Are two readings as good as three for monthly linear accelerator output quality assurance?

6 affiliations
  1. 1Department of Radiology University of Chicago Chicago Illinois USA
  2. 2Department of Medical Physics University of Wisconsin Madison Wisconsin USA
  3. 3Department of Biology University of Central Florida Orlando Florida USA
  4. 4Neuroscience Program, School of Molecular and Cellular Biology University of Illinois Urbana‐Champaign Urbana Illinois USA
  5. 5Department of Radiation Oncology Virginia Commonwealth University Richmond Virginia USA
  6. 6Department of Radiation Medicine University of Wisconsin Madison Wisconsin USA

Abstract

Background: During monthly linear accelerator (linac) output quality assurance (QA), it is common practice to acquire three consecutive ionization chamber readings per beam energy and average them. Whether two readings carry the same QA decisions in practice has not been directly quantified against historical output data.

Purpose: This study evaluates whether reducing the number of readings per monthly output measurement from three to two preserves QA decision accuracy against the TG‐142 ± 2% action threshold. Within‐session variation was additionally evaluated against the ± 0.5% between‐reading criterion of TG‐198.

Methods: A retrospective analysis was performed on three years of monthly output QA data from four linear accelerators at a single institution, across eight beam energies (6, 9, 12, and 16 MeV electrons; 6, 6 FFF, 10, and 15 MV photons), with each check comprising three consecutive readings per energy. The analysis combined four components: (i) a paired‐difference comparison of per‐session means from two‐reading versus three‐reading protocols, summarized with mean differences, percent differences, and Cohen's d; (ii) an evaluation of within‐session reading variation against the TG‐198 criterion; (iii) a scenario analysis of false‐positive and false‐negative classifications against the TG‐142 action threshold, with the three‐reading decision taken as ground truth; and (iv) generalized linear mixed‐effects modeling of individual readings, with random intercepts for machine and measurement session, to test for systematic reading‐order shifts.

Results: Across all eight energies, the absolute mean difference between two‐ and three‐reading protocols was at most 0.006 nC, corresponding to less than 0.04% of the corrected output. The two protocols produced concordant pass/fail decisions in 961 of 963 recorded sessions (99.1–100% per energy). The two‐reading protocol produced no false negatives: the two sessions above the action threshold were detected by both protocols, and the only disagreements were two isolated false alarms (one at 6 MeV, one at 6 MV). Within‐session reading variation satisfied the TG‐198 ± 0.5% criterion in 91.9–100% of sessions per energy, and restricting attention to the first two readings never increased the fraction of sessions exceeding that criterion. Mixed‐effects modeling identified a small but statistically detectable reading‐order shift: relative to the first reading, the second and third photon readings were lower by approximately 0.05% and 0.12%, and the corresponding electron shifts were near 0.05%. In all cases the absolute magnitude of these shifts was below 0.18%, well within the ± 0.5% between‐reading variation limit recommended by TG‐198.

Conclusions: For the eight beam energies studied, monthly output QA decisions based on two readings were consistent with those based on three readings, with no false negatives and a reading‐order effect well within the TG‐198 ± 0.5% limit. Low‐energy electron beams (non‐normal repeatability, the largest single outlier) and 10 MV (most frequent exceedance of the 0.5% criterion) warrant closer attention. These results indicate that two‐reading protocols can maintain QA decision accuracy while reducing measurement time.

