Non-contrast renal MR angiography: a within-subject comparison of imaging performance at 1.5 T, 3 T, and 5 T
8 affiliations
- 1The Graduate School of Fujian Medical University, Fuzhou, China
- 2Department of Radiology, Zhongshan Hospital (Xiamen), Fudan University, Xiamen, China
- 3Collaborative Innovation Department, United Imaging Healthcare, Shanghai, China
- 4Department of Electrocardiography Room, Zhongshan Hospital (Xiamen), Fudan University, Xiamen, China
- 5Department of Medical Imaging, Xiamen Municipal Clinical Research Center for Medical Imaging, Xiamen, China
- 6Department of Medical Imaging, Fujian Province Key Clinical Specialty for Medical Imaging, Xiamen, China
- 7Department of Imaging Big Data and Artificial Intelligence, Xiamen Key Laboratory of Clinical Transformation of Imaging Big Data and Artificial Intelligence, Xiamen, China
- 8Department of Radiology, Zhongshan Hospital, Fudan University, Shanghai, China
Abstract
Objective: This within-subject study compared non-contrast renal MRA (NC-MRA) at 1.5 T, 3 T and 5 T via comprehensive quantitative and subjective imaging assessments, and further explored imaging disparities between standard and high-resolution 5 T sequences.
Materials and methods: Thirty-nine healthy volunteers underwent standardized same-day non-contrast MRA scans across the three magnetic field strengths; both standard and high-resolution protocols were implemented on the 5 T scanner. Quantitative parameters including apparent signal-to-noise ratio (aSNR), apparent contrast-to-noise ratio (aCNR) and vessel sharpness were calculated using the erector spinae muscle as the local noise reference, an approach intended to reduce g-factor related noise bias between different scanners. Two senior radiologists performed subjective image scoring independently as blinded readers. All images were fully anonymized by removing scanner and sequence identifiers, and the presentation order of image datasets was randomly rearranged for each reviewer to minimize potential reading bias.
Results: Statistically significant inter-group differences were detected for all quantitative indicators (all p < 0.05). The 5 T standard protocol yielded the highest median aSNR and aCNR values, while the 5 T high-resolution sequence achieved optimal vessel sharpness and clearer visualization of distal subsegmental renal branches. Both 5 T protocols were superior to 1.5 T and 3 T sequences in terms of artifact suppression, vascular contour delineation and subjective diagnostic confidence.
Conclusion: The two tested 5 T NC-MRA protocols were associated with improved image-quality metrics and artifact scores in healthy volunteers. These findings provide quantitative baseline data for normal renal vasculature and support further clinical validation of NC-MRA protocols in patients with renal artery disease or contraindications to gadolinium-based contrast agents.
Related
- Hypertrophic obstructive cardiomyopathy complicated by severe midventricular obstruction treated with percutaneous transluminal septal myocardial ablation using simultaneous contrast imaging with left ventriculography and coronary angiography: a case report
- Patient-level diagnostic performance of myocardial contrast echocardiography using ultrasound-enhancing agents after acute myocardial infarction: comparison with radionuclide perfusion-metabolism imaging
- Quantitative comparison of virtual non-contrast and true non-contrast computed tomography pulmonary angiography images across eight thoracic tissue types: A retrospective cross-sectional study
- Diagnostic Performance of Photon-Counting Coronary CT Angiography in Patients Undergoing TAVI: A Comparison With Invasive Angiography.
- Comparison between angiography or intravascular imaging guided coronary rotational atherectomy
BibTeX
@article{xie2026noncontrast,
title = {{Non-contrast renal MR angiography: a within-subject comparison of imaging performance at 1.5 T, 3 T, and 5 T}},
author = {Anjie Xie and Min Li and Jiaqi Wang and Chunmei Wang and Xiaoxia Li and Liuhong Zhu and Funan Wang and Jianjun Zhou},
journal = {BMC Medical Imaging},
year = {2026},
month = aug,
doi = {10.1186/s12880-026-02723-3},
url = {https://doi.org/10.1186/s12880-026-02723-3},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Non-contrast renal MR angiography: a within-subject comparison of imaging performance at 1.5 T, 3 T, and 5 T
AU - Anjie Xie
AU - Min Li
AU - Jiaqi Wang
AU - Chunmei Wang
AU - Xiaoxia Li
AU - Liuhong Zhu
AU - Funan Wang
AU - Jianjun Zhou
PY - 2026
DA - 2026/08/22
JO - BMC Medical Imaging
DO - 10.1186/s12880-026-02723-3
UR - https://doi.org/10.1186/s12880-026-02723-3
AB - Objective: This within-subject study compared non-contrast renal MRA (NC-MRA) at 1.5 T, 3 T and 5 T via comprehensive quantitative and subjective imaging assessments, and further explored imaging disparities between standard and high-resolution 5 T sequences. Materials and methods: Thirty-nine healthy volunteers underwent standardized same-day non-contrast MRA scans across the three magnetic field strengths; both standard and high-resolution protocols were implemented on the 5 T scanner. Quantitative parameters including apparent signal-to-noise ratio (aSNR), apparent contrast-to-noise ratio (aCNR) and vessel sharpness were calculated using the erector spinae muscle as the local noise reference, an approach intended to reduce g-factor related noise bias between different scanners. Two senior radiologists performed subjective image scoring independently as blinded readers. All images were fully anonymized by removing scanner and sequence identifiers, and the presentation order of image datasets was randomly rearranged for each reviewer to minimize potential reading bias. Results: Statistically significant inter-group differences were detected for all quantitative indicators (all p < 0.05). The 5 T standard protocol yielded the highest median aSNR and aCNR values, while the 5 T high-resolution sequence achieved optimal vessel sharpness and clearer visualization of distal subsegmental renal branches. Both 5 T protocols were superior to 1.5 T and 3 T sequences in terms of artifact suppression, vascular contour delineation and subjective diagnostic confidence. Conclusion: The two tested 5 T NC-MRA protocols were associated with improved image-quality metrics and artifact scores in healthy volunteers. These findings provide quantitative baseline data for normal renal vasculature and support further clinical validation of NC-MRA protocols in patients with renal artery disease or contraindications to gadolinium-based contrast agents.