Method Comparison of Three Automated D-Dimer Assays.
4 affiliations
- 1Department of Medical Biochemistry, Faculty of Medicine, Atlas University, 34408 Istanbul, Türkiye.
- 2Department of Chemistry, Yildiz Technical University, 34220 Istanbul, Türkiye.
- 3Department of Biophysics, Faculty of Medicine, Kütahya Health Sciences University, 43100 Kütahya, Türkiye.
- 4Department of Biostatistics and Medical Informatics, Faculty of Medicine, Istanbul Atlas University, 34408 Istanbul, Türkiye.
Abstract
Background and Objective: D-dimer results differ across analytical platforms, which may affect their clinical interpretation. This study aimed to compare the analytical agreement and threshold-based classification of three routinely used D-dimer analyzers. Methods: Of 150 consecutive residual citrated plasma specimens, 81 patient specimens with valid results on the Sysmex CS-2500, Tokra Novae MTI, and VIDAS 3 analyzers were included. Agreement was assessed using the intraclass correlation coefficient, Passing-Bablok regression, Bland-Altman analysis, and classification concordance at the fixed (500 ng/mL FEU) and age-adjusted thresholds. VIDAS 3 served solely as the comparator method. Results: Overall agreement was moderate to good; however, the platforms were not fully interchangeable. Sysmex CS-2500 and VIDAS 3 showed the closest agreement, with no significant constant or proportional bias, whereas Tokra Novae MTI exhibited significant proportional bias. Mixed classification across the three analyzers occurred in 21/81 patients (25.9%) at the fixed threshold and in 18/81 patients (22.2%) at the age-adjusted thresholds. Conclusions: Clinically relevant inter-platform differences persisted despite acceptable average agreement. Therefore, D-dimer results should not be regarded as interchangeable across analyzers, particularly near clinical decision thresholds, and platform-specific verification remains essential.
Related
- Agreement and reproducibility of intraoral scan- versus CBCT-based three-dimensional implant position assessment in the posterior mandible — a paired in vitro method-comparison study
- High-throughput, low volume d-ROMs and BAP assays: 384-well plate method for large-scale studies
- Multi‐platform comparison of plasma phosphorylated tau assays: analytical and workflow factors drive clinical implementation decisions
- Comparison of two automated PGMI-based software tools for breast positioning quality assessment in DBT screening
- Underestimation of Low-Density Lipoprotein Cholesterol (LDL-C) Using the Friedewald Equation at Higher Triglyceride Levels: Evidence From a Method-Comparison Study
BibTeX
@article{dundar2026method,
title = {{Method Comparison of Three Automated D-Dimer Assays.}},
author = {Bağnu Dündar and Betül Özbek Kurtul and Beyzanur Badilli and Sevgi Koçyiğit Sevinç and Elif Ertaş and Asiye Gök Yurttaş},
journal = {Diagnostics (Basel, Switzerland)},
year = {2026},
month = sep,
doi = {10.3390/diagnostics16193119},
url = {https://europepmc.org/article/MED/42882274},
note = {CC BY},
}
RIS
TY - JOUR
TI - Method Comparison of Three Automated D-Dimer Assays.
AU - Bağnu Dündar
AU - Betül Özbek Kurtul
AU - Beyzanur Badilli
AU - Sevgi Koçyiğit Sevinç
AU - Elif Ertaş
AU - Asiye Gök Yurttaş
PY - 2026
DA - 2026/09/25
JO - Diagnostics (Basel, Switzerland)
DO - 10.3390/diagnostics16193119
UR - https://europepmc.org/article/MED/42882274
AB - Background and Objective: D-dimer results differ across analytical platforms, which may affect their clinical interpretation. This study aimed to compare the analytical agreement and threshold-based classification of three routinely used D-dimer analyzers. Methods: Of 150 consecutive residual citrated plasma specimens, 81 patient specimens with valid results on the Sysmex CS-2500, Tokra Novae MTI, and VIDAS 3 analyzers were included. Agreement was assessed using the intraclass correlation coefficient, Passing-Bablok regression, Bland-Altman analysis, and classification concordance at the fixed (500 ng/mL FEU) and age-adjusted thresholds. VIDAS 3 served solely as the comparator method. Results: Overall agreement was moderate to good; however, the platforms were not fully interchangeable. Sysmex CS-2500 and VIDAS 3 showed the closest agreement, with no significant constant or proportional bias, whereas Tokra Novae MTI exhibited significant proportional bias. Mixed classification across the three analyzers occurred in 21/81 patients (25.9%) at the fixed threshold and in 18/81 patients (22.2%) at the age-adjusted thresholds. Conclusions: Clinically relevant inter-platform differences persisted despite acceptable average agreement. Therefore, D-dimer results should not be regarded as interchangeable across analyzers, particularly near clinical decision thresholds, and platform-specific verification remains essential.
SN - 2075-4418
ER -