Agreement, Reliability, and Feasibility of Melbourne Rapid Fields Tablet Perimetry Compared with Humphrey Field Analyzer in Glaucoma

  1. 1Department of Ophthalmology, Faculty of Medicine, Universiti Kebangsaan Malaysia, Jalan Yaakob Latiff, Bandar Tun Razak, Cheras, Kuala Lumpur 56000, Malaysia; (M.H.B.); (T.C.C.)
  2. 2Department of Ophthalmology, Hospital Canselor Tuanku Muhriz, Jalan Yaakob Latiff, Bandar Tun Razak, Cheras, Kuala Lumpur 56000, Malaysia

Abstract

Objectives: We aimed to evaluate the agreement, test–retest reliability, test duration, and patient acceptability of Melbourne Rapid Fields (MRF) tablet perimetry compared with Humphrey Field Analyzer (HFA) 24-2 SITA-Standard perimetry in glaucoma patients. Methods: This cross-sectional study evaluated 64 eyes of 64 patients with primary or secondary glaucoma. Participants underwent both HFA and MRF perimetry in a non-randomized sequence. Agreement was analyzed using intraclass correlation coefficients (ICC) and Bland–Altman analysis. Test–retest reliability of MRF was evaluated in 47 participants after 4 weeks. Acceptability was assessed via questionnaires. Results: MRF demonstrated comparable reliability indices to HFA (p > 0.05), although these findings do not establish equivalent clinical reliability due to differences in platform-specific definitions. MRF demonstrated excellent test–retest reliability for MD (ICC 0.993), PD (ICC 0.992), and VC (ICC 0.992; all p < 0.001). Compared with HFA, MRF produced significantly less negative MD (−6.27 ± 5.55 vs. −8.32 ± 5.47 dB, p < 0.001) and higher PSD/PD (7.48 ± 3.26 vs. 6.33 ± 4.11 dB, p < 0.001), while VFI/VC did not differ significantly (p = 0.808). Overall agreement was good-to-excellent MD (ICC 0.949), PSD/PD (ICC 0.880), and VFI/VC (ICC 0.930). However, exploratory severity-stratified Bland–Altman analysis suggested greater variability in moderate and severe glaucoma. MRF was significantly faster than HFA (4.22 ± 0.81 vs. 5.46 ± 1.33 min, p < 0.001). All patients reported it was easy and comfortable. Conclusions: MRF provides excellent test–retest reliability, shorter testing duration, and high patient acceptability. While overall agreement is high, systematic differences and greater advanced-stage variability prevent direct interchangeability with HFA. Instead, MRF serves as a practical, complementary platform for accessible, frequent testing.

Related

  1. Remote Assessment of Mobility in People with Parkinson's Disease: A Feasibility, Safety, Validity, Agreement, and Reliability Study
    Carla Silva‐Batista, Kathleen T. Scanlan, Margaret E. Stojak +7
    Movement Disorders Clinical PracticePeer-reviewedLicence not statedFull textNeurologyPMC
  2. Feasibility and performance of eye movement perimetry for glaucoma triage in Tanzania: A prospective cross-sectional study in a real-world setting
    Arun J. Thirunavukarasu, Victor H. Hu, Pete R. Jones +9
    PLOS Global Public HealthPeer-reviewedCC BY 4.0Full textNeurosciencePhysiologyPLOS
  3. AI‐Based Wound Measurement in a Telemedicine‐Style Workflow: Agreement With In‐Person Reference Measurements Compared With Individual Physicians and Expert Consensus
    Tassilo Dege, Vanessa Borst, Luisa Deutzmann +3
    International wound journalPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  4. Large language models for systematic review tasks compared with Cochrane reviews: a protocol for evaluating accuracy, agreement, and time efficiency.
    Josue Rivadeneira, Luis Fuenmayor-González, Daniel Peñaherrera-Vásquez +10
    Systematic reviewsPeer-reviewedCC BY-NC-NDFull textClinical trials & evidence synthesisEurope PMC
  5. The reliability conundrum of DWI-ASPECTS in stroke assessment: a systematic review of agreement on DWI-ASPECTS in anterior circulation stroke
    Mobin Jamal, Vincent Brissette, Gabriele Zitikyte +2
    Frontiers in NeurologyPeer-reviewedCC BY 4.0Full textEpidemiology & public healthCrossrefDOAJEurope PMCPMC
BibTeX
@article{basri2026agreement,
  title = {{Agreement, Reliability, and Feasibility of Melbourne Rapid Fields Tablet Perimetry Compared with Humphrey Field Analyzer in Glaucoma}},
  author = {Murshidah Hassan Basri and Teck Chee Cheng and Rona Asnida Nasaruddin and Jemaima Che Hamzah},
  journal = {Diagnostics},
  year = {2026},
  month = oct,
  doi = {10.3390/diagnostics16213528},
  url = {https://doi.org/10.3390/diagnostics16213528},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Agreement, Reliability, and Feasibility of Melbourne Rapid Fields Tablet Perimetry Compared with Humphrey Field Analyzer in Glaucoma
AU  - Murshidah Hassan Basri
AU  - Teck Chee Cheng
AU  - Rona Asnida Nasaruddin
AU  - Jemaima Che Hamzah
PY  - 2026
DA  - 2026/10/27
JO  - Diagnostics
DO  - 10.3390/diagnostics16213528
UR  - https://doi.org/10.3390/diagnostics16213528
AB  - Objectives: We aimed to evaluate the agreement, test–retest reliability, test duration, and patient acceptability of Melbourne Rapid Fields (MRF) tablet perimetry compared with Humphrey Field Analyzer (HFA) 24-2 SITA-Standard perimetry in glaucoma patients. Methods: This cross-sectional study evaluated 64 eyes of 64 patients with primary or secondary glaucoma. Participants underwent both HFA and MRF perimetry in a non-randomized sequence. Agreement was analyzed using intraclass correlation coefficients (ICC) and Bland–Altman analysis. Test–retest reliability of MRF was evaluated in 47 participants after 4 weeks. Acceptability was assessed via questionnaires. Results: MRF demonstrated comparable reliability indices to HFA (p > 0.05), although these findings do not establish equivalent clinical reliability due to differences in platform-specific definitions. MRF demonstrated excellent test–retest reliability for MD (ICC 0.993), PD (ICC 0.992), and VC (ICC 0.992; all p < 0.001). Compared with HFA, MRF produced significantly less negative MD (−6.27 ± 5.55 vs. −8.32 ± 5.47 dB, p < 0.001) and higher PSD/PD (7.48 ± 3.26 vs. 6.33 ± 4.11 dB, p < 0.001), while VFI/VC did not differ significantly (p = 0.808). Overall agreement was good-to-excellent MD (ICC 0.949), PSD/PD (ICC 0.880), and VFI/VC (ICC 0.930). However, exploratory severity-stratified Bland–Altman analysis suggested greater variability in moderate and severe glaucoma. MRF was significantly faster than HFA (4.22 ± 0.81 vs. 5.46 ± 1.33 min, p < 0.001). All patients reported it was easy and comfortable. Conclusions: MRF provides excellent test–retest reliability, shorter testing duration, and high patient acceptability. While overall agreement is high, systematic differences and greater advanced-stage variability prevent direct interchangeability with HFA. Instead, MRF serves as a practical, complementary platform for accessible, frequent testing.
SN  - 2075-4418
ER  -