Impact of tissue heterogeneity on intracranial Cesium-131 brachytherapy dosimetry: A patient-specific Monte Carlo Study.

5 affiliations
  1. 1University of Minnesota, Minneapolis, Minnesota, USA.
  2. 2East Carolina University, Greenville, North Carolina, USA.
  3. 3University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  4. 4Kathmandu University, Nepal.
  5. 5Loyola University Medical Center, Maywood, Illinois, USA.

Abstract

Background

AAPM TG-43-based dose calculations for intracranial cesium-131 (¹³¹Cs) brachytherapy ignore patient-specific heterogeneities, which can affect dose accuracy. This study investigates the dosimetric impact of these heterogeneity effects using Monte Carlo (MC) simulations and patient-specific imaging data.Purpose

To quantify the impact of tissue heterogeneity on dose distributions in intracranial brachytherapy using ¹³¹Cs sources and to evaluate the accuracy of AAPM TG-43-based treatment planning system (TPS) calculations.Methods

TG-43 dosimetric parameters for the ¹³¹Cs source were independently characterized and benchmarked against published consensus data. MC simulations using MCNP were performed in both homogeneous water phantoms and heterogeneous voxelized lattice phantoms derived from CT DICOM RT data. Single and multiple-seed configurations were investigated. TPS and MC dose distributions were compared using dose-volume histogram (DVH) metrics such as (V₆₀, V₁₀₀, D2cc, and mean dose), isodose distributions, dose difference maps, and local and global dose difference analyzes. Spatial agreement was quantified using gamma analysis.Results

Strong agreement was observed between TPS and MC calculations in homogeneous media. However, heterogeneous patient-specific MC simulations revealed systematic differences in DVH metrics for the target volume and organs at risk, including the skull and skin. Local dose differences revealed pronounced deviations near high-density bony regions, where dose enhancement effects were observed, while global dose difference analyzes demonstrated an overall dosimetric bias. Gamma pass rates decreased in heterogeneous regions, consistent with these spatial dose discrepancies. Compared with the MC simulations, the TG-43-based TPS overestimated the mean dose to the target by approximately 23.63±4.97% , V₆₀ and V₁₀₀ by approximately 24.42±4.97% , and 23.75±4.97% , respectively, and the mean dose to the skin by approximately 73.60±4.97% , while underestimating the mean dose to the skull by 220.90±4.97% .Conclusion

Although the TG-43 formalism provides accurate dose estimates in homogeneous media, it introduces clinically meaningful local and global dosimetric deviations in heterogeneous intracranial environments. Careful consideration is required when reporting doses to the target and OARs, particularly given the potential for substantially higher doses to the irradiated skull in certain cases (up to 3.2-fold in this study), as well as the possibility of target underdosage within the high-risk clinical target volume (HR-CTV). Patient-specific MC modeling highlights the importance of incorporating tissue heterogeneity into intracranial brachytherapy dose calculations to improve dosimetric accuracy and optimize treatment planning.

