Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults
Abstract
What are the main findings?• A seated bioelectrical impedance sensor accurately estimated appendicular muscle mass against DXA in older adults. • The Sergi prediction equation provided higher agreement and lower estimation error than the Kyle equation.
What are the implications of the main findings?
• Seated bioelectrical impedance enables rapid, practical muscle mass assessment without compromising measurement validity. • The phase angle derived from the sensor provides complementary information on muscle function and sarcopenia.
Related
- Agreement of Bioelectrical Impedance Analysis Prediction Equations with Dual-Energy X-Ray Absorptiometry for Tracking Resistance Training-Induced Changes in Fat-Free Mass in Older Women.
- Bioelectrical Impedance Analysis–Derived Phase Angle in Initial Medical and Nursing Assessment for Cardiac Rehabilitation: Association with Cardiac Rehabilitation Risk Category and Longitudinal Change During the Program—An Exploratory Prospective Observational Study
- Missing Link between Glycemic Control and Bioelectrical Impedance Analysis
- Bioelectrical impedance-derived body composition and systemic therapy outcomes in non-small cell lung cancer
- Association of electrical properties of bioelectrical impedance spectroscopy with sedentary time, physical activity and muscle function in older adults
BibTeX
@article{piche2026toward,
title = {{Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults}},
author = {Elodie Piche and Emeline Michel and Olivier Guerin and Etienne Gouraud and Frédéric Chorin},
journal = {Sensors},
year = {2026},
month = aug,
doi = {10.3390/s26165112},
url = {https://hal.science/hal-05733600v1},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults
AU - Elodie Piche
AU - Emeline Michel
AU - Olivier Guerin
AU - Etienne Gouraud
AU - Frédéric Chorin
PY - 2026
DA - 2026/08/12
JO - Sensors
DO - 10.3390/s26165112
UR - https://hal.science/hal-05733600v1
AB - What are the main findings?• A seated bioelectrical impedance sensor accurately estimated appendicular muscle mass against DXA in older adults. • The Sergi prediction equation provided higher agreement and lower estimation error than the Kyle equation. What are the implications of the main findings? • Seated bioelectrical impedance enables rapid, practical muscle mass assessment without compromising measurement validity. • The phase angle derived from the sensor provides complementary information on muscle function and sarcopenia.
SN - 1424-8220
ER -