Peer-reviewedCC BY 4.0Frontiers in Cardiovascular Medicine

Evaluation of a radiofrequency catheter ablation target based on averaged impedance drop for pulmonary vein isolation

  1. 1Department of Cardiology, Karolinska University Hospital
  2. 2Integrative Cardiovascular, Cancer and Ageing Research Unit (ICCA), Department of Medicine, Karolinska Institute

Abstract

Background: Data regarding optimal ablation parameters for the TactiFlex Sensor Enabled (TFSE) radiofrequency ablation (RF) catheter to guide pulmonary vein isolation (PVI) are limited. The use of impedance drop correlates with lesion formation but is hampered by cardiac and respiratory oscillation artifacts. A filtered impedance signal, average impedance drop (AID), is now available in the EnSite X EP system. The study aim was to evaluate the efficacy and safety of a novel ablation protocol based on proportional (%)AID using the TFSE catheter, including mandatory remapping to assess PVI durability.

Methods: Pulmonary vein isolation with the TFSE was performed in patients with atrial fibrillation (AF) undergoing first-time PVI. The ablation protocol consisted of reaching %AID ≥9 or reaching maximum allowed RF time of 20 s and 10 s anteriorly and posteriorly. Protocol-mandated remapping was performed ≥2 months following index ablation.

Results: The study included 26 patients (paroxysmal/persistent AF = 17/9). Protocol adherence was 93%, and first-pass isolation was achieved in 85% of left and 93% of right pulmonary veins (PVs). Mean %AID was 9.4 ± 0.98 and proportion of lesions with %AID ≥9 was 62.2 ± 15.5%. Durable PVI was observed in 93% of all PVs at remapping (88% of patients) with no ablation-related complications. Inter-lesion distance, but not %AID, was significantly different in reconnected PV segments compared to segments with durable isolation (Firth logit <0.001).

Conclusion: Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.

Related

  1. Left Inferior Pulmonary Vein Thrombosis and Secondary Lung Infarction Following Radiofrequency Catheter Ablation: A Diagnostic Challenge.
    Karen M'bie, Pascale Bohy, Aurore Margot
    CureusPeer-reviewedCC BYFull textCardiovascular medicineEurope PMC
  2. Grid annotation vs ablation tags in predicting lesion area coverage for pulmonary vein isolation using a variable-loop circular catheter.
    Giampaolo Vetta, Charles Audiat, Lorenzo Marcon +9
    Heart rhythm O2Peer-reviewedCC BY-NC-NDFull textCardiovascular medicineEurope PMC
  3. Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility Study
    Masahiro Mizobuchi, Junya Matsumoto, Kaito Oka +3
    Journal of ArrhythmiaPeer-reviewedCC BY-NC-ND 4.0Full textCardiovascular medicinePMCEurope PMC
  4. Radiofrequency Catheter Ablation of Atrial Fibrillation in Patients with Anomalous Pulmonary Venous Return: A Multicenter Case Series.
    C. Estrada, O. Baqal, M. Singh Sidhu +6
    medRxivPreprintCC BY-NC-NDFull textCardiovascular medicinemedRxiv
  5. Pulmonary Vein Isolation with Different Single-Shot Ablation Devices in Patients with Atrial Fibrillation: A Prospective Observational Single-Center Clinical Evaluation
    Wiktor Czarkowski, Bartłomiej Popielarz, Agata Kukfisz +3
    Medicina (Kaunas, Lithuania)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
BibTeX
@article{akerstrom2026evaluation,
  title = {{Evaluation of a radiofrequency catheter ablation target based on averaged impedance drop for pulmonary vein isolation}},
  author = {Finn Åkerström and Emmanouil Charitakis and Fahd Asaad and Serkan Saygi and Emma Svennberg and Frieder Braunschweig and Mats Jensen-Urstad and Nikola Drca},
  journal = {Frontiers in Cardiovascular Medicine},
  year = {2026},
  month = nov,
  doi = {10.3389/fcvm.2026.1983617},
  url = {https://www.frontiersin.org/articles/10.3389/fcvm.2026.1983617/full},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Evaluation of a radiofrequency catheter ablation target based on averaged impedance drop for pulmonary vein isolation
AU  - Finn Åkerström
AU  - Emmanouil Charitakis
AU  - Fahd Asaad
AU  - Serkan Saygi
AU  - Emma Svennberg
AU  - Frieder Braunschweig
AU  - Mats Jensen-Urstad
AU  - Nikola Drca
PY  - 2026
DA  - 2026/11/16
JO  - Frontiers in Cardiovascular Medicine
DO  - 10.3389/fcvm.2026.1983617
UR  - https://www.frontiersin.org/articles/10.3389/fcvm.2026.1983617/full
AB  - Background: Data regarding optimal ablation parameters for the TactiFlex Sensor Enabled (TFSE) radiofrequency ablation (RF) catheter to guide pulmonary vein isolation (PVI) are limited. The use of impedance drop correlates with lesion formation but is hampered by cardiac and respiratory oscillation artifacts. A filtered impedance signal, average impedance drop (AID), is now available in the EnSite X EP system. The study aim was to evaluate the efficacy and safety of a novel ablation protocol based on proportional (%)AID using the TFSE catheter, including mandatory remapping to assess PVI durability.  Methods: Pulmonary vein isolation with the TFSE was performed in patients with atrial fibrillation (AF) undergoing first-time PVI. The ablation protocol consisted of reaching %AID ≥9 or reaching maximum allowed RF time of 20 s and 10 s anteriorly and posteriorly. Protocol-mandated remapping was performed ≥2 months following index ablation.  Results: The study included 26 patients (paroxysmal/persistent AF = 17/9). Protocol adherence was 93%, and first-pass isolation was achieved in 85% of left and 93% of right pulmonary veins (PVs). Mean %AID was 9.4 ± 0.98 and proportion of lesions with %AID ≥9 was 62.2 ± 15.5%. Durable PVI was observed in 93% of all PVs at remapping (88% of patients) with no ablation-related complications. Inter-lesion distance, but not %AID, was significantly different in reconnected PV segments compared to segments with durable isolation (Firth logit <0.001).  Conclusion: Point-by-point RF ablation using the TFSE catheter and a prespecified ablation protocol resulted in a high rate of first pass isolation and durable PVI (93%) at protocol-mandated remapping, without serious procedure-related complications.
SN  - 2297-055X
ER  -