Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter

6 affiliations
  1. 1IR Innovate Research Pvt. Ltd., Noida, Uttar Pradesh, India
  2. 2City X-Ray and Scan Clinic Pvt. Ltd., New Delhi, India
  3. 3Citizen Hospital, Kalasipalya, Bengaluru, India
  4. 4Dr. Dipti’s Multispeciality Clinics, Sahibabad, Uttar Pradesh, India
  5. 5Panchsheel Hospital, New Delhi, India
  6. 6Department of Surgery, IMU University, Malaysia

Abstract

Background: Non-invasive hemoglobin measurement offers a painless and rapid alternative to conventional blood-based testing. The Vivaray Hb pro is a handheld photoplethysmography-based device designed for point-of-care hemoglobin assessment without blood sampling. We evaluated the clinical performance of the Vivaray Hb pro by comparing device-generated hemoglobin values with those obtained from a calibrated laboratory blood cell counter. Methods: In this cross-sectional, non-randomized clinical performance study, participants aged ≥8 years were prospectively recruited. Hemoglobin was measured non-invasively using the Vivaray Hb pro and compared with venous blood samples analyzed on a calibrated Coulter counter. Agreement between methods was assessed using Bland–Altman analysis, including regression-based evaluation for proportional bias. Mean difference, mean absolute error (MAE) and proportions of measurements within tolerance limits were also calculated. Complete paired measurements were available for 763 individuals. Results: Bland–Altman analysis returned a mean difference of −0.40 g/dL. The bias was hemoglobin-dependent, being higher than the true value at low hemoglobin levels and underestimation at higher levels. Regression-based analysis showed proportional bias (β1 = −0.178), indicating decreasing difference with increasing hemoglobin concentration. The MAE was 1.5 g/dL but was lower (1.2) in the clinically predominant ranges of 8.1–13 g/dL. Conclusion: The Vivaray Hb pro can function as a useful non-invsive screening and point-of-care assessment device for measuring hemoglobin, particularly in settings where rapid, painless and repeatable measurements are desirable.

Related

  1. Blood loss due to diagnostic testing in extremely preterm infants in 22 European countries: a prospective observational study
    Nina Houben, Suzanne Fustolo-Gunnink, Karin Fijnvandraat +106
    EClinicalMedicinePeer-reviewedCC BYGeneral biologyHAL
  2. Origins of allostery in vertebrate hemoglobin evolution
    C. R. Cortez-Romero, N. M. Bautista, C. Nisler +3
    bioRxivPreprintCC BY-NC-NDFull textEvolutionary biologybioRxiv
  3. Engineering a hemoglobin variant with enhanced stability and oxygen affinity using a protein language model.
    Sergey Shityakov, Alexander S Novikov, Ekaterina V Skorb +1
    Frontiers in chemistryPeer-reviewedCC BYFull textCell biologyEurope PMC
  4. Evidence-Calibrated Reanalysis of an Otosclerosis Locus Generates a Testable Myeloid State Hypothesis
    Weijun Ma, Ying Cheng, Heng Li +1
    GenesPeer-reviewedCC BYFull textGenetics & genomicsEurope PMC
  5. Hemoglobin-based oxygen carriers: evolution, translational applications and clinical considerations.
    Huan Tang, Xinyi Liu, Hai Zhang +3
    Frontiers in medicinePeer-reviewedCC BYFull textCell biologyEurope PMC
BibTeX
@article{kumar2026vivaray,
  title = {{Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter}},
  author = {Devesh Kumar and Sunita Kapoor and Ambanna Gowda and Dipti Gupta and Harsh Mittal and Suneet Sood},
  journal = {PLOS One},
  year = {2026},
  month = nov,
  doi = {10.1371/journal.pone.0353640},
  url = {https://doi.org/10.1371/journal.pone.0353640},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Vivaray Hb pro non-invasive hemoglobin device: A prospective diagnostic accuracy study measuring agreement with a calibrated blood cell counter
AU  - Devesh Kumar
AU  - Sunita Kapoor
AU  - Ambanna Gowda
AU  - Dipti Gupta
AU  - Harsh Mittal
AU  - Suneet Sood
PY  - 2026
DA  - 2026/11/03
JO  - PLOS One
DO  - 10.1371/journal.pone.0353640
UR  - https://doi.org/10.1371/journal.pone.0353640
AB  - Background: Non-invasive hemoglobin measurement offers a painless and rapid alternative to conventional blood-based testing. The Vivaray Hb pro is a handheld photoplethysmography-based device designed for point-of-care hemoglobin assessment without blood sampling. We evaluated the clinical performance of the Vivaray Hb pro by comparing device-generated hemoglobin values with those obtained from a calibrated laboratory blood cell counter. Methods: In this cross-sectional, non-randomized clinical performance study, participants aged ≥8 years were prospectively recruited. Hemoglobin was measured non-invasively using the Vivaray Hb pro and compared with venous blood samples analyzed on a calibrated Coulter counter. Agreement between methods was assessed using Bland–Altman analysis, including regression-based evaluation for proportional bias. Mean difference, mean absolute error (MAE) and proportions of measurements within tolerance limits were also calculated. Complete paired measurements were available for 763 individuals. Results: Bland–Altman analysis returned a mean difference of −0.40 g/dL. The bias was hemoglobin-dependent, being higher than the true value at low hemoglobin levels and underestimation at higher levels. Regression-based analysis showed proportional bias (β1 = −0.178), indicating decreasing difference with increasing hemoglobin concentration. The MAE was 1.5 g/dL but was lower (1.2) in the clinically predominant ranges of 8.1–13 g/dL. Conclusion: The Vivaray Hb pro can function as a useful non-invsive screening and point-of-care assessment device for measuring hemoglobin, particularly in settings where rapid, painless and repeatable measurements are desirable.
SN  - 1932-6203
ER  -