Peer-reviewedCC BY-NC-NDQuantitative imaging in medicine and surgery

Image quality and radiation dose optimization in CT pulmonary angiography using 120, 100, and 80 kV protocols for pulmonary thromboembolism diagnosis.

4 affiliations
  1. 1Centre of Diagnostic, Therapeutic & Investigative Studies (CODTIS), Faculty of Health Sciences, Universiti Kebangsaan Malaysia, Kuala Lumpur, Malaysia.
  2. 2Radiology Department, First Floor, Main Building, Hospital Kuala Lumpur, Kuala Lumpur, Malaysia.
  3. 3Department of Radiology, Universiti Kebangsaan Malaysia, Kuala Lumpur, Malaysia.
  4. 4Department of Public Health, Faculty of Allied Health Sciences, Daffodil International University, Bangladesh, Dhaka.

Abstract

Computed tomography pulmonary angiography (CTPA) is the standard imaging modality for diagnosing pulmonary thromboembolism (PTE). Lowering tube potential enhances iodine attenuation but may affect image noise. This retrospective study evaluated quantitative image quality, subjective ratings, and radiation dose across 120, 100, and 80 kV CTPA protocols on a dual-source CT scanner. Ninety adult patients (n=30 per protocol) undergoing clinically indicated CTPA were evaluated across three tube potential settings (120, 100, and 80 kV). Quantitative parameters pulmonary artery attenuation [Hounsfield unit (HU)], image noise [standard deviation (SD)], signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and size-specific dose estimates (SSDE)-were evaluated. Three cardiothoracic radiologists performed blinded qualitative assessments using a 5-point Likert scale. Radiation dose metrics [volumetric computed tomography dose index (CTDIvol), dose-length product (DLP), and estimated effective dose] were recorded. Baseline characteristics [age, sex, body mass index (BMI)] were comparable across groups (P>0.05). Pulmonary artery attenuation increased significantly as tube potential decreased, reaching highest values at 80 kV (633.9±92.1 vs. 338.5±42.1 HU at 120 kV, P<0.001). Image noise increased modestly at 80 kV (25.88±5.0 vs. 20.23±4.1 HU, P=0.26). SNR and CNR were significantly higher at 80 kV (25.93±2.6 and 25.20±3.4) compared with 120 kV (18.19±4.8 and 9.31±2.1, P<0.001). Effective dose decreased significantly by 46.4% from 3.34±0.67 mSv (120 kV) to 1.79±0.35 mSv (80 kV) (P<0.001). Qualitative scores for overall image quality, vascular enhancement, and motion artifacts showed no significant differences across protocols (P>0.05). Lowering tube potential to 80 kV in CTPA significantly increases vascular attenuation, SNR, and CNR while offering substantial radiation dose savings and maintaining subjective image quality scores. These findings support the feasibility of an 80-kV protocol for radiation dose optimization in CTPA.

