Skip to content
DraconiticResearch
Search Sources API
203,234 studies31 sourcesUpdated 11 min ago
Peer-reviewedCC BYPhysics and imaging in radiation oncology

Three-dimensional deep learning nomogram for post-plan quality assurance in high-dose-rate prostate brachytherapy.

  1. 1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
  2. 2Medical Physics Unit, Department of Oncology, McGill University, Montreal, Quebec, Canada.
  3. 3Lady Davis Institute for Medical Research, Jewish General Hospital, Montreal, Quebec, Canada.

Abstract

Purpose

Classical nomograms for high-dose-rate (HDR) prostate brachytherapy quality assurance (QA) evaluate agreement between planned and expected total dwell time but do not directly assess the three-dimensional treated volume. This study's aim was to develop a deep learning-based 3D nomogram to predict institution-specific clinical 100% isodose volumes from anatomy and dwell positions for post-plan QA.Methods

A total of 2394 single-fraction 15 Gy HDR prostate brachytherapy plans were included. Input label maps encoded the 3D prostate, urethra, rectum, and dwell positions. The prediction target was the clinical 100% isodose volume. A five-fold cross-validation strategy was used for model development, with 2155 plans used for training and validation and 239 plans held out for testing. A 3D U-Net was trained using combined Dice and binary cross-entropy loss. Agreement was evaluated using prostate V100%, 100% isodose volume, Dice similarity coefficient, and average symmetric surface distance (ASSD). Validation-derived thresholds were used for outlier screening and targeted review.Results

The model predicted prostate V100% with a mean error of 0.73 percentage points. Predicted and clinical 100% isodose volumes showed strong agreement, with R² = 0.990, slope = 0.99, and intercept = 0.77 cm³. Median Dice similarity coefficient was 0.960, and median ASSD was 0.48 mm. A 95% threshold flagged 6.3% of plans as atypical relative to validation data.Conclusions

A deep learning-based nomogram predicted an institution-specific 100% isodose volume with low coverage error and high spatial agreement. This geometry aware post-plan QA tool can complement equation-based nomograms by directly evaluating the spatial treated volume.

Related

  1. Dosimetric equivalence of deep learning prostate contours after LDR brachytherapy: pre-declared margins, patient-level acceptance thresholds and the incremental predictive value of DVH indices
    Louis-Bernard St-Cyr, Anne Saint-Laurent, José Angel Lesteiro-Tejeda +7
    arXivPreprintCC BY-NC-ND 4.0Full textBiological & medical physicsarXiv
  2. Using artificial intelligence for clinical trial quality assurance in a multi-institutional neurovascular-sparing prostate radiotherapy trial
    Matthew Doucette, Ying Zhang, Chien-Yi Liao +10
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Applied physics & instrumentationCrossref
  3. Independent Three-dimensional Dose Calculation from Log Files for Breast VMAT: A Measurement-free Quality Assurance Approach
    Rosa Angelica Petit Sevilla, Oscar Abel Apaza Blanco, Nayron Medeiros Soares +6
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
  4. A deep learning approach for EPID-based patient-specific quality assurance incorporating daily machine-state variations
    Kihong Pak, Sung-woo Kim, Kyeongyun Park +5
    Frontiers in OncologyPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationCrossrefEurope PMCPMC
  5. Daily Multi‐Leaf Collimator positional error prediction using machine learning: A Dynalog‐based approach for patient specific quality assurance
    Samira Aghaei Akizi, Seyed Mahmoud Reza Aghamiri, Zahra Siavashpour +2
    Journal of Applied Clinical Medical PhysicsPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationPMC
BibTeX
@article{beaudry2026threedimensional,
  title = {{Three-dimensional deep learning nomogram for post-plan quality assurance in high-dose-rate prostate brachytherapy.}},
  author = {Joel Beaudry and Tonghe Wang and Shirin A Enger and Antonio L Damato},
  journal = {Physics and imaging in radiation oncology},
  year = {2026},
  month = sep,
  doi = {10.1016/j.phro.2026.101100},
  url = {https://europepmc.org/article/MED/42898876},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Three-dimensional deep learning nomogram for post-plan quality assurance in high-dose-rate prostate brachytherapy.
AU  - Joel Beaudry
AU  - Tonghe Wang
AU  - Shirin A Enger
AU  - Antonio L Damato
PY  - 2026
DA  - 2026/09/01
JO  - Physics and imaging in radiation oncology
DO  - 10.1016/j.phro.2026.101100
UR  - https://europepmc.org/article/MED/42898876
AB  - Purpose  Classical nomograms for high-dose-rate (HDR) prostate brachytherapy quality assurance (QA) evaluate agreement between planned and expected total dwell time but do not directly assess the three-dimensional treated volume. This study's aim was to develop a deep learning-based 3D nomogram to predict institution-specific clinical 100% isodose volumes from anatomy and dwell positions for post-plan QA.Methods  A total of 2394 single-fraction 15 Gy HDR prostate brachytherapy plans were included. Input label maps encoded the 3D prostate, urethra, rectum, and dwell positions. The prediction target was the clinical 100% isodose volume. A five-fold cross-validation strategy was used for model development, with 2155 plans used for training and validation and 239 plans held out for testing. A 3D U-Net was trained using combined Dice and binary cross-entropy loss. Agreement was evaluated using prostate V100%, 100% isodose volume, Dice similarity coefficient, and average symmetric surface distance (ASSD). Validation-derived thresholds were used for outlier screening and targeted review.Results  The model predicted prostate V100% with a mean error of 0.73 percentage points. Predicted and clinical 100% isodose volumes showed strong agreement, with R² = 0.990, slope = 0.99, and intercept = 0.77 cm³. Median Dice similarity coefficient was 0.960, and median ASSD was 0.48 mm. A 95% threshold flagged 6.3% of plans as atypical relative to validation data.Conclusions  A deep learning-based nomogram predicted an institution-specific 100% isodose volume with low coverage error and high spatial agreement. This geometry aware post-plan QA tool can complement equation-based nomograms by directly evaluating the spatial treated volume.
SN  - 2405-6316
ER  -