Magnetically-Controlled Capsule Endoscopy Method to Study Gastric Motility: A Feasibility Study.
5 affiliations
- 1Vatche and Tamar Manoukian Division of Digestive Diseases, University of California Los Angeles, Los Angeles, California, USA.
- 2G. Oppenheimer Center for Neurobiology of Stress and Resilience, Los Angeles, California, USA.
- 3Division of Gastroenterology, University of Washington, Seattle, Washington, USA.
- 4Division of Gastroenterology, Baylor Scott and White Health, Dallas, Texas, USA.
- 5Division of Gastroenterology and Hepatology, Mayo Clinic Scottsdale, Scottsdale, Arizona, USA.
Abstract
Background/aims
Current methods to study antropyloric contractility in functional dyspepsia (FD) and gastroparesis (GP) have drawbacks partly because their invasive nature may alter normal stomach physiology. This pilot study evaluated the technical feasibility of a non-invasive magnetically-controlled capsule endoscopy (MCCE) system to characterize gastric motility in FD and GP.Methods
Six healthy controls (HCs), 6 FD participants, and 5 GP participants underwent MCCE testing to measure pyloric diameter and antral wave period before and after sham feeding with assistance from an artificial intelligence deep learning model. Results were related to symptoms enumerated by the Gastroparesis Cardinal Symptom Index (GCSI) and Dyspepsia Severity Scale (DSS) questionnaires.Results
MCCE successfully assessed antropyloric motility in all participants. Pyloric diameters were 10.6 ± 3.7 mm in HC, 8.4 ± 4.5 mm in FD, and 14.1 ± 4.9 mm in GP. Antral wave periods in HC, FD, and GP were 20.0 ± 2.3, 25.1 ± 4.4, and 24.1 ± 5.5 seconds before sham feeding and 20.0 ± 1.5, 25.8 ± 4.5, and 27.6 ± 9.7 seconds after sham feeding. Effect sizes (Cohen's d > 0.8) were observed across multiple comparisons. DSS, but not GCSI, scores correlated positively with wave periods after sham feeding (rho = 0.69, p = 0.04) but not pyloric diameters (rho = 0.19, p = 0.54).Conclusions
MCCE enabled the acquisition of analyzable antral contraction and pyloric opening measurements in FD and GP. These findings support technical feasibility. Large effect sizes suggest differences in pyloric diameter and antral periodicity between groups, although further studies are warranted to provide analytical validation of measurement accuracy, reproducibility, and physiological comparability.
Related
- Investigating the contextual modulation of postoperative gastrointestinal motility: a methodological feasibility study combining a standardized intervention protocol with wireless motility capsule monitoring.
- Magnetically gated millimeter-scale capsule power system for leadless electrotherapy
- Concordancia diagnóstica entre la cápsula endoscópica y la enteroscopía de doble balón en la hemorragia digestiva de intestino medio en un hospital de referencia en Perú Diagnostic Agreement Between Capsule Endoscopy and Double-Balloon Enteroscopy in Mid-Gut Gastrointestinal Bleeding at a Referral Hospital in Peru
- Rare Disease Classification in Video Capsule Endoscopy
- Artificial Intelligence Approaches for Gastrointestinal Bleeding Detection in Capsule Endoscopy: A Systematic Review and Critical Analysis
BibTeX
@article{sohn2026magneticallycontrolled,
title = {{Magnetically-Controlled Capsule Endoscopy Method to Study Gastric Motility: A Feasibility Study.}},
author = {Jessica Sohn and Rahul Kamath and Dean Ehrlich and Swapna Mahurkar-Joshi and Alireza Sedarat and William L Hasler and Lin Chang},
journal = {Neurogastroenterology and motility},
year = {2026},
month = nov,
doi = {10.1111/nmo.70516},
url = {https://europepmc.org/article/MED/43079784},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Magnetically-Controlled Capsule Endoscopy Method to Study Gastric Motility: A Feasibility Study.
AU - Jessica Sohn
AU - Rahul Kamath
AU - Dean Ehrlich
AU - Swapna Mahurkar-Joshi
AU - Alireza Sedarat
AU - William L Hasler
AU - Lin Chang
PY - 2026
DA - 2026/11/01
JO - Neurogastroenterology and motility
DO - 10.1111/nmo.70516
UR - https://europepmc.org/article/MED/43079784
AB - Background/aims Current methods to study antropyloric contractility in functional dyspepsia (FD) and gastroparesis (GP) have drawbacks partly because their invasive nature may alter normal stomach physiology. This pilot study evaluated the technical feasibility of a non-invasive magnetically-controlled capsule endoscopy (MCCE) system to characterize gastric motility in FD and GP.Methods Six healthy controls (HCs), 6 FD participants, and 5 GP participants underwent MCCE testing to measure pyloric diameter and antral wave period before and after sham feeding with assistance from an artificial intelligence deep learning model. Results were related to symptoms enumerated by the Gastroparesis Cardinal Symptom Index (GCSI) and Dyspepsia Severity Scale (DSS) questionnaires.Results MCCE successfully assessed antropyloric motility in all participants. Pyloric diameters were 10.6 ± 3.7 mm in HC, 8.4 ± 4.5 mm in FD, and 14.1 ± 4.9 mm in GP. Antral wave periods in HC, FD, and GP were 20.0 ± 2.3, 25.1 ± 4.4, and 24.1 ± 5.5 seconds before sham feeding and 20.0 ± 1.5, 25.8 ± 4.5, and 27.6 ± 9.7 seconds after sham feeding. Effect sizes (Cohen's d > 0.8) were observed across multiple comparisons. DSS, but not GCSI, scores correlated positively with wave periods after sham feeding (rho = 0.69, p = 0.04) but not pyloric diameters (rho = 0.19, p = 0.54).Conclusions MCCE enabled the acquisition of analyzable antral contraction and pyloric opening measurements in FD and GP. These findings support technical feasibility. Large effect sizes suggest differences in pyloric diameter and antral periodicity between groups, although further studies are warranted to provide analytical validation of measurement accuracy, reproducibility, and physiological comparability.
SN - 1350-1925
SN - 1365-2982
ER -