Related

  1. Radiation Dose Optimization in Coronary CT Angiography
    Muhammad Kashif, Danish Waris Ali, Adeel Ur Rahman Siddiqui +4
    JACC Case ReportsPeer-reviewedCC BY-NC-ND 4.0Full textMedical imaging & radiologyPMC
  2. Kartagener syndrome with pulmonary mycetoma and recurrent thromboembolism : A case report
    Muhammad Asghar Khan, Okasha Tahir, Abbas Khan +8
    MedicinePeer-reviewedCC BYFull textGeneral medicineEurope PMC
  3. Impact of Optical Coherence Tomography Angiography Scan Protocols on Quantitative Microvascular Biomarkers in Diabetic Retinopathy.
    Wenjun Song, Ryan Zi Lin Tan, Ayush Jayakrishnan +19
    Translational vision science & technologyPeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMC
  4. Incidental pulmonary nodules detected on computed tomography pulmonary angiography: prevalence and malignancy outcomes from a Thai tertiary-care center
    Korkate Taikerd, Thitiwat Sriprasart
    Journal of thoracic diseasePeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMC
  5. Pulmonary thromboembolism with prolonged pulmonary infarction during intensive decongestion for severe heart failure : A case report
    A-Reum Roh, Jong Hyun Baek, Jong-Il Park
    MedicinePeer-reviewedCC BYFull textGeneral medicineEurope PMCPMC
BibTeX
@article{shazli2026image,
  title = {{Image quality and radiation dose optimization in CT pulmonary angiography using 120, 100, and 80 kV protocols for pulmonary thromboembolism diagnosis.}},
  author = {Zubir Shazli and Hanani Abdul Manan and Syeda Humayra and Zaidah Maspin and Kam Wai Chan and Victor Xing Dao Chong and Sarawana Chelwan Muniandy and Sian Zhen Khoo and Akmal Sabarudin},
  journal = {Quantitative imaging in medicine and surgery},
  year = {2026},
  month = sep,
  doi = {10.21037/qims-2026-1209},
  url = {https://europepmc.org/article/MED/42997798},
  note = {CC BY-NC-ND},
}
RIS
TY  - JOUR
TI  - Image quality and radiation dose optimization in CT pulmonary angiography using 120, 100, and 80 kV protocols for pulmonary thromboembolism diagnosis.
AU  - Zubir Shazli
AU  - Hanani Abdul Manan
AU  - Syeda Humayra
AU  - Zaidah Maspin
AU  - Kam Wai Chan
AU  - Victor Xing Dao Chong
AU  - Sarawana Chelwan Muniandy
AU  - Sian Zhen Khoo
AU  - Akmal Sabarudin
PY  - 2026
DA  - 2026/09/30
JO  - Quantitative imaging in medicine and surgery
DO  - 10.21037/qims-2026-1209
UR  - https://europepmc.org/article/MED/42997798
AB  - Computed tomography pulmonary angiography (CTPA) is the standard imaging modality for diagnosing pulmonary thromboembolism (PTE). Lowering tube potential enhances iodine attenuation but may affect image noise. This retrospective study evaluated quantitative image quality, subjective ratings, and radiation dose across 120, 100, and 80 kV CTPA protocols on a dual-source CT scanner. Ninety adult patients (n=30 per protocol) undergoing clinically indicated CTPA were evaluated across three tube potential settings (120, 100, and 80 kV). Quantitative parameters pulmonary artery attenuation [Hounsfield unit (HU)], image noise [standard deviation (SD)], signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and size-specific dose estimates (SSDE)-were evaluated. Three cardiothoracic radiologists performed blinded qualitative assessments using a 5-point Likert scale. Radiation dose metrics [volumetric computed tomography dose index (CTDIvol), dose-length product (DLP), and estimated effective dose] were recorded. Baseline characteristics [age, sex, body mass index (BMI)] were comparable across groups (P>0.05). Pulmonary artery attenuation increased significantly as tube potential decreased, reaching highest values at 80 kV (633.9±92.1 vs. 338.5±42.1 HU at 120 kV, P<0.001). Image noise increased modestly at 80 kV (25.88±5.0 vs. 20.23±4.1 HU, P=0.26). SNR and CNR were significantly higher at 80 kV (25.93±2.6 and 25.20±3.4) compared with 120 kV (18.19±4.8 and 9.31±2.1, P<0.001). Effective dose decreased significantly by 46.4% from 3.34±0.67 mSv (120 kV) to 1.79±0.35 mSv (80 kV) (P<0.001). Qualitative scores for overall image quality, vascular enhancement, and motion artifacts showed no significant differences across protocols (P>0.05). Lowering tube potential to 80 kV in CTPA significantly increases vascular attenuation, SNR, and CNR while offering substantial radiation dose savings and maintaining subjective image quality scores. These findings support the feasibility of an 80-kV protocol for radiation dose optimization in CTPA.
SN  - 2223-4292
SN  - 2223-4306
ER  -