Physics-informed depth-enhanced surface tracking for cumulative dose reconstruction in breast radiotherapy: An in silico feasibility study.
- 1Department of Medical Engineering, The Fourth Affiliated Hospital of Guangzhou Medical University, Guangzhou, Guangdong, China.
- 2Department of Radiation Oncology, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou, Guangdong, China.
Abstract
Purpose
Surface guided radiotherapy (SGRT) provides nonionizing motion monitoring but lacks direct correlation with internal dosimetric changes. Purely data-driven models for surface-to-internal deformation prediction often yield biomechanically implausible results. This study proposes a physics-informed depth-enhanced (PIDE) framework, validated in silico, to reconstruct internal deformation and cumulative dose from noisy surface measurements.Methods
A virtual clinical trial (VCT) was conducted using weekly cone-beam CT (CBCT) data from 29 breast cancer patients. We simulated consumer-grade RGB-D sensor noise ( σ∝z2 ) on extracted patient surfaces. To map external surface features to internal deformation vector fields (DVF), we developed a regularized ridge regression model. Unlike traditional models, ours integrates a spectral energy penalty derived from patient-specific CBCT motion priors, constraining predictions to physically plausible deformations. Patient-specific clinical RTDOSE distributions and 348 anatomical landmarks were utilized to evaluate the dosimetric and geometric accuracy. Vector field smoothness was quantified using the Jacobian determinant.Results
Under realistic simulated sensor noise, the PIDE framework achieved an anatomical target registration error (TRE) of 1.33 ± 1.42 mm (compared to an initial misalignment of 2.14 ± 2.38 mm, representing a 37.8% error reduction). The physics-constrained regression yielded an RMSE of 2.31 mm (r = 0.720) for motion prediction and effectively eliminated nonphysical tissue "folding" artifacts (reducing negative Jacobian voxels from 0.54% to 0.10%, p = 0.0209). Dosimetric evaluation using actual patient treatment plans achieved a mean Gamma passing rate (3%/3 mm) of 99.68 ± 0.30% (range: 99.00%-100.00%), with 100% of patients (29/29) exceeding the 90% clinical safety threshold. The end-to-end algorithmic inference latency was 2.71 ms per frame, demonstrating high computational efficiency.Conclusion
The integration of geometric surface features with patient-specific physical constraints successfully prevents nonphysiological deformation artifacts caused by sensor noise. This in silico study demonstrates the feasibility and computational efficiency of the PIDE framework for rapid cumulative dose monitoring.
Related
- Feasibility of Urethra-Sparing Ultra-Hypofractionated Three-Dose-Level Radiotherapy for Prostate Cancer with a Dominant Intraprostatic Lesion Abutting the High-Dose Avoidance Zone.
- Evaluating a Large Language Model for Radiotherapy Patient Scheduling under Realistic Clinical and Operational Constraints: a Feasibility Study
- Diagnostic CT-Based Palliative Radiotherapy Using kV Image Verification: A Prospective Feasibility Study.
- Dosimetric feasibility of lattice radiotherapy for bulky tumors: a paired VMAT vs. helical tomotherapy comparison - the STRYKE trial
- Feasibility of reduced PTV margins in magnetic resonance image-guided online adaptive radiotherapy for bladder cancer: a personalised approach.
BibTeX
@article{tang2026physicsinformed,
title = {{Physics-informed depth-enhanced surface tracking for cumulative dose reconstruction in breast radiotherapy: An in silico feasibility study.}},
author = {Xixue Tang and Minmin Qiu and Yongjin Deng},
journal = {Journal of applied clinical medical physics},
year = {2026},
month = oct,
doi = {10.1002/acm2.70890},
url = {https://europepmc.org/article/MED/42949306},
note = {CC BY},
}
RIS
TY - JOUR
TI - Physics-informed depth-enhanced surface tracking for cumulative dose reconstruction in breast radiotherapy: An in silico feasibility study.
AU - Xixue Tang
AU - Minmin Qiu
AU - Yongjin Deng
PY - 2026
DA - 2026/10/28
JO - Journal of applied clinical medical physics
DO - 10.1002/acm2.70890
UR - https://europepmc.org/article/MED/42949306
AB - Purpose Surface guided radiotherapy (SGRT) provides nonionizing motion monitoring but lacks direct correlation with internal dosimetric changes. Purely data-driven models for surface-to-internal deformation prediction often yield biomechanically implausible results. This study proposes a physics-informed depth-enhanced (PIDE) framework, validated in silico, to reconstruct internal deformation and cumulative dose from noisy surface measurements.Methods A virtual clinical trial (VCT) was conducted using weekly cone-beam CT (CBCT) data from 29 breast cancer patients. We simulated consumer-grade RGB-D sensor noise ( σ∝z2 ) on extracted patient surfaces. To map external surface features to internal deformation vector fields (DVF), we developed a regularized ridge regression model. Unlike traditional models, ours integrates a spectral energy penalty derived from patient-specific CBCT motion priors, constraining predictions to physically plausible deformations. Patient-specific clinical RTDOSE distributions and 348 anatomical landmarks were utilized to evaluate the dosimetric and geometric accuracy. Vector field smoothness was quantified using the Jacobian determinant.Results Under realistic simulated sensor noise, the PIDE framework achieved an anatomical target registration error (TRE) of 1.33 ± 1.42 mm (compared to an initial misalignment of 2.14 ± 2.38 mm, representing a 37.8% error reduction). The physics-constrained regression yielded an RMSE of 2.31 mm (r = 0.720) for motion prediction and effectively eliminated nonphysical tissue "folding" artifacts (reducing negative Jacobian voxels from 0.54% to 0.10%, p = 0.0209). Dosimetric evaluation using actual patient treatment plans achieved a mean Gamma passing rate (3%/3 mm) of 99.68 ± 0.30% (range: 99.00%-100.00%), with 100% of patients (29/29) exceeding the 90% clinical safety threshold. The end-to-end algorithmic inference latency was 2.71 ms per frame, demonstrating high computational efficiency.Conclusion The integration of geometric surface features with patient-specific physical constraints successfully prevents nonphysiological deformation artifacts caused by sensor noise. This in silico study demonstrates the feasibility and computational efficiency of the PIDE framework for rapid cumulative dose monitoring.
SN - 1526-9914
ER -