Methodology for computational assessment of hemolysis in under-occlusive roller pumps operated at clinically relevant conditions.

  1. 1Faculty of Mechanical Engineering Brno University of Technology Brno Czech Republic.
  2. 2Department of Anaesthesia and Intensive Care St. Anne's University Hospital and Masaryk University Brno Czech Republic.

Abstract

Hemolysis remains a major concern in clinical procedures involving extracorporeal blood circuits, with the blood delivery pump often being its considerable source. This study focuses on roller pumps, which-alongside centrifugal pumps-are the main types used in clinical practice. The objective is to establish a comprehensive methodology for the computational evaluation of hemolysis in under-occlusive roller pumps, which, compared to fully occlusive operation, have the potential to reduce hemolysis while maintaining sufficient flow performance. The methodology employs transient 2D CFD simulations with a dynamic mesh, while hemolysis is evaluated using a shear-induced hemolysis model. The governing equations are solved using the transitional low-Re k-ω SST turbulence model with appropriate prism-layer refinement at the wall, adhering to y+ values lower than 1, which is essential for accurate hemolysis estimation. In contrast to existing studies, the simulations are complemented by experimental measurements from an in vitro circuit operated under clinically relevant settings, providing realistic boundary conditions and validation support. The core component of the circuit is an in-house roller pump that enables precise setting and control of the under-occlusion degree. The proposed methodology enables to estimate and compare hemolysis across five under-occlusion settings with progressively increasing gap size, each characterized by the corresponding reduction in flow rate. The results indicate that hemolysis levels remain comparable across under-occlusive settings with flow reductions of up to 37%, with no substantial decrease observed over this range. A more pronounced reduction in hemolysis is observed only at higher degrees of under-occlusion, corresponding to 50% and 63% flow reductions, which, however, result in clinically unacceptable losses in flow performance. Thus, the simulations suggest that the tightest under-occlusive setting with 10% flow reduction provides the most favorable balance between hemolysis and flow performance.

Related

  1. An Autonomous Robotic Phlebotomy Device: Extended Analytical Validation and Real-World Hemolysis Assessment
    Jeroen A. Roest, Luuk F. P. Giesen, Thijs C. van Holten +6
    Diagnostics (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. Clinically Informed Optic Disc and Optic Cup Segmentation for Automated Glaucoma Assessment Using Adaptive Feature Fusion Multi-Task U-Net++
    Burak Ören, Eser Sert
    BiomedicinesPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  3. Toward Clinically Feasible Assessment of Muscle Mass: Validation of a Seated Bioelectrical Impedance Device Against Dual-Energy X-Ray Absorptiometry (DXA) in Older Adults
    Elodie Piche, Emeline Michel, Olivier Guerin +2
    SensorsPeer-reviewedCC BY 4.0Full textPhysiologyHAL
  4. Evaluation of the psychometric properties of patient-reported outcome measures for the assessment of health-related quality of life across the European cancer continuum: a systematic review applying COSMIN methodology.
    Hugo Vachon, K Martičić Giljević, C Piccinin +21
    Health and quality of life outcomesPeer-reviewedCC BY-NC-NDFull textOncologyEurope PMCPMC
  5. Standardized Performance Assessment Methodology and Evaluation Framework for Tactical Combat Casualty Care Autonomous Documentation Algorithms: Framework Development and Evaluation.
    Jaeyeon Lee, Jeanette Little, Ethan Quist +2
    JMIR medical informaticsPeer-reviewedCC BYFull textEpidemiology & public healthEurope PMC
BibTeX
@article{kohut2026methodology,
  title = {{Methodology for computational assessment of hemolysis in under-occlusive roller pumps operated at clinically relevant conditions.}},
  author = {Jiří Kohút and Jiří Jagoš and Pavel Suk and Martin Formánek and Jiří Burša},
  journal = {Bioengineering \& translational medicine},
  year = {2026},
  month = nov,
  doi = {10.1002/btm2.70191},
  url = {https://europepmc.org/article/MED/43011939},
  note = {CC BY},
}
RIS
TY  - JOUR
TI  - Methodology for computational assessment of hemolysis in under-occlusive roller pumps operated at clinically relevant conditions.
AU  - Jiří Kohút
AU  - Jiří Jagoš
AU  - Pavel Suk
AU  - Martin Formánek
AU  - Jiří Burša
PY  - 2026
DA  - 2026/11/10
JO  - Bioengineering & translational medicine
DO  - 10.1002/btm2.70191
UR  - https://europepmc.org/article/MED/43011939
AB  - Hemolysis remains a major concern in clinical procedures involving extracorporeal blood circuits, with the blood delivery pump often being its considerable source. This study focuses on roller pumps, which-alongside centrifugal pumps-are the main types used in clinical practice. The objective is to establish a comprehensive methodology for the computational evaluation of hemolysis in under-occlusive roller pumps, which, compared to fully occlusive operation, have the potential to reduce hemolysis while maintaining sufficient flow performance. The methodology employs transient 2D CFD simulations with a dynamic mesh, while hemolysis is evaluated using a shear-induced hemolysis model. The governing equations are solved using the transitional low-Re k-ω SST turbulence model with appropriate prism-layer refinement at the wall, adhering to y+ values lower than 1, which is essential for accurate hemolysis estimation. In contrast to existing studies, the simulations are complemented by experimental measurements from an in vitro circuit operated under clinically relevant settings, providing realistic boundary conditions and validation support. The core component of the circuit is an in-house roller pump that enables precise setting and control of the under-occlusion degree. The proposed methodology enables to estimate and compare hemolysis across five under-occlusion settings with progressively increasing gap size, each characterized by the corresponding reduction in flow rate. The results indicate that hemolysis levels remain comparable across under-occlusive settings with flow reductions of up to 37%, with no substantial decrease observed over this range. A more pronounced reduction in hemolysis is observed only at higher degrees of under-occlusion, corresponding to 50% and 63% flow reductions, which, however, result in clinically unacceptable losses in flow performance. Thus, the simulations suggest that the tightest under-occlusive setting with 10% flow reduction provides the most favorable balance between hemolysis and flow performance.
SN  - 2380-6761
ER  -