A Multicenter Prospective Feasibility Study of a Novel Rapid-Exchange Biliary Stent Delivery System Enabling Sequential Stent Placement with Index Guidewire Retention
6 affiliations
- 1Center for Interventional Gastroenterology at UTHealth Houston The University of Texas Health Science Center at Houston Houston Texas United States
- 2Division of Gastroenterology and Hepatology Johns Hopkins Medicine Baltimore Maryland United States
- 3Division of Gastroenterology HCA Florida Largo Hospital Largo Florida United States
- 4Gastroenterology Ziv Medical Center Safed Israel
- 5Texas Digestive Disease Consultants PLLC Southlake Texas United States
- 6Gastroenterology Montefiore Health System, Albert Einstein College of Medicine New York New York United States
Abstract
Background and Aims Biliary stenting is essential for the management of biliary obstruction. All plastic biliary stent delivery systems in the United States require withdrawal of the guidewire upon stent deployment, mandating repeat cannulation when additional stents are required. A novel multiple biliary stent delivery system with a self-detaching tip preserves the index guidewire in situ after stent release, enabling rapid sequential placement of additional plastic stents. We report the first-in-human multicenter clinical experience with this device.
Methods This was a prospective feasibility pilot study at 7 international tertiary referral centers. Endoscopists were trained using a 2-minute video and bedside proctoring. Adult patients undergoing endoscopic retrograde cholangiopancreatography (ERCP) with the novel system between 2023 and 2024 were enrolled. Primary endpoints were technical success of stent deployment and index guidewire retention. Secondary endpoints included stents placed per session, ability to advance the system to the intended location, immediate adverse events (AE), and endoscopist-reported usability.
Results Twenty-five patients underwent ERCP with the novel system. Between 1 and 4 plastic stents were placed in parallel per session. Technical success was achieved in 24 of 25 cases (96.0%); one stent could not be delivered owing to severe ampullary angulation. The index guidewire was retained in situ in 100% of successfully delivered stents.
Conclusion In this first-in-human multicenter experience, a novel rapid-exchange biliary stent delivery system with a self-detaching tip enabled sequential parallel plastic stent placement without removal of the index guidewire, with high technical success and no immediate procedure-related adverse events. Further prospective comparative study is warranted.
Related
- Prophylactic endoscopic transpapillary gallbladder stenting at fully covered self-expandable metal stent placement to prevent cholecystitis in malignant distal biliary obstruction: A multicenter prospective single-arm study protocol (PROTECT study)
- Prospective multicenter feasibility study of omnipolar near-field peak-frequency mapping for slow-pathway localization in typical atrioventricular nodal re-entrant tachycardia.
- Comparison of feasibility, inter-instrument agreement, and prognostic value of four frailty assessment instruments in geriatric colorectal cancer patients: a 3-year multicenter prospective cohort study
- Pan-European Training Program for Robotic Pancreatoduodenectomy (LEARNBOT): Prospective Multicenter Study on Feasibility and Outcome.
- Feasibility and placement accuracy of a sensor-guided intraosseous device with an automated stop in a small-bone rat cadaver model: a single-arm pilot study
BibTeX
@article{ma2026multicenter,
title = {{A Multicenter Prospective Feasibility Study of a Novel Rapid-Exchange Biliary Stent Delivery System Enabling Sequential Stent Placement with Index Guidewire Retention}},
author = {Jennifer Ma and Madhav Desai and Vikesh K. Singh and Meir Mizrahi and Stanislav Bezobchuk and Dang Nguyen and Mairin Joseph-Talreja and Isaac Raijman and Sammy Ho and Mouen A. Khashab and Nirav Thosani},
journal = {Endoscopy International Open},
year = {2026},
month = oct,
doi = {10.1055/a-2942-4784},
url = {https://doi.org/10.1055/a-2942-4784},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - A Multicenter Prospective Feasibility Study of a Novel Rapid-Exchange Biliary Stent Delivery System Enabling Sequential Stent Placement with Index Guidewire Retention
AU - Jennifer Ma
AU - Madhav Desai
AU - Vikesh K. Singh
AU - Meir Mizrahi
AU - Stanislav Bezobchuk
AU - Dang Nguyen
AU - Mairin Joseph-Talreja
AU - Isaac Raijman
AU - Sammy Ho
AU - Mouen A. Khashab
AU - Nirav Thosani
PY - 2026
DA - 2026/10/21
JO - Endoscopy International Open
DO - 10.1055/a-2942-4784
UR - https://doi.org/10.1055/a-2942-4784
AB - Background and Aims Biliary stenting is essential for the management of biliary obstruction. All plastic biliary stent delivery systems in the United States require withdrawal of the guidewire upon stent deployment, mandating repeat cannulation when additional stents are required. A novel multiple biliary stent delivery system with a self-detaching tip preserves the index guidewire in situ after stent release, enabling rapid sequential placement of additional plastic stents. We report the first-in-human multicenter clinical experience with this device. Methods This was a prospective feasibility pilot study at 7 international tertiary referral centers. Endoscopists were trained using a 2-minute video and bedside proctoring. Adult patients undergoing endoscopic retrograde cholangiopancreatography (ERCP) with the novel system between 2023 and 2024 were enrolled. Primary endpoints were technical success of stent deployment and index guidewire retention. Secondary endpoints included stents placed per session, ability to advance the system to the intended location, immediate adverse events (AE), and endoscopist-reported usability. Results Twenty-five patients underwent ERCP with the novel system. Between 1 and 4 plastic stents were placed in parallel per session. Technical success was achieved in 24 of 25 cases (96.0%); one stent could not be delivered owing to severe ampullary angulation. The index guidewire was retained in situ in 100% of successfully delivered stents. Conclusion In this first-in-human multicenter experience, a novel rapid-exchange biliary stent delivery system with a self-detaching tip enabled sequential parallel plastic stent placement without removal of the index guidewire, with high technical success and no immediate procedure-related adverse events. Further prospective comparative study is warranted.
SN - 2364-3722
SN - 2196-9736
ER -