Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial

5 affiliations
  1. 1Rainier Clinical Research Center, Renton, WA USA
  2. 2Sinocare Meditech, Inc., Lauderdale, FL USA
  3. 3Sinocare, 265 GuYuan Road, Hi-Tech Zone, Changsha, Hunan 410205 People’s Republic of China
  4. 4Flacke Consulting GmbH, Haydnstr. 12, 55130 Mainz, Germany
  5. 5Integrated Medical Development, 186 Princeton Hightstown Rd, Princeton Junction, NJ 08550 USA

Abstract

Introduction: The performance of a new continuous glucose monitoring system (CGM) with flavin adenine dinucleotide-dependent glucose dehydrogenase (GDH-FAD) technology in adults with type 1 diabetes mellitus (T1DM) was assessed, with a focus on the impact of ascorbic acid and exercise on the accuracy of glucose measurements relative to Yellow Springs Instrument (YSI) 2300 control.

Methods: Adults with T1DM aged 18–65 years using a stable glucose management treatment regimen for at least 3 months with multiple daily injections or continuous subcutaneous insulin infusion participated in this study at a single center. Participants attended a screening visit and three in-clinic visits for frequent blood sampling with blood glucose measurements using YSI measurements as the reference measurement as endorsed by the Food and Drug Administration (FDA). Participants had two meals at each clinic visit leading to pronounced glucose excursions. On the second and third in-clinic visits, participants performed moderate exercise in between the two meals and were randomized to receive 1000 mg of ascorbic acid orally at either of those visits. Accuracy evaluation included the proportion of CGM values within 15% of YSI glucose values > 100 mg/dL or within 15 mg/dL of YSI values ≤ 100 mg/dL (%15/15), along with %20/20, %30/30, and %40/40 agreement rates (AR) across CGM ranges. The mean absolute difference (MAD) and mean absolute relative difference (MARD) with and without ascorbic acid and exercise, as well as their differences, are presented across glucose ranges.

Results: Data from 16 participants were analyzed. In the overall CGM range (35–450 mg/dL), the %20/20 AR was 92.0%. The overall changes in MAD between the exercise/nonexercise and the ascorbic acid/nonascorbic acid conditions were −1.3 mg/dL and −1.1 mg/dL. The overall changes in MARD between those two sets of conditions were −1.5% and −1.6%, respectively, none of which were considered to be clinically relevant. There were no serious adverse events.

Conclusions: This study provides evidence that neither exercise nor ascorbic acid results in any clinically significant interference on glucose measurements using the iCan CGM system.

Trial Registration: ClinicalTrials.gov identifier, NCT05348928.

Supplementary Information: The online version contains supplementary material available at https://doi.org/10.1007/s13300-026-01918-9.

