzo Strani and Daniele Tanzilli and Rebecca Barbieri and Raniero Marabottini and Francesca Martone and Angela Napolitano and Marina Cocchi and Caterina Durante},
  journal = {Analytica Chimica Acta},
  year = {2026},
  month = oct,
  doi = {10.1016/j.aca.2026.346326},
  url = {https://hal.inrae.fr/hal-05797544v1},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - A multiway approach to overcome the peak resolution challenges in gas-chromatography coupled to ion mobility spectrometry
AU  - Samuele Pellacani
AU  - Lorenzo Strani
AU  - Daniele Tanzilli
AU  - Rebecca Barbieri
AU  - Raniero Marabottini
AU  - Francesca Martone
AU  - Angela Napolitano
AU  - Marina Cocchi
AU  - Caterina Durante
PY  - 2026
DA  - 2026/10/13
JO  - Analytica Chimica Acta
DO  - 10.1016/j.aca.2026.346326
UR  - https://hal.inrae.fr/hal-05797544v1
AB  - Background: Gas-chromatography coupled to ion mobility spectrometry (GC-IMS) is a powerful platform for untargeted analysis of volatile compounds, combining high sensitivity and short analysis times. However, its practical implementation still presents several analytical and computational challenges. From the instrumental perspective, the high sensitivity of isothermal GC–IMS makes it prone to signal saturation and memory effects, which can compromise the reproducibility of measurements. On the data-analysis side, the nature of GC–IMS outputs produce large and collinear datasets, in which baseline drift, shift in ion mobility and retention time modes, and peak overlap hinder the reliable extraction of relevant chemical information. Results: This work proposes a comprehensive workflow from the analytical strategy to the data analysis using as benchmark a data set of honey samples of different origin. Analysing the GC-IMS data, obtained with an instrument operating in isothermal condition, a prominent and marked memory effect was observed; it could be contained only through the systematic use of frequent blanks. Notably, most papers in the literature do not adopt blanks at such frequency. Attention was then devoted to the preprocessing, to improve data quality prior to modelling while limiting the number of preprocessing steps. A multiway approach, i.e. Parallel Factor Analysis 2 (PARAFAC2) applied interval wise, was adopted, for the first time, to resolve overlapping peaks and extract chemically meaningful components from the GC-IMS landscape. The results obtained were also compared with Multivariate Curve Resolution–Alternating Least Squares (MCR–ALS) and recent improved PARAFAC2 algorithms. Significance: A complete pipeline for the GC-IMS analysis was carefully developed and validated. PARAFAC2, was successfully applied for the deconvolution of GC-IMS data. This approach provided comparable results with previous methods and indicated that the same grade of automation reached in the analysis of GC-MS data with the same approach is feasible.
SN  - 0003-2670
ER  -