Interpreting Dose×LET volume histograms using difference mapping in Spot Delete proton planning.
- 1Covenant Health Proton Center, Knoxville, Tennessee, USA.
Abstract
Purpose
Dose-LET volume histograms (DLVHs) provide a more complete representation of proton dose deposition than conventional dose-volume histograms (DVHs), but their clinical interpretation remains challenging. We developed a clinically interpretable DLVH analysis framework and applied it to Spot Delete (SD) proton planning to determine whether biologically relevant changes in skin exposure could be identified beyond those detected by conventional dose-based evaluation.Methods
Twenty-one head-and-neck patients treated with pencil-beam scanning proton therapy were retrospectively replanned with and without SD using identical beam arrangements, optimization objectives, and robustness settings. Skin dose and dose-averaged linear energy transfer (LETd) were evaluated using DLVHs. Dose×LET difference mapping combined paired nonparametric testing, false-discovery-rate correction, effect-size analysis, and spatial-contiguity filtering to identify dominant regions of systematic change. Conventional target and skin DVH metrics were also compared.Results
Dose×LET difference mapping consistently identified a single dominant contiguous region characterized by moderate-to-high dose (≥10 GyRBE) and moderately elevated LET (≥3.0 keV/µm) that was reduced in SD plans across all patients. This region exhibited large effect sizes and strong statistical significance. The resulting volume reduction averaged 10.5 ± 6.8 cc per patient and was observed in all patients. Threshold-sensitivity analysis demonstrated persistence of the dominant elevated-LET region across LETd thresholds from 2.5 to 4.0 keV/µm. In contrast, conventional DVHs demonstrated only modest differences, while target coverage was preserved (ΔD95% < 0.2 GyRBE).Conclusions
Spot Delete planning produced a consistent reduction in elevated-LET skin exposure while maintaining target dose coverage. Conventional DVH analysis captured only small changes, whereas DLVH-based analysis revealed a systematic redistribution of biologically intensified dose away from skin. More broadly, these results show that reproducible differences in joint dose and LETd exposure arise from the interaction of dose and LET rather than extreme values of either alone and illustrate how incorporating LET-aware analysis can enhance routine plan comparison and support biologically informed decision-making in proton therapy.
Related
- PRIME-SVR: Physics-infoRmed Implicit Multi-Echo Slice-to-Volume Reconstruction for Fetal T2 mapping
- Interpreting the top quark mass using simulations of top-flavoured mesons
- Tokamak-SI: Interpreting Tokamak Discharges with Labeled Catalogs
- Interpreting the LZ 248 keV Event using Dark QCD
- Interpreting Liouvillian Eigenmodes: Quantized Orbits in Dissipative Systems
BibTeX
@article{hedrick2026interpreting,
title = {{Interpreting Dose×LET volume histograms using difference mapping in Spot Delete proton planning.}},
author = {Samantha G Hedrick and Laura Buchanan and Chester Ramsey},
journal = {Journal of applied clinical medical physics},
year = {2026},
month = nov,
doi = {10.1002/acm2.70901},
url = {https://europepmc.org/article/MED/43075780},
note = {CC BY},
}
RIS
TY - JOUR
TI - Interpreting Dose×LET volume histograms using difference mapping in Spot Delete proton planning.
AU - Samantha G Hedrick
AU - Laura Buchanan
AU - Chester Ramsey
PY - 2026
DA - 2026/11/26
JO - Journal of applied clinical medical physics
DO - 10.1002/acm2.70901
UR - https://europepmc.org/article/MED/43075780
AB - Purpose Dose-LET volume histograms (DLVHs) provide a more complete representation of proton dose deposition than conventional dose-volume histograms (DVHs), but their clinical interpretation remains challenging. We developed a clinically interpretable DLVH analysis framework and applied it to Spot Delete (SD) proton planning to determine whether biologically relevant changes in skin exposure could be identified beyond those detected by conventional dose-based evaluation.Methods Twenty-one head-and-neck patients treated with pencil-beam scanning proton therapy were retrospectively replanned with and without SD using identical beam arrangements, optimization objectives, and robustness settings. Skin dose and dose-averaged linear energy transfer (LETd) were evaluated using DLVHs. Dose×LET difference mapping combined paired nonparametric testing, false-discovery-rate correction, effect-size analysis, and spatial-contiguity filtering to identify dominant regions of systematic change. Conventional target and skin DVH metrics were also compared.Results Dose×LET difference mapping consistently identified a single dominant contiguous region characterized by moderate-to-high dose (≥10 GyRBE) and moderately elevated LET (≥3.0 keV/µm) that was reduced in SD plans across all patients. This region exhibited large effect sizes and strong statistical significance. The resulting volume reduction averaged 10.5 ± 6.8 cc per patient and was observed in all patients. Threshold-sensitivity analysis demonstrated persistence of the dominant elevated-LET region across LETd thresholds from 2.5 to 4.0 keV/µm. In contrast, conventional DVHs demonstrated only modest differences, while target coverage was preserved (ΔD95% < 0.2 GyRBE).Conclusions Spot Delete planning produced a consistent reduction in elevated-LET skin exposure while maintaining target dose coverage. Conventional DVH analysis captured only small changes, whereas DLVH-based analysis revealed a systematic redistribution of biologically intensified dose away from skin. More broadly, these results show that reproducible differences in joint dose and LETd exposure arise from the interaction of dose and LET rather than extreme values of either alone and illustrate how incorporating LET-aware analysis can enhance routine plan comparison and support biologically informed decision-making in proton therapy.
SN - 1526-9914
ER -