Skip to content
DraconiticResearch
Search Sources API
197,768 studies31 sourcesUpdated 55 min ago
PreprintarXiv licencearXiv

Quantitative Interleaved-Pulse Acquisition Extends the Range of Isotope-Ratio Measurements in ToF-SIMS

Abstract

Isotope ratios underpin tracer studies of transport in oxides, battery materials, and biological systems, yet time-of-flight secondary ion mass spectrometry (ToF-SIMS) struggles to quantify ratios whose members differ in intensity by orders of magnitude: a primary-ion pulse long enough to count the minor isotope precisely can saturate the major isotope, whereas a pulse short enough to keep the major isotope linear yields too few minor-isotope counts. We show that multi-pulse-width acquisition yields quantitative isotope ratios when a dose factor measured within each crater links the phases and each isotope is taken only from phases in which it remains linear; we call this quantitative reconstruction interleaved-pulse acquisition. Pulse widths are selected from measured transfer curves and peak shapes. On thermal SiO2, a tuned (6, 35) ns pair achieves the same per-layer precision as short-pulse acquisition with a 13-fold reduction in the number of analysis frames. On 18O-enriched WOx films, interleaving extends the quantifiable range at the low-fraction end by about 14-fold. Across isotope ratios spanning approximately 1:500 to 1:1, the interleaved reconstruction agrees with the corresponding linear reference measurements, showing no measurable bias introduced by interleaving, and natural-abundance ratios of oxygen, strontium and titanium are reproduced within several percent. The method requires no hardware additions and enables depth-resolved isotope-ratio measurements over a wide range within a single crater.

Related

  1. XPCS-Echo and broad relaxation measurements using a bunch-mode data acquisition scheme
    William Chèvremont, Thomas Gibaud, Yuriy Chushkin +1
    arXivPreprintarXiv licenceFull textCondensed matterarXiv
  2. From Dye to Dynamics: Quantitative Measurements of Fluid Flows from Vortex Geometry
    Dustin Kleckner, Kevin Collins, Cole Cooper
    arXivPreprintarXiv licenceFull textFluid dynamicsarXiv
  3. Quantitative interpretation of Brookfield DV3TLV measurements: shear rate conversion, correction factors, and applicability limits
    A. E. Vasiliev, A. S. Besov, D. O. Andreev
    arXivPreprintarXiv licenceFull textGeophysics & seismologyApplied physics & instrumentationarXiv
  4. TETRIS-Q: Tiling-based Effective Transient-fault Reduction on Interleaved Superconducting Qubits
    Marzio Vallero, Gioele Casagranda, Flavio Vella +1
    arXivPreprintCC BY-SA 4.0Full textQuantum physics & quantum informationarXiv
  5. Phase-Altered Interleaved Randomized Benchmarking for Compiled Non-Clifford Gates
    Simona K. Grigorova, Nikolay V. Vitanov, Boyan T. Torosov
    arXivPreprintarXiv licenceFull textQuantum physics & quantum informationarXiv
BibTeX
@misc{ievlev2026quantitative,
  title = {{Quantitative Interleaved-Pulse Acquisition Extends the Range of Isotope-Ratio Measurements in ToF-SIMS}},
  author = {Anton V. Ievlev},
  howpublished = {arXiv},
  year = {2026},
  month = sep,
  doi = {10.48550/arxiv.2609.37413},
  eprint = {2609.37413},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2609.37413},
}
RIS
TY  - UNPB
TI  - Quantitative Interleaved-Pulse Acquisition Extends the Range of Isotope-Ratio Measurements in ToF-SIMS
AU  - Anton V. Ievlev
PY  - 2026
DA  - 2026/09/29
PB  - arXiv
DO  - 10.48550/arxiv.2609.37413
UR  - https://arxiv.org/abs/2609.37413
AB  - Isotope ratios underpin tracer studies of transport in oxides, battery materials, and biological systems, yet time-of-flight secondary ion mass spectrometry (ToF-SIMS) struggles to quantify ratios whose members differ in intensity by orders of magnitude: a primary-ion pulse long enough to count the minor isotope precisely can saturate the major isotope, whereas a pulse short enough to keep the major isotope linear yields too few minor-isotope counts. We show that multi-pulse-width acquisition yields quantitative isotope ratios when a dose factor measured within each crater links the phases and each isotope is taken only from phases in which it remains linear; we call this quantitative reconstruction interleaved-pulse acquisition. Pulse widths are selected from measured transfer curves and peak shapes. On thermal SiO2, a tuned (6, 35) ns pair achieves the same per-layer precision as short-pulse acquisition with a 13-fold reduction in the number of analysis frames. On 18O-enriched WOx films, interleaving extends the quantifiable range at the low-fraction end by about 14-fold. Across isotope ratios spanning approximately 1:500 to 1:1, the interleaved reconstruction agrees with the corresponding linear reference measurements, showing no measurable bias introduced by interleaving, and natural-abundance ratios of oxygen, strontium and titanium are reproduced within several percent. The method requires no hardware additions and enables depth-resolved isotope-ratio measurements over a wide range within a single crater.
ER  -