Peer-reviewedCC BY-NC-ND 4.0Journal of Arrhythmia
Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility Study
- 1Cardiovascular Center Kyoto‐Katsura Hospital Kyoto Japan
- 2Department of Clinical Engineering Kyoto‐Katsura Hospital Kyoto Japan
Abstract
Background: PulseSelect pulmonary vein isolation (PVI) lacks a dedicated contact indicator.
Methods: We retrospectively compared 63 patients undergoing PVI before and after EnSite Contact Index (CI) implementation. Because electrical isolation does not confirm lesion transmurality, post‐procedural unipolar electrograms (u‐EGMs) were assessed as acute surrogate markers.
Results: CI guidance was associated with more applications, shorter fluoroscopy, and in an exploratory analysis of analyzable patients, fewer residual high‐voltage u‐EGMs (1/20 vs. 8/24; p = 0.027). No complications occurred.
Conclusions: CI‐guided PVI was feasible; these procedural and acute electrophysiological associations require prospective validation.
Related
- Evaluation of a radiofrequency catheter ablation target based on averaged impedance drop for pulmonary vein isolation
- Monitoring multiple rib fractures using pulmonary electrical impedance tomography technology: correlation analysis of pulmonary ventilation changes
- Durability and reconnection patterns at time of repeat ablation after first-time pulmonary vein isolation by monopolar focal pulsed field ablation
- Effects of Pulmonary Vein Isolation on Left Atrial Hemodynamic and Mechanical Function: Insights from Advanced and Emerging Echocardiographic Parameters in Responders and Non-Responders
- Sinus arrest during right pulmonary vein isolation related to right sinus node artery injury: a case report
BibTeX
@article{mizobuchi2026impedancebased,
title = {{Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility Study}},
author = {Masahiro Mizobuchi and Junya Matsumoto and Kaito Oka and Hotaka Takemura and Tomoki Yamashita and Shigeru Nakamura},
journal = {Journal of Arrhythmia},
year = {2026},
month = nov,
doi = {10.1002/joa3.70507},
url = {https://doi.org/10.1002/joa3.70507},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Impedance‐Based Contact Visualization for PulseSelect Pulmonary Vein Isolation: A Single‐Center Feasibility Study
AU - Masahiro Mizobuchi
AU - Junya Matsumoto
AU - Kaito Oka
AU - Hotaka Takemura
AU - Tomoki Yamashita
AU - Shigeru Nakamura
PY - 2026
DA - 2026/11/08
JO - Journal of Arrhythmia
DO - 10.1002/joa3.70507
UR - https://doi.org/10.1002/joa3.70507
AB - Background: PulseSelect pulmonary vein isolation (PVI) lacks a dedicated contact indicator. Methods: We retrospectively compared 63 patients undergoing PVI before and after EnSite Contact Index (CI) implementation. Because electrical isolation does not confirm lesion transmurality, post‐procedural unipolar electrograms (u‐EGMs) were assessed as acute surrogate markers. Results: CI guidance was associated with more applications, shorter fluoroscopy, and in an exploratory analysis of analyzable patients, fewer residual high‐voltage u‐EGMs (1/20 vs. 8/24; p = 0.027). No complications occurred. Conclusions: CI‐guided PVI was feasible; these procedural and acute electrophysiological associations require prospective validation.
SN - 1880-4276
SN - 1883-2148
ER -