Advancing GaN detectors toward proton therapy applications
Abstract
Proton therapy offers highly localized dose delivery for cancer treatment, but its clinical precision critically depends on accurate beam monitoring and patient imaging. Conventional detector technologies remain constrained by limited spatial resolution, dynamic range, and radiation tolerance. Here we demonstrate gallium nitride (GaN) diode arrays as a new platform for proton imaging. Using a one-dimensional GaN array, we image a 65 MeV clinical proton beam with linear response across three orders of magnitude, high sensitivity down to 10 pA, and robustness under irradiation up to 100 kGy. 2D profiles are reconstructed through transverse mechanical scanning with uncertainties in the beam position and linewidth as low as ± 0.07 mm and ± 0.14 mm, respectively. The intrinsic radiation hardness of GaN was experimentally confirmed, while key requirements for 65 MeV proton beam quality assurance and patient radiography were addressed. This proof of concept establishes the foundations for extending the approach to 2D detector matrices, combining high-resolution and real-time imaging.
Related
- Toward High-Resolution Detection of Target Fragmentation: TEA-Sensitized NIT for Proton Therapy Applications
- Review of Hybrid and Opaque Scintillator Technologies Advancing Neutrino and Dark Matter Detectors
- Clinical Commissioning of the First Ultra-Compact Gantry-less Upright Proton Therapy System
- Can future reirradiation justify upfront proton therapy? An extension of model-based proton therapy selection beyond primary-course NTCP benefit
- Impact of anatomical changes on simulated proton arc therapy and intensity‐modulated proton therapy for lung cancer
BibTeX
@article{siviero2026advancing,
title = {{Advancing GaN detectors toward proton therapy applications}},
author = {Matilde Siviero and Lucas Lesourd and Eric Frayssinet and Johan-Petter Hofverberg and Marie Vidal and Jean-Yves Duboz and Maxime Hugues},
journal = {Scientific Reports},
year = {2026},
month = aug,
doi = {10.1038/s41598-026-68432-7},
url = {https://hal.science/hal-05774815v1},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Advancing GaN detectors toward proton therapy applications
AU - Matilde Siviero
AU - Lucas Lesourd
AU - Eric Frayssinet
AU - Johan-Petter Hofverberg
AU - Marie Vidal
AU - Jean-Yves Duboz
AU - Maxime Hugues
PY - 2026
DA - 2026/08/26
JO - Scientific Reports
DO - 10.1038/s41598-026-68432-7
UR - https://hal.science/hal-05774815v1
AB - Proton therapy offers highly localized dose delivery for cancer treatment, but its clinical precision critically depends on accurate beam monitoring and patient imaging. Conventional detector technologies remain constrained by limited spatial resolution, dynamic range, and radiation tolerance. Here we demonstrate gallium nitride (GaN) diode arrays as a new platform for proton imaging. Using a one-dimensional GaN array, we image a 65 MeV clinical proton beam with linear response across three orders of magnitude, high sensitivity down to 10 pA, and robustness under irradiation up to 100 kGy. 2D profiles are reconstructed through transverse mechanical scanning with uncertainties in the beam position and linewidth as low as ± 0.07 mm and ± 0.14 mm, respectively. The intrinsic radiation hardness of GaN was experimentally confirmed, while key requirements for 65 MeV proton beam quality assurance and patient radiography were addressed. This proof of concept establishes the foundations for extending the approach to 2D detector matrices, combining high-resolution and real-time imaging.
SN - 2045-2322
ER -