Related

  1. Fast Patient-Specific Breast CT Dosimetry: 22-Fold Acceleration of Monte Carlo MGD Estimation
    Amir Entezam, Ashkan Pakzad, Christopher J. Hall +9
    arXivPreprintCC BY 4.0Full textBiological & medical physicsarXiv
  2. Contact-resolved deployment of the Contour Neurovascular System in patient-specific intracranial aneurysms
    Ratnadeep Pramanik, Fina Gießler, Martin Frank +5
    arXivPreprintarXiv licenceFull textApplied physics & instrumentationBiological & medical physicsarXiv
  3. Effect of heart rate on hemodynamics in patient-specific intracranial aneurysms: An in vitro study
    Baha Al-Deen T El-Khader, Nikhil Shirdade, Jana Korte +4
    PLOS OnePeer-reviewedCC BYFull textCardiovascular medicineFluid dynamicsPLOSEurope PMCPMC
  4. GammaTile® Brachytherapy as Spatially Fractionated Radiation Therapy for Postoperative Brain Cancer: A Paradigm for Precision Intracranial Radiotherapy
    Hualin Zhang, Nguyen Phuong Dang, Jason C. Ye +11
    CancersPeer-reviewedCC BYFull textApplied physics & instrumentationEurope PMCCrossref
  5. Dosimetric equivalence of deep learning prostate contours after LDR brachytherapy: pre-declared margins, patient-level acceptance thresholds and the incremental predictive value of DVH indices
    Louis-Bernard St-Cyr, Anne Saint-Laurent, José Angel Lesteiro-Tejeda +7
    arXivPreprintCC BY-NC-ND 4.0Full textBiological & medical physicsarXiv
BibTeX
@article{adhikari2026impact,
  title = {{Impact of tissue heterogeneity on intracranial Cesium-131 brachytherapy dosimetry: A patient-specific Monte Carlo Study.}},
  author = {Tirthraj Adhikari and Jae W Jung and Sherif M Gadoue and Puja Sharma and Tomas Montegegro and Farhad Jafari and Clara Ferreira},
  journal = {Journal of applied clinical medical physics},
  year = {2026},
  month = oct,
  doi = {10.1002/acm2.70872},
  url = {https://europepmc.org/article/MED/42958620},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Impact of tissue heterogeneity on intracranial Cesium-131 brachytherapy dosimetry: A patient-specific Monte Carlo Study.
AU  - Tirthraj Adhikari
AU  - Jae W Jung
AU  - Sherif M Gadoue
AU  - Puja Sharma
AU  - Tomas Montegegro
AU  - Farhad Jafari
AU  - Clara Ferreira
PY  - 2026
DA  - 2026/10/29
JO  - Journal of applied clinical medical physics
DO  - 10.1002/acm2.70872
UR  - https://europepmc.org/article/MED/42958620
AB  - Background  AAPM TG-43-based dose calculations for intracranial cesium-131 (¹³¹Cs) brachytherapy ignore patient-specific heterogeneities, which can affect dose accuracy. This study investigates the dosimetric impact of these heterogeneity effects using Monte Carlo (MC) simulations and patient-specific imaging data.Purpose  To quantify the impact of tissue heterogeneity on dose distributions in intracranial brachytherapy using ¹³¹Cs sources and to evaluate the accuracy of AAPM TG-43-based treatment planning system (TPS) calculations.Methods  TG-43 dosimetric parameters for the ¹³¹Cs source were independently characterized and benchmarked against published consensus data. MC simulations using MCNP were performed in both homogeneous water phantoms and heterogeneous voxelized lattice phantoms derived from CT DICOM RT data. Single and multiple-seed configurations were investigated. TPS and MC dose distributions were compared using dose-volume histogram (DVH) metrics such as (V₆₀, V₁₀₀, D2cc, and mean dose), isodose distributions, dose difference maps, and local and global dose difference analyzes. Spatial agreement was quantified using gamma analysis.Results  Strong agreement was observed between TPS and MC calculations in homogeneous media. However, heterogeneous patient-specific MC simulations revealed systematic differences in DVH metrics for the target volume and organs at risk, including the skull and skin. Local dose differences revealed pronounced deviations near high-density bony regions, where dose enhancement effects were observed, while global dose difference analyzes demonstrated an overall dosimetric bias. Gamma pass rates decreased in heterogeneous regions, consistent with these spatial dose discrepancies. Compared with the MC simulations, the TG-43-based TPS overestimated the mean dose to the target by approximately 23.63±4.97% , V₆₀ and V₁₀₀ by approximately 24.42±4.97% , and 23.75±4.97% , respectively, and the mean dose to the skin by approximately 73.60±4.97% , while underestimating the mean dose to the skull by 220.90±4.97% .Conclusion  Although the TG-43 formalism provides accurate dose estimates in homogeneous media, it introduces clinically meaningful local and global dosimetric deviations in heterogeneous intracranial environments. Careful consideration is required when reporting doses to the target and OARs, particularly given the potential for substantially higher doses to the irradiated skull in certain cases (up to 3.2-fold in this study), as well as the possibility of target underdosage within the high-risk clinical target volume (HR-CTV). Patient-specific MC modeling highlights the importance of incorporating tissue heterogeneity into intracranial brachytherapy dose calculations to improve dosimetric accuracy and optimize treatment planning.
SN  - 1526-9914
ER  -