Twenty-Year Wear Progression and Long-Term Outcomes of Conventional Polyethylene Liners After Cementless Total Hip Arthroplasty

  1. 1Department of Orthopaedic Surgery, Faculty of Medicine, The University of Tokyo, Tokyo, Japan
  2. 2Sensory & Motor System Medicine, Faculty of Medicine, The University of Tokyo, Tokyo, Japan; Corresponding author. 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8655, Japan. Tel.: +813 3815 5411.
  3. 3Department of Orthopaedic Surgery, NTT Medical Center Tokyo, Tokyo, Japan

Abstract

Background: Polyethylene (PE) wear debris can cause periprosthetic osteolysis and late failure after total hip arthroplasty, but long-term wear behavior of conventional non–cross-linked PE remains unclear. Methods: We reviewed 85 consecutive cementless total hip arthroplasties performed between 2002 and 2005 using ArCom non–cross-linked PE liners (Biomet Inc., Warsaw, IN). After excluding 31 hips lost to follow-up and 5 revised hips, 49 hips (45 patients) with ≥20 years of follow-up were analyzed. Femoral head penetration was measured using PolyWare (Draftware Inc., Vevay, IN) by three independent observers. The first postoperative year was defined as the bedding-in phase, and subsequent penetration as the steady-state phase (SSP). Associations between wear and patient, implant, surgical, and clinical factors were assessed. Results: The mean Japanese Orthopaedic Association score improved from 39.6 ± 10.6 preoperatively to 84.4 ± 12.9 at 20 years ( P < .001). Osteolysis was observed in 18 hips. Mean bedding-in phase and SSP wear rates were 0.235 ± 0.155 mm/y and 0.031 ± 0.023 mm/y, respectively. SSP wear correlated linearly with postoperative time (R 2 = 0.99). Hips with higher SSP wear (>0.025 mm/y) had higher 20-year Japanese Orthopaedic Association gait ( P = .033), activities of daily living ( P = .018), and total scores ( P = .013). Charnley category A had higher SSP wear than category B ( P = .024), whereas osteolysis did not affect wear rates. Conclusions: ArCom conventional PE showed a nearly constant SSP wear rate over 20 years. Exploratory analyses suggested associations with functional status and Charnley category, but not with other patient, implant, or surgical factors.

Related

  1. Early Revision Rates and Clinical Outcomes of Two Cementless Cruciate-Retaining Total Knee Arthroplasty Designs at Two-Year Follow-Up: A Propensity Score-Adjusted Comparative Analysis
    Medicina (Kaunas, Lithuania)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. Age May Not be a Barrier: Cementless Total Knee Arthroplasty Yields Equivalent or Superior PROMs in Older Adults at One Year
    Benjamin E. Jevnikar, Khaled A. Elmenawi, Yuxuan Jin +2
    Indian journal of orthopaedicsPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
  3. Long-term outcomes of bioabsorbable polylactide scaffold-based versus conventional arthroplasty for trapeziometacarpal osteoarthritis.
    Antek Nicklas, Feras Taqatqeh, Adrian Dragu +1
    GMS Interdisciplinary plastic and reconstructive surgery DGPWPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
  4. Same patient, same total knee arthroplasty, different outcomes ? A pilot study comparing conventional vs. OMNIBotics Robotic-Assisted HLS Kneetec TM (Corin ® )
    Galdéric Roblot, Marie Moret, Quentin Seguret +6
    SICOT-JPeer-reviewedCC BYFull textGeneral biologySurgery & anaesthesiaHAL
  5. Final Femoral Rasp Predicts Intraoperative Stem Alignment in Primary Total Hip Arthroplasty Using a Cementless Triple-Tapered Stem.
    Yannick Alexander Ahlheit, Gautier Beckers, Dominic Simon +4
    Arthroplasty todayPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMC
BibTeX
@article{yasutakawatanabe2026twentyyear,
  title = {{Twenty-Year Wear Progression and Long-Term Outcomes of Conventional Polyethylene Liners After Cementless Total Hip Arthroplasty}},
  author = {Yasutaka Watanabe, MD and Toru Moro, MD, PhD and Hisatoshi Ishikura, MD, PhD and Hirofumi Oshima, MD and Takeyuki Tanaka, MD, PhD and Sakae Tanaka, MD, PhD},
  journal = {Arthroplasty Today},
  year = {2026},
  month = oct,
  doi = {10.1016/j.artd.2026.102187},
  url = {https://doi.org/10.1016/j.artd.2026.102187},
  note = {CC BY-NC-ND},
}
RIS
TY  - JOUR
TI  - Twenty-Year Wear Progression and Long-Term Outcomes of Conventional Polyethylene Liners After Cementless Total Hip Arthroplasty
AU  - Yasutaka Watanabe, MD
AU  - Toru Moro, MD, PhD
AU  - Hisatoshi Ishikura, MD, PhD
AU  - Hirofumi Oshima, MD
AU  - Takeyuki Tanaka, MD, PhD
AU  - Sakae Tanaka, MD, PhD
PY  - 2026
DA  - 2026/10/22
JO  - Arthroplasty Today
DO  - 10.1016/j.artd.2026.102187
UR  - https://doi.org/10.1016/j.artd.2026.102187
AB  - Background: Polyethylene (PE) wear debris can cause periprosthetic osteolysis and late failure after total hip arthroplasty, but long-term wear behavior of conventional non–cross-linked PE remains unclear. Methods: We reviewed 85 consecutive cementless total hip arthroplasties performed between 2002 and 2005 using ArCom non–cross-linked PE liners (Biomet Inc., Warsaw, IN). After excluding 31 hips lost to follow-up and 5 revised hips, 49 hips (45 patients) with ≥20 years of follow-up were analyzed. Femoral head penetration was measured using PolyWare (Draftware Inc., Vevay, IN) by three independent observers. The first postoperative year was defined as the bedding-in phase, and subsequent penetration as the steady-state phase (SSP). Associations between wear and patient, implant, surgical, and clinical factors were assessed. Results: The mean Japanese Orthopaedic Association score improved from 39.6 ± 10.6 preoperatively to 84.4 ± 12.9 at 20 years ( P < .001). Osteolysis was observed in 18 hips. Mean bedding-in phase and SSP wear rates were 0.235 ± 0.155 mm/y and 0.031 ± 0.023 mm/y, respectively. SSP wear correlated linearly with postoperative time (R 2 = 0.99). Hips with higher SSP wear (>0.025 mm/y) had higher 20-year Japanese Orthopaedic Association gait ( P = .033), activities of daily living ( P = .018), and total scores ( P = .013). Charnley category A had higher SSP wear than category B ( P = .024), whereas osteolysis did not affect wear rates. Conclusions: ArCom conventional PE showed a nearly constant SSP wear rate over 20 years. Exploratory analyses suggested associations with functional status and Charnley category, but not with other patient, implant, or surgical factors.
SN  - 2352-3441
ER  -