Related

  1. Monitoring multiple rib fractures using pulmonary electrical impedance tomography technology: correlation analysis of pulmonary ventilation changes
    Liming Wang, Lingqing Zhou, Picong You +4
    Journal of thoracic diseasePeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMC
  2. 341 SMART G-Rex® & SensoRexTM technology for in-line, non-invasive, and real-time monitoring of TIL expansion
    Chantale Bernatchez, Marie-Andree Forget, Donastas Sakellariou-Thompson +7
    Journal for ImmunoTherapy of CancerPeer-reviewedCC BYOncologyDOAJ
  3. Accuracy and fairness of facial recognition technology in real-world, low-quality police images: An experiment
    Maria Cuellar, Hon Kiu James To, Arush Mehrotra
    PLOS OnePeer-reviewedCC BYFull textNeurosciencePhysiologyPLOSEurope PMC
  4. Practical Guidance for the Use of Control‐IQ Technology During Physical Activity and Exercise
    Sergio Di Molfetta, Jonida Haxhi, Francesca Franco +5
    Diabetes, Obesity & MetabolismPeer-reviewedCC BY-NC-ND 4.0Full textEndocrinology & metabolismPMC
  5. Real-world accuracy, extended-wear performance, and user satisfaction with a novel 21-day continuous glucose monitoring system in pediatric type 1 diabetes.
    Erdal Eren, Yasemin Denkboy Ongen, Burcu Eryilmaz +2
    BMJ open diabetes research & carePeer-reviewedCC BY-NCFull textEndocrinology & metabolismEurope PMC
BibTeX
@article{brazg2026impact,
  title = {{Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial}},
  author = {Ronald Brazg and Tina Mitchell and Jiangfeng Fei and Jiyun Zheng and Fei Gao and Ao Gao and Siting Zhu and Frank Flacke and Leon Shi and Poul Strange},
  journal = {Diabetes Therapy},
  year = {2026},
  month = oct,
  doi = {10.1007/s13300-026-01918-9},
  url = {https://doi.org/10.1007/s13300-026-01918-9},
  note = {CC BY-NC 4.0},
}
RIS
TY  - JOUR
TI  - Impact of Ascorbic Acid and Exercise on Accuracy of a Glucose Monitoring System with GDH-FAD Technology in Adult Patients with Type 1 Diabetes Mellitus: A Double-Blinded, Randomized Trial
AU  - Ronald Brazg
AU  - Tina Mitchell
AU  - Jiangfeng Fei
AU  - Jiyun Zheng
AU  - Fei Gao
AU  - Ao Gao
AU  - Siting Zhu
AU  - Frank Flacke
AU  - Leon Shi
AU  - Poul Strange
PY  - 2026
DA  - 2026/10/09
JO  - Diabetes Therapy
DO  - 10.1007/s13300-026-01918-9
UR  - https://doi.org/10.1007/s13300-026-01918-9
AB  - Introduction: The performance of a new continuous glucose monitoring system (CGM) with flavin adenine dinucleotide-dependent glucose dehydrogenase (GDH-FAD) technology in adults with type 1 diabetes mellitus (T1DM) was assessed, with a focus on the impact of ascorbic acid and exercise on the accuracy of glucose measurements relative to Yellow Springs Instrument (YSI) 2300 control.  Methods: Adults with T1DM aged 18–65 years using a stable glucose management treatment regimen for at least 3 months with multiple daily injections or continuous subcutaneous insulin infusion participated in this study at a single center. Participants attended a screening visit and three in-clinic visits for frequent blood sampling with blood glucose measurements using YSI measurements as the reference measurement as endorsed by the Food and Drug Administration (FDA). Participants had two meals at each clinic visit leading to pronounced glucose excursions. On the second and third in-clinic visits, participants performed moderate exercise in between the two meals and were randomized to receive 1000 mg of ascorbic acid orally at either of those visits. Accuracy evaluation included the proportion of CGM values within 15% of YSI glucose values > 100 mg/dL or within 15 mg/dL of YSI values ≤ 100 mg/dL (%15/15), along with %20/20, %30/30, and %40/40 agreement rates (AR) across CGM ranges. The mean absolute difference (MAD) and mean absolute relative difference (MARD) with and without ascorbic acid and exercise, as well as their differences, are presented across glucose ranges.  Results: Data from 16 participants were analyzed. In the overall CGM range (35–450 mg/dL), the %20/20 AR was 92.0%. The overall changes in MAD between the exercise/nonexercise and the ascorbic acid/nonascorbic acid conditions were −1.3 mg/dL and −1.1 mg/dL. The overall changes in MARD between those two sets of conditions were −1.5% and −1.6%, respectively, none of which were considered to be clinically relevant. There were no serious adverse events.  Conclusions: This study provides evidence that neither exercise nor ascorbic acid results in any clinically significant interference on glucose measurements using the iCan CGM system.  Trial Registration: ClinicalTrials.gov identifier, NCT05348928.  Supplementary Information: The online version contains supplementary material available at https://doi.org/10.1007/s13300-026-01918-9.
SN  - 1869-6953
SN  - 1869-6961
ER  -