Cryogenic performance and long-term stability of Hamamatsu R8520-506 photomultiplier tubes in liquid argon

Abstract

Seven Hamamatsu R8520-506 photomultiplier tubes were characterized at room temperature and in liquid argon to evaluate their response and operating stability for liquid-argon light readout. A 405~nm pulsed light source was used to measure gain, peak-to-valley ratio, single-photoelectron charge resolution, and afterpulsing, while dark-count data were acquired with random triggers in the absence of active illumination. Gain and dark-count stability were monitored during a ten-day operating period in liquid argon. At a motherboard input voltage of 900~V, the extracted gains in liquid argon were approximately 27.8\%--51.3\% higher than the corresponding room-temperature values. For all seven tubes, the peak-to-valley ratios were higher and the single-photoelectron charge resolution improved in liquid argon. The dark count rates ranged from approximately 143 to 252~Hz at room temperature and from 166 to 199~Hz in liquid argon, with three tubes showing increases and four showing decreases. The mean number of detected afterpulses per selected primary event was approximately 21.0\%--63.6\% lower in liquid argon. The relative standard deviations across the ten daily means of gain and dark count rate did not exceed approximately 2.1\% and 3.3\%, respectively. The measurements provide a quantitative basis for considering R8520-506 tubes as photosensor candidates for liquid-argon detectors.

Related

  1. Evaluation of the upgraded 3-inch Hamamatsu photomultiplier for the KM3NeT Neutrino Telescope
    S Aiello, A Albert, A.R Alhebsi +272
    JINSTPeer-reviewedCC BY2 citedApplied physics & instrumentationHigh-energy & particle physicsHAL
  2. Liquid-Nitrogen Micropillar-Wick Cooling for Cryogenic Electronics: A Numerical Study of Thermal Performance and Capillary Dry-Out Limits
    Felix Mende, Marcus Wislicenus, Dianping Jiang +2
    arXivPreprintarXiv licenceFull textApplied physics & instrumentationComputational physicsarXiv
  3. Quantifying Phase Transformations in Alloying Anodes via in-situ Liquid Cell Hard X-ray Spectroscopy and Cryogenic Microscopy
    Neil Mulcahy, Syeda Ramin Jannat, Yaqi Li +7
    PRX EnergyPeer-reviewedCC BY 4.0Full textCondensed matterHAL
  4. Physics-integrated neural network modeling of heat and mass transfer in cryogenic liquid storage under static and dynamic conditions
    Pedro Afonso Marques, Samuel Akatchi Ahizi, Miguel Alfonso Mendez
    arXivPreprintCC BY 4.0Full textApplied physics & instrumentationarXiv
  5. A dry cryogenic platform with picometre stability for nanoscale and quantum technologies
    William G. Roth, Andreas Thurn, Alexander M. Stramma +4
    arXivPreprintCC BY 4.0Full textApplied physics & instrumentationarXiv
BibTeX
@misc{yin2026cryogenic,
  title = {{Cryogenic performance and long-term stability of Hamamatsu R8520-506 photomultiplier tubes in liquid argon}},
  author = {Jilong Yin and Gaoshuang Li and Yike Shu and Yi Wang},
  howpublished = {arXiv},
  year = {2026},
  month = sep,
  doi = {10.48550/arxiv.2609.27718},
  eprint = {2609.27718},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2609.27718},
}
RIS
TY  - UNPB
TI  - Cryogenic performance and long-term stability of Hamamatsu R8520-506 photomultiplier tubes in liquid argon
AU  - Jilong Yin
AU  - Gaoshuang Li
AU  - Yike Shu
AU  - Yi Wang
PY  - 2026
DA  - 2026/09/23
PB  - arXiv
DO  - 10.48550/arxiv.2609.27718
UR  - https://arxiv.org/abs/2609.27718
AB  - Seven Hamamatsu R8520-506 photomultiplier tubes were characterized at room temperature and in liquid argon to evaluate their response and operating stability for liquid-argon light readout. A 405~nm pulsed light source was used to measure gain, peak-to-valley ratio, single-photoelectron charge resolution, and afterpulsing, while dark-count data were acquired with random triggers in the absence of active illumination. Gain and dark-count stability were monitored during a ten-day operating period in liquid argon. At a motherboard input voltage of 900~V, the extracted gains in liquid argon were approximately 27.8\%--51.3\% higher than the corresponding room-temperature values. For all seven tubes, the peak-to-valley ratios were higher and the single-photoelectron charge resolution improved in liquid argon. The dark count rates ranged from approximately 143 to 252~Hz at room temperature and from 166 to 199~Hz in liquid argon, with three tubes showing increases and four showing decreases. The mean number of detected afterpulses per selected primary event was approximately 21.0\%--63.6\% lower in liquid argon. The relative standard deviations across the ten daily means of gain and dark count rate did not exceed approximately 2.1\% and 3.3\%, respectively. The measurements provide a quantitative basis for considering R8520-506 tubes as photosensor candidates for liquid-argon detectors.
ER  -