The MIME framework for non-invasive longitudinal assessment of inspiratory effort during pressure support ventilation: a physiological validation study.
17 affiliations
- 1School of Electrical Engineering, Computing and Mathematical Sciences, Curtin University, Perth, Australia.
- 2Department of Emergency, ASST Spedali Civili University Hospital, Brescia, Italy.
- 3Department of Medicine and Surgery, University of Parma, Parma, Italy.
- 4Department of Precision-Regenerative Medicine and Jonic Area (DiMePRe-J), Section of Anesthesiology and Intensive Care Medicine, University of Bari "Aldo Moro", Bari, Italy.
- 5Department of Translational Medicine, Azienda Ospedaliera Universitaria di Ferrara, Università di Ferrara, Ferrara, Italy.
- 6Departamento de Anestesia y Cuidados Críticos, Hospital Clínic, Institut D'Investigació August Pi i Sunyer, Barcelona, España.
- 7CIBER of Respiratory Disease, Instituto de Salud Carlos III, Madrid, Spain.
- 8Department of Anesthesia and Critical Care, Policlinico di Bari, Bari, Italy.
- 9Department of Mechanical Engineering, University of Canterbury, Christchurch, 8041, New Zealand.
- 10Department of Medical and Surgical Specialties, Radiological Science and Public Health, Università di Brescia, Brescia, Italy.
- 11Alessandra BONO' Interdepartmental University Research Center on LOng Term Outcome (LOTO) in Critical Illness Survivors, University of Brescia, Brescia, Italy.
- 12Department of Mechanical and Industrial Engineering, Università di Brescia, Brescia, Italy.
- 13Interdepartmental Division of Critical Care Medicine, University of Toronto, Ontario, Canada.
- 14Department of Physiology, University of Toronto, Toronto, ON, Canada.
- 15Curtin Centre for Optimisation and Decision Science, Curtin University, Brescia, Australia.
- 16Department of Emergency, ASST Spedali Civili University Hospital, Brescia, Italy. [email protected].
- 17Department of Medical and Surgical Specialties, Radiological Science and Public Health, Università di Brescia, Brescia, Italy. [email protected].
Abstract
Background
Continuous monitoring of inspiratory effort during assisted mechanical ventilation remains challenging, yet it is vital as the transition from controlled to assisted mechanical ventilation requires careful titration of inspiratory effort to ensure patient safety. The Mixed Integer Multi-Effort (MIME) method is a physiological modeling approach that uses standard ventilator signals to non-invasively estimate inspiratory muscle pressure (Pmus). We evaluated the accuracy of MIME during pressure support ventilation (PSV) in patients with acute hypoxemic respiratory failure (AHRF).Methods
We conducted a prospective physiological validation study in adult patients with AHRF within 48 h of their first transition from controlled mechanical ventilation to PSV. Airway and esophageal pressure, flow, and volume, were recorded and analyzed offline with the MIME algorithm, which estimates Pmus by reconstructing airway pressure and separating ventilator-delivered from patient-generated pressure using a respiratory mechanics-based model. MIME-derived Pmus estimates were compared with esophageal pressure-derived Pmus using Bland-Altman analysis with mixed-effects modeling to account for repeated measurements within patients. Receiver operating characteristic analysis was used to evaluate MIME's ability to detect high and low inspiratory effort and to compare its performance with end-expiratory occlusion pressure (ΔPocc).Results
Sixty-two independent model-fitting windows (1240 breaths) from 18 patients were analyzed. Median inspiratory Pmus was 7.5 (IQR 6.0-9.1) cmH₂O for MIME-derived estimates and 6.9 (5.9-8.9) cmH₂O for esophageal pressure-derived measurements. Breath-by-breath analysis demonstrated a mean difference of 0.4 cmH₂O with limits of agreement ranging from - 3.4 to 4.2 cmH₂O. Agreement between methods improved when inspiratory effort was averaged over multiple breaths. Averaged values within each analyzed recording set showed a mean difference of 0.4 cmH₂O with limits of agreement ranging from - 2.2 to 3.0 cmH₂O. MIME demonstrated good discriminative performance in detecting both high and low inspiratory effort (AUROC 0.83 and 0.88, respectively). Compared with ΔPocc, MIME showed similar performance for detecting low inspiratory effort, whereas ΔPocc performed better for detecting high inspiratory effort.Conclusions
The MIME method provided non-invasive estimates of inspiratory effort during pressure support ventilation and may complement established physiological tools such as ΔPocc by supporting longitudinal assessment. Validation in broader populations and physiological settings, together with dedicated assessment of convergence and computational performance, is required before real-time bedside implementation.
