Automated verification of source dwell position in lunar ovoid applicators.
6 affiliations
- 1University of Barcelona, Barcelona, Spain.
- 2Computer Vision and Robotics Institute, University of Girona, Girona, Spain.
- 3Medical Physics Section, Radiation Oncology Department, Hospital Clínic Barcelona, Barcelona, Spain.
- 4Radiation Protection Department, Hospital Clínic Barcelona, Barcelona, Spain.
- 5Infrastructure Management, Hospital Clínic Barcelona, Barcelona, Spain.
- 6Laboratory of Translational Genomics and Targeted therapies in Solid Tumors, Institut d'Investigacions Biomèdiques August Pi i Sunyer, Barcelona, Spain.
Abstract
Background
Accurate reconstruction of the source dwell positions in high-dose-rate (HDR) brachytherapy is essential for treatment quality. The Venezia™ gynaecological applicator, with its lunar-shaped ovoids, poses challenges because conventional dummy markers do not match the actual source path, introducing systematic uncertainties.Purpose
The aim of this work was to develop and demonstrate an automated, reproducible, and open-source software for autoradiographic verification of dwell positions in Venezia lunar ovoids, and to compare measured positions with manufacturer-provided applicator library data. This pipeline established a standardized approach for clinical commissioning and quality assurance (QA) workflows.Methods
A custom 3D-printed holder was designed to standardize autoradiograph acquisition using radiochromic films. An open-source Python pipeline, referred to as SPICA (Source Position Identification in Circular Applicators), was developed to process scanned films, isolate dwell position, and compute their coordinates with associated uncertainties. The code allows direct comparison between measured positions and manufacturer-provided applicator library data. We performed preliminary validation of the software by measuring the trajectory of an Ir-192 source for the 22 and 26 mm ovoids with a step size of 2.5 mm.Results
Mean discrepancies between measured and reference positions were 1.2 ± 0.5 and 1.5 ± 0.4 mm for the 22 and 26 mm ovoids, respectively, with a maximum deviation of 2.6 mm. Distal positions showed better agreement. These findings confirm that, in the tested applicator-afterloader combination, vendor libraries provide a generally reliable baseline but may exhibit systematic differences.Conclusions
The proposed methodology offers a practical, reproducible solution to improve source path verification while reducing manual intervention. It supports more robust commissioning and QA of complex applicators and using the coordinates calculated in the treatment planning software can potentially reduce geometric uncertainties in clinical cases. Individualized commissioning remains essential for applicators with complex geometries.
Related
- Efficient multi-basis quantum position verification secure against generalized adversaries
- Arbitrarily Loss-Tolerant Quantum Position Verification in a Single Execution
- Classical Verifier Position Verification from Non-Local Games
- Loss-Tolerant Quantum Position Verification for Metropolitan Area Networks
- Breaking the Bounded Entanglement Barrier for Quantum Position Verification
BibTeX
@article{gubaubesalu2026automated,
title = {{Automated verification of source dwell position in lunar ovoid applicators.}},
author = {Miquel Gubau-Besalú and Adrià Casamitjana and Cristian Candela-Juan and Eduard Agusti-Camprubi and Gislè Segura-Roca and Gerard Trias and Antonio Herreros},
journal = {Journal of applied clinical medical physics},
year = {2026},
month = oct,
doi = {10.1002/acm2.70887},
url = {https://europepmc.org/article/MED/42963831},
note = {CC BY},
}
RIS
TY - JOUR
TI - Automated verification of source dwell position in lunar ovoid applicators.
AU - Miquel Gubau-Besalú
AU - Adrià Casamitjana
AU - Cristian Candela-Juan
AU - Eduard Agusti-Camprubi
AU - Gislè Segura-Roca
AU - Gerard Trias
AU - Antonio Herreros
PY - 2026
DA - 2026/10/30
JO - Journal of applied clinical medical physics
DO - 10.1002/acm2.70887
UR - https://europepmc.org/article/MED/42963831
AB - Background Accurate reconstruction of the source dwell positions in high-dose-rate (HDR) brachytherapy is essential for treatment quality. The Venezia™ gynaecological applicator, with its lunar-shaped ovoids, poses challenges because conventional dummy markers do not match the actual source path, introducing systematic uncertainties.Purpose The aim of this work was to develop and demonstrate an automated, reproducible, and open-source software for autoradiographic verification of dwell positions in Venezia lunar ovoids, and to compare measured positions with manufacturer-provided applicator library data. This pipeline established a standardized approach for clinical commissioning and quality assurance (QA) workflows.Methods A custom 3D-printed holder was designed to standardize autoradiograph acquisition using radiochromic films. An open-source Python pipeline, referred to as SPICA (Source Position Identification in Circular Applicators), was developed to process scanned films, isolate dwell position, and compute their coordinates with associated uncertainties. The code allows direct comparison between measured positions and manufacturer-provided applicator library data. We performed preliminary validation of the software by measuring the trajectory of an Ir-192 source for the 22 and 26 mm ovoids with a step size of 2.5 mm.Results Mean discrepancies between measured and reference positions were 1.2 ± 0.5 and 1.5 ± 0.4 mm for the 22 and 26 mm ovoids, respectively, with a maximum deviation of 2.6 mm. Distal positions showed better agreement. These findings confirm that, in the tested applicator-afterloader combination, vendor libraries provide a generally reliable baseline but may exhibit systematic differences.Conclusions The proposed methodology offers a practical, reproducible solution to improve source path verification while reducing manual intervention. It supports more robust commissioning and QA of complex applicators and using the coordinates calculated in the treatment planning software can potentially reduce geometric uncertainties in clinical cases. Individualized commissioning remains essential for applicators with complex geometries.
SN - 1526-9914
ER -