Perceived sound quality and speech recognition are not interchangeable outcome measures in adult cochlear implant users.

  1. 1Department of Otolaryngology, Hannover Medical School, Hannover, Germany.
  2. 2Cluster of Excellence "Hearing4All", Hannover, Germany.
  3. 3College of Health Solutions, Arizona State University, Tempe, AZ, United States.

Abstract

Introduction

This study examined whether descriptor-based reports of perceived sound quality in adult cochlear implant users are associated with speech-recognition outcomes, preoperative maximum monosyllabic word recognition, and patient- or device-related factors.Methods

In this retrospective single-center analysis, 249 adult cochlear implant users completed a 25-item descriptor-based sound-quality assessment covering retrospectively recalled early (T1) and current (T2) listening experience. Descriptor profiles and superordinate descriptor groups were linked to preoperative maximum monosyllabic word recognition and postoperative speech-recognition outcomes in quiet and noise using group comparisons, univariate logistic regression, and multiple linear regression of rank-transformed objective outcomes, with false-discovery-rate correction.Results

Individual sound-quality descriptors were not robustly associated with concurrent or later speech-recognition performance. Preoperative maximum monosyllabic word recognition was not robustly associated with later descriptor profiles. Patient- and device-related factors showed selective associations with descriptor endorsement. Among nominally significant associations in adjusted models, descriptors and supergroups explained 1.5-3.2 percentage points of additional variance in speech-recognition outcomes and 1.8-3.3 percentage points in preoperative maximum monosyllabic word recognition beyond patient- and device-related factors.Conclusion

Speech-recognition measures do not reliably capture perceived cochlear implant sound quality. Structured sound-quality assessment may add clinically relevant information beyond standard speech testing, particularly when subjective listening experience and conventional outcome measures do not align.

Related

  1. Associations of Word Recognition in Noise with Perceived Hearing Handicap and Quality of Life in Older Adults with Sensorineural Hearing Loss.
    Ebru Sönmez, Çağlar Batman
    Audiology researchPeer-reviewedCC BYFull textNeuroscienceEurope PMC
  2. Spectro-temporal modulation sensitivity prospectively predicts speech-in-noise recognition in cochlear implant users
    Julia Erb, Malte Wöstmann, Jens Kreitewolf +3
    Ear and hearingPeer-reviewedLicence not statedFull textNeurosciencePMC
  3. Perceived stress mediates the longitudinal effect of sleep quality on internalizing symptoms
    Runtang Meng, Jiale Xu, Yi Luo +10
    Journal of Affective DisordersPeer-reviewedLicence not stated13 citedGeneral biologyHAL
  4. Comparing prescribing and spending across originator, follow-on, and interchangeable biosimilar insulin glargine: Opportunities to improve uptake of biosimilar medications
    Lucille M. Abrams, Kylee Funk, Joel F. Farley
    Journal of Managed Care & Specialty PharmacyPeer-reviewedLicence not statedFull textImmunologyPMC
  5. Long-term impact of the diagnosis on quality of life, social skills, and person recognition strategies in congenital prosopagnosics: A cross-sectional study
    Martina Grüter, Claus-Christian Carbon, Thomas Grüter
    i-PerceptionPeer-reviewedCC BY-NC 4.0Full textNeurosciencePMCEurope PMC
BibTeX
@article{buczak2026perceived,
  title = {{Perceived sound quality and speech recognition are not interchangeable outcome measures in adult cochlear implant users.}},
  author = {Nadine Buczak and Michael F Dorman and Anke Leichtle and Andreas Büchner},
  journal = {Frontiers in human neuroscience},
  year = {2026},
  month = nov,
  doi = {10.3389/fnhum.2026.1968712},
  url = {https://europepmc.org/article/MED/43080347},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Perceived sound quality and speech recognition are not interchangeable outcome measures in adult cochlear implant users.
AU  - Nadine Buczak
AU  - Michael F Dorman
AU  - Anke Leichtle
AU  - Andreas Büchner
PY  - 2026
DA  - 2026/11/10
JO  - Frontiers in human neuroscience
DO  - 10.3389/fnhum.2026.1968712
UR  - https://europepmc.org/article/MED/43080347
AB  - Introduction  This study examined whether descriptor-based reports of perceived sound quality in adult cochlear implant users are associated with speech-recognition outcomes, preoperative maximum monosyllabic word recognition, and patient- or device-related factors.Methods  In this retrospective single-center analysis, 249 adult cochlear implant users completed a 25-item descriptor-based sound-quality assessment covering retrospectively recalled early (T1) and current (T2) listening experience. Descriptor profiles and superordinate descriptor groups were linked to preoperative maximum monosyllabic word recognition and postoperative speech-recognition outcomes in quiet and noise using group comparisons, univariate logistic regression, and multiple linear regression of rank-transformed objective outcomes, with false-discovery-rate correction.Results  Individual sound-quality descriptors were not robustly associated with concurrent or later speech-recognition performance. Preoperative maximum monosyllabic word recognition was not robustly associated with later descriptor profiles. Patient- and device-related factors showed selective associations with descriptor endorsement. Among nominally significant associations in adjusted models, descriptors and supergroups explained 1.5-3.2 percentage points of additional variance in speech-recognition outcomes and 1.8-3.3 percentage points in preoperative maximum monosyllabic word recognition beyond patient- and device-related factors.Conclusion  Speech-recognition measures do not reliably capture perceived cochlear implant sound quality. Structured sound-quality assessment may add clinically relevant information beyond standard speech testing, particularly when subjective listening experience and conventional outcome measures do not align.
SN  - 1662-5161
ER  -