Related
- The Missing Dimension in Lung-Protective Ventilation: Physiologically Directed Inspiratory and Expiratory Time
- Airway occlusions to measure inspiratory effort, respiratory drive, and lung mechanics during noninvasive ventilation
- Post-Exercise Increase in Maximal Voluntary Ventilation in Swimmers and Water Polo Players: Different Changes in Spirometric Indices and Inspiratory Flow–Volume Loop Parameters
- Suboptimal peak inspiratory flow, COPD exacerbations, hospitalisations and disease burden: a longitudinal follow-up in a multicentre real-world study.
- Predictive performance of bedside physiological parameters for the failure of prophylactic non-invasive ventilation after extubation.
BibTeX
@article{lindup2026mime,
title = {{The MIME framework for non-invasive longitudinal assessment of inspiratory effort during pressure support ventilation: a physiological validation study.}},
author = {K Lindup and S Giovanazzi and S Renzetti and F Murgolo and M Falappi and D Nodari and S Spadaro and C Ferrando and S Grasso and C Zhou and J Chase and N Latronico and A Visioli and E C Goligher and S Piva and F Padula and Michele Bertoni},
journal = {Critical care (London, England)},
year = {2026},
month = sep,
doi = {10.1186/s13054-026-06292-3},
url = {https://europepmc.org/article/MED/43079619},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - The MIME framework for non-invasive longitudinal assessment of inspiratory effort during pressure support ventilation: a physiological validation study.
AU - K Lindup
AU - S Giovanazzi
AU - S Renzetti
AU - F Murgolo
AU - M Falappi
AU - D Nodari
AU - S Spadaro
AU - C Ferrando
AU - S Grasso
AU - C Zhou
AU - J Chase
AU - N Latronico
AU - A Visioli
AU - E C Goligher
AU - S Piva
AU - F Padula
AU - Michele Bertoni
PY - 2026
DA - 2026/09/01
JO - Critical care (London, England)
DO - 10.1186/s13054-026-06292-3
UR - https://europepmc.org/article/MED/43079619
AB - Background Continuous monitoring of inspiratory effort during assisted mechanical ventilation remains challenging, yet it is vital as the transition from controlled to assisted mechanical ventilation requires careful titration of inspiratory effort to ensure patient safety. The Mixed Integer Multi-Effort (MIME) method is a physiological modeling approach that uses standard ventilator signals to non-invasively estimate inspiratory muscle pressure (Pmus). We evaluated the accuracy of MIME during pressure support ventilation (PSV) in patients with acute hypoxemic respiratory failure (AHRF).Methods We conducted a prospective physiological validation study in adult patients with AHRF within 48 h of their first transition from controlled mechanical ventilation to PSV. Airway and esophageal pressure, flow, and volume, were recorded and analyzed offline with the MIME algorithm, which estimates Pmus by reconstructing airway pressure and separating ventilator-delivered from patient-generated pressure using a respiratory mechanics-based model. MIME-derived Pmus estimates were compared with esophageal pressure-derived Pmus using Bland-Altman analysis with mixed-effects modeling to account for repeated measurements within patients. Receiver operating characteristic analysis was used to evaluate MIME's ability to detect high and low inspiratory effort and to compare its performance with end-expiratory occlusion pressure (ΔPocc).Results Sixty-two independent model-fitting windows (1240 breaths) from 18 patients were analyzed. Median inspiratory Pmus was 7.5 (IQR 6.0-9.1) cmH₂O for MIME-derived estimates and 6.9 (5.9-8.9) cmH₂O for esophageal pressure-derived measurements. Breath-by-breath analysis demonstrated a mean difference of 0.4 cmH₂O with limits of agreement ranging from - 3.4 to 4.2 cmH₂O. Agreement between methods improved when inspiratory effort was averaged over multiple breaths. Averaged values within each analyzed recording set showed a mean difference of 0.4 cmH₂O with limits of agreement ranging from - 2.2 to 3.0 cmH₂O. MIME demonstrated good discriminative performance in detecting both high and low inspiratory effort (AUROC 0.83 and 0.88, respectively). Compared with ΔPocc, MIME showed similar performance for detecting low inspiratory effort, whereas ΔPocc performed better for detecting high inspiratory effort.Conclusions The MIME method provided non-invasive estimates of inspiratory effort during pressure support ventilation and may complement established physiological tools such as ΔPocc by supporting longitudinal assessment. Validation in broader populations and physiological settings, together with dedicated assessment of convergence and computational performance, is required before real-time bedside implementation.