Balloon shaft fracture, deflation failure, and intracoronary entrapment: a stepwise percutaneous bailout
- 1Department of Interventional Cardiology, Nhan dan Gia Dinh Hospital, 1 No Trang Long Street, Gia Dinh Ward, Ho Chi Minh City 700000, Vietnam.
Abstract
Background
Balloon catheter fracture during percutaneous coronary intervention (PCI) is rare but potentially catastrophic. The literature describes three distinct forms: shaft fracture, undeflatable balloon, and entrapped balloon. Published bailout strategies typically address each problem in isolation; however, no specific guidance exists when multiple forms occur simultaneously.Case summary
A 62-year-old man with multivessel coronary disease underwent staged PCI of a left circumflex-obtuse marginal bifurcation. During lesion preparation, a reused semi-compliant balloon (2.5 × 14 mm) fractured after multiple inflations, leaving an inflated, entrapped balloon fragment at the bifurcation with impaired antegrade flow. A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the tip of the guiding catheter. The procedure was completed successfully with stenting, and the patient recovered uneventfully.Discussion
Sequential management converted a high-risk situation into a tractable single problem. Distal displacement of the inflated fragment relieved ischaemia before any retrieval attempt was made. Trapping with a second balloon straddling the guide catheter tip may improve grip when the fragment partly protrudes from the guiding catheter.
Related
- Assessing Calcified Coronary Lesions to Guide Percutaneous Coronary Intervention: From Angiography to Intracoronary Imaging
- Single-Stage Computed Tomography Angiography, Percutaneous Coronary Intervention, and Transcatheter Aortic Valve Replacement in an Octogenarian With Femoral Neck Fracture, Non-ST-Elevation Myocardial Infarction, and Severe Aortic Stenosis: A Case Report and Literature Review of Multidisciplinary Approaches
- Functional outcome of closed reduction and percutaneous multiple K-wire fixation of displaced intra-articular calcaneal fracture
- Development and external validation of a nomogram for predicting re-fracture after percutaneous kyphoplasty: A two-center study
- Percutaneous Vertebroplasty versus Non-Surgical Controls for Osteoporotic Vertebral Compression Fractures: A Systematic Review and Meta-Analysis of Pain, Fracture, and Safety Outcomes
BibTeX
@article{nguyen2026balloon,
title = {{Balloon shaft fracture, deflation failure, and intracoronary entrapment: a stepwise percutