Characterization and Validation of MicroDiamond Detectors in Patient-specific Quality Assurance of Magnetic Resonance-guided Stereotactic Radiotherapy

Abstract

Introduction: A dosimetric performance analysis of microDiamond (MD) detector in magnetic resonance imaging linear accelerators (MR-Linac) was performed to evaluate its utility for patient-specific quality assurance (PSQA) in stereotactic radiotherapy (SRT). Methods: Output factors (OFs), depth dose curves, and dose profiles for field sizes from 1 cm × 1 cm to 10 cm × 10 cm were measured in an Elekta Unity 1.5 T MR-Linac. Measurements were conducted using an MR-Safe three-dimensional (3D) water phantom with a PTW Semiflex 3D ionization chamber and a PTW MD detector. Angular response was assessed for orientations perpendicular to the beam and parallel to the magnetic field; correction factors were determined for gantry angles from 0° to 360° in 20° increments. Detectors were cross-calibrated against a 0.6 cc Farmer chamber, and patient-specific point-dose verification was performed for SRT plans. Results: Average dmax measured with the MD detector (1.28 ± 0.1 cm) agreed closely with the Semiflex 3D (1.27 ± 0.1 cm). The Semiflex 3D overestimated penumbra by 1.8 ± 0.6 mm and underestimated small-field OFs, while the MD detector showed slight overresponse (0.5%). The angular dependence remains substantial and highly sensitive to field size, requiring field-size-specific angular correction factors to achieve clinical accuracy better than 2%. After cross-calibration, the difference between the measured and planned doses for the 1.2 cc target was − 6.92% for Semiflex 3D and 1.36% for MD detector. Conclusion: The MD detector was successfully validated for MR-Linac. Asymmetric angular dependence due to the Lorentz force was confirmed, necessitating gantry angle-specific corrections for dose estimation. The MD detector demonstrated superior accuracy for small-field SRT targets, confirming its significance for precise PSQA.

Related

  1. Efficacy and safety of single-fraction stereotactic body radiotherapy (SBRT) for the curative treatment of extracranial primary Tumors: A collaborative systematic review on behalf of the SBRT Focus Group of the European Society for Radiotherapy & Oncology (ESTRO).
    Francesco Cuccia, Ciro Franzese, Suraiya Dubash +9
    Clinical and translational radiation oncologyPeer-reviewedCC BY-NC-NDFull textApplied physics & instrumentationEurope PMC
  2. Dosimetric and Treatment Delivery Efficiency Comparison of Two Ring-gantry Linear Accelerator Platforms for Liver Stereotactic Body Radiotherapy
    Srikanth Guduru, Rohith Singareddy, N. V. N. Madhusudhana Sresty +2
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
  3. Retrospective intrafraction motion verification using triggered kilovoltage images during bone metastases stereotactic body radiotherapy
    Quentin Techer, Noé Grandgirard, Saturnin Sandjong +2
    Physics and Imaging in Radiation OncologyPeer-reviewedCC BY 4.0Full textSystems & synthetic biologyMedical imaging & radiologyHAL
  4. 3D analysis and evolution of prostate and tumor oxygenation using MRI-derived relative oxygen extraction fraction during MR-guided stereotactic body radiotherapy☆
    Thomas Berger, Ariane Lapierre, Perla Zidane +8
    Physics and imaging in radiation oncologyPeer-reviewedCC BYFull textApplied physics & instrumentationEurope PMCHAL
  5. Assessment of Calculation Grid Size Impact on Plan Quality and Gamma Evaluation Metrics in Intracranial Stereotactic Radiotherapy
    Ravindra Shende, Joyce Alexis, R. S. Rahul
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
BibTeX
@article{choudhary2026characterization,
  title = {{Characterization and Validation of MicroDiamond Detectors in Patient-specific Quality Assurance of Magnetic Resonance-guided Stereotactic Radiotherapy}},
  author = {Sanjana Choudhary and Chithra Krishnamoorthy and Radhakrishnan B. Nair and S. D. Sharma and Vinayaka Shenoy},
  journal = {Journal of Medical Physics},
  year = {2026},
  month = jul,
  doi = {10.4103/jmp.jmp_123_26},
  url = {https://doi.org/10.4103/jmp.jmp_123_26},
  note = {CC BY-NC-SA},
}
RIS
TY  - JOUR
TI  - Characterization and Validation of MicroDiamond Detectors in Patient-specific Quality Assurance of Magnetic Resonance-guided Stereotactic Radiotherapy
AU  - Sanjana Choudhary
AU  - Chithra Krishnamoorthy
AU  - Radhakrishnan B. Nair
AU  - S. D. Sharma
AU  - Vinayaka Shenoy
PY  - 2026
DA  - 2026/07/01
JO  - Journal of Medical Physics
DO  - 10.4103/jmp.jmp_123_26
UR  - https://doi.org/10.4103/jmp.jmp_123_26
AB  - Introduction: A dosimetric performance analysis of microDiamond (MD) detector in magnetic resonance imaging linear accelerators (MR-Linac) was performed to evaluate its utility for patient-specific quality assurance (PSQA) in stereotactic radiotherapy (SRT). Methods: Output factors (OFs), depth dose curves, and dose profiles for field sizes from 1 cm × 1 cm to 10 cm × 10 cm were measured in an Elekta Unity 1.5 T MR-Linac. Measurements were conducted using an MR-Safe three-dimensional (3D) water phantom with a PTW Semiflex 3D ionization chamber and a PTW MD detector. Angular response was assessed for orientations perpendicular to the beam and parallel to the magnetic field; correction factors were determined for gantry angles from 0° to 360° in 20° increments. Detectors were cross-calibrated against a 0.6 cc Farmer chamber, and patient-specific point-dose verification was performed for SRT plans. Results: Average dmax measured with the MD detector (1.28 ± 0.1 cm) agreed closely with the Semiflex 3D (1.27 ± 0.1 cm). The Semiflex 3D overestimated penumbra by 1.8 ± 0.6 mm and underestimated small-field OFs, while the MD detector showed slight overresponse (0.5%). The angular dependence remains substantial and highly sensitive to field size, requiring field-size-specific angular correction factors to achieve clinical accuracy better than 2%. After cross-calibration, the difference between the measured and planned doses for the 1.2 cc target was − 6.92% for Semiflex 3D and 1.36% for MD detector. Conclusion: The MD detector was successfully validated for MR-Linac. Asymmetric angular dependence due to the Lorentz force was confirmed, necessitating gantry angle-specific corrections for dose estimation. The MD detector demonstrated superior accuracy for small-field SRT targets, confirming its significance for precise PSQA.
SN  - 0971-6203
SN  - 1998-3913
ER  -