Application of continuous intraoperative neuromonitoring (C-IONM) for recurrent laryngeal nerve preservation in transoral endoscopic thyroidectomy: a preliminary report from Vietnam.
5 affiliations
- 1Department of Oncology, Hanoi Medical University, No 1 Ton That Tung Street, Dong Da District, Hanoi, 100000, Viet Nam.
- 2Oncology Center, Hanoi Medical University Hospital, No 1 Ton That Tung Street, Dong Da District, Hanoi, 100000, Viet Nam.
- 3Vietnam National Cancer Hospital, No.43 Quan Su Street, Hoan Kiem District, Hanoi, Vietnam.
- 4Department of Oncology, Hanoi Medical University, No 1 Ton That Tung Street, Dong Da District, Hanoi, 100000, Viet Nam. [email protected].
- 5Oncology Center, Hanoi Medical University Hospital, No 1 Ton That Tung Street, Dong Da District, Hanoi, 100000, Viet Nam. [email protected].
Abstract
Background
The transoral endoscopic thyroidectomy vestibular approach (TOETVA) eliminates visible scarring but introduces technical challenges for recurrent laryngeal nerve (RLN) preservation. Continuous intraoperative neuromonitoring (C-IONM) provides dynamic assessment of nerve function, potentially reducing injury risk. This study reports the initial Vietnamese experience with C-IONM during TOETVA.Methods
Thirty-two patients undergoing TOETVA with C-IONM (NIM Vital system, Medtronic) for benign nodules or differentiated thyroid carcinoma were prospectively enrolled. An Automatic Periodic Stimulation (APS) electrode was placed on the vagus nerve for continuous monitoring. Intraoperative loss of signal (LOS) events were recorded, and all patients underwent postoperative laryngoscopy to assess vocal cord function.Results
C-IONM was successfully established in all 33 nerves at risk. Baseline median amplitude was 645.0 (IQR: 520.0-780.0) µV. Intraoperative LOS events occurred in 3 nerves (9.1%): 2 transient events resolved with maneuver adjustment without postoperative deficit; 1 sustained LOS (3.0%) resulted in temporary vocal cord palsy (full recovery within 3 months). The positive predictive value of sustained LOS for temporary palsy was 100%, and the negative predictive value of stable EMG was 100%. No permanent nerve injuries occurred.Conclusion
C-IONM is feasible during TOETVA and may provide useful real-time functional information to guide surgical decision-making. The two transient loss of signal events successfully reversed through timely surgical adjustments suggest that C-IONM may help prevent nerve injury during this technically demanding approach. Larger comparative studies are needed to confirm these preliminary findings.
Related
- The effect of routine intraoperative neuromonitoring on thyroidectomy complications : A propensity score-matched retrospective cohort study
- Thyroid-cartilage and anterior-laryngeal electrodes provide higher and more frequently adequate electromyographic signals in intraoperative neuromonitoring during thyroid surgery: a systematic review and meta-analysis
- Association of intraoperative mitoxantrone hydrochloride injection with incidental parathyroidectomy and parathyroid autotransplantation during gasless unilateral axillary endoscopic thyroidectomy: a propensity score-matched retrospective cohort study
- Motor Deficits After Insular Glioma Surgery: Association with Postoperative Ischemia and Intraoperative Neuromonitoring
- Intraoperative preservation of an extravascular implantable cardioverter-defibrillator lead during coronary artery bypass grafting.
BibTeX
@article{nguyen2026application,
title = {{Application of continuous intraoperative neuromonitoring (C-IONM) for recurrent laryngeal nerve preservation in transoral endoscopic thyroidectomy: a preliminary report from Vietnam.}},
author = {Xuan Hau Nguyen and Xuan Hien Nguyen and Nhat Tan Nguyen and Quang Le Van},
journal = {BMC surgery},
year = {2026},
month = aug,
doi = {10.1186/s12893-026-04137-y},
url = {https://europepmc.org/article/MED/43009906},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Application of continuous intraoperative neuromonitoring (C-IONM) for recurrent laryngeal nerve preservation in transoral endoscopic thyroidectomy: a preliminary report from Vietnam.
AU - Xuan Hau Nguyen
AU - Xuan Hien Nguyen
AU - Nhat Tan Nguyen
AU - Quang Le Van
PY - 2026
DA - 2026/08/13
JO - BMC surgery
DO - 10.1186/s12893-026-04137-y
UR - https://europepmc.org/article/MED/43009906
AB - Background The transoral endoscopic thyroidectomy vestibular approach (TOETVA) eliminates visible scarring but introduces technical challenges for recurrent laryngeal nerve (RLN) preservation. Continuous intraoperative neuromonitoring (C-IONM) provides dynamic assessment of nerve function, potentially reducing injury risk. This study reports the initial Vietnamese experience with C-IONM during TOETVA.Methods Thirty-two patients undergoing TOETVA with C-IONM (NIM Vital system, Medtronic) for benign nodules or differentiated thyroid carcinoma were prospectively enrolled. An Automatic Periodic Stimulation (APS) electrode was placed on the vagus nerve for continuous monitoring. Intraoperative loss of signal (LOS) events were recorded, and all patients underwent postoperative laryngoscopy to assess vocal cord function.Results C-IONM was successfully established in all 33 nerves at risk. Baseline median amplitude was 645.0 (IQR: 520.0-780.0) µV. Intraoperative LOS events occurred in 3 nerves (9.1%): 2 transient events resolved with maneuver adjustment without postoperative deficit; 1 sustained LOS (3.0%) resulted in temporary vocal cord palsy (full recovery within 3 months). The positive predictive value of sustained LOS for temporary palsy was 100%, and the negative predictive value of stable EMG was 100%. No permanent nerve injuries occurred.Conclusion C-IONM is feasible during TOETVA and may provide useful real-time functional information to guide surgical decision-making. The two transient loss of signal events successfully reversed through timely surgical adjustments suggest that C-IONM may help prevent nerve injury during this technically demanding approach. Larger comparative studies are needed to confirm these preliminary findings.
SN - 1471-2482
ER -