Related

  1. Six-year data from three magnetic resonance-guided radiotherapy systems demonstrate long-term stability of linear accelerator quality assurance
    Guido C. Hilgers, Simon J. Woodings, Thomas A. Foppen +4
    Physics and Imaging in Radiation OncologyPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationPMCEurope PMC
  2. Using artificial intelligence for clinical trial quality assurance in a multi-institutional neurovascular-sparing prostate radiotherapy trial
    Matthew Doucette, Ying Zhang, Chien-Yi Liao +10
    Scientific ReportsPeer-reviewedCC BY-NC-ND 4.0Applied physics & instrumentationCrossref
  3. Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator☆
    Nienke D. Sijtsema, Guus Grimbergen, Eline N. de Groot-van Breugel +2
    Physics and Imaging in Radiation OncologyPeer-reviewedCC BY 4.0Full textApplied physics & instrumentationPMC
  4. EP09.28 Quality assurance in a national randomized trial of radiation target volume reduction in glioblastoma: results from the contouring and dosimetry dummy run
    D Di Perri, C De Laet, M Lambrecht +28
    Neuro-OncologyPeer-reviewedPublisher termsFull textApplied physics & instrumentationPMC
  5. Dosimetric and Treatment Delivery Efficiency Comparison of Two Ring-gantry Linear Accelerator Platforms for Liver Stereotactic Body Radiotherapy
    Srikanth Guduru, Rohith Singareddy, N. V. N. Madhusudhana Sresty +2
    Journal of Medical PhysicsPeer-reviewedCC BY-NC-SAFull textApplied physics & instrumentationDOAJ
BibTeX
@article{stolen2026readings,
  title = {{Are two readings as good as three for monthly linear accelerator output quality assurance?}},
  author = {Ethan D. Stolen and Joseph B. Schulz and Eric D. Stolen and Ava Paek and Siyong Kim and James J. Sohn},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  month = nov,
  doi = {10.1002/acm2.70912},
  url = {https://doi.org/10.1002/acm2.70912},
  note = {CC BY 4.0},
}
RIS
TY  - JOUR
TI  - Are two readings as good as three for monthly linear accelerator output quality assurance?
AU  - Ethan D. Stolen
AU  - Joseph B. Schulz
AU  - Eric D. Stolen
AU  - Ava Paek
AU  - Siyong Kim
AU  - James J. Sohn
PY  - 2026
DA  - 2026/11/22
JO  - Journal of Applied Clinical Medical Physics
DO  - 10.1002/acm2.70912
UR  - https://doi.org/10.1002/acm2.70912
AB  - Background: During monthly linear accelerator (linac) output quality assurance (QA), it is common practice to acquire three consecutive ionization chamber readings per beam energy and average them. Whether two readings carry the same QA decisions in practice has not been directly quantified against historical output data.  Purpose: This study evaluates whether reducing the number of readings per monthly output measurement from three to two preserves QA decision accuracy against the TG‐142 ± 2% action threshold. Within‐session variation was additionally evaluated against the ± 0.5% between‐reading criterion of TG‐198.  Methods: A retrospective analysis was performed on three years of monthly output QA data from four linear accelerators at a single institution, across eight beam energies (6, 9, 12, and 16 MeV electrons; 6, 6 FFF, 10, and 15 MV photons), with each check comprising three consecutive readings per energy. The analysis combined four components: (i) a paired‐difference comparison of per‐session means from two‐reading versus three‐reading protocols, summarized with mean differences, percent differences, and Cohen's d; (ii) an evaluation of within‐session reading variation against the TG‐198 criterion; (iii) a scenario analysis of false‐positive and false‐negative classifications against the TG‐142 action threshold, with the three‐reading decision taken as ground truth; and (iv) generalized linear mixed‐effects modeling of individual readings, with random intercepts for machine and measurement session, to test for systematic reading‐order shifts.  Results: Across all eight energies, the absolute mean difference between two‐ and three‐reading protocols was at most 0.006 nC, corresponding to less than 0.04% of the corrected output. The two protocols produced concordant pass/fail decisions in 961 of 963 recorded sessions (99.1–100% per energy). The two‐reading protocol produced no false negatives: the two sessions above the action threshold were detected by both protocols, and the only disagreements were two isolated false alarms (one at 6 MeV, one at 6 MV). Within‐session reading variation satisfied the TG‐198 ± 0.5% criterion in 91.9–100% of sessions per energy, and restricting attention to the first two readings never increased the fraction of sessions exceeding that criterion. Mixed‐effects modeling identified a small but statistically detectable reading‐order shift: relative to the first reading, the second and third photon readings were lower by approximately 0.05% and 0.12%, and the corresponding electron shifts were near 0.05%. In all cases the absolute magnitude of these shifts was below 0.18%, well within the ± 0.5% between‐reading variation limit recommended by TG‐198.  Conclusions: For the eight beam energies studied, monthly output QA decisions based on two readings were consistent with those based on three readings, with no false negatives and a reading‐order effect well within the TG‐198 ± 0.5% limit. Low‐energy electron beams (non‐normal repeatability, the largest single outlier) and 10 MV (most frequent exceedance of the 0.5% criterion) warrant closer attention. These results indicate that two‐reading protocols can maintain QA decision accuracy while reducing measurement time.
SN  - 1526-9914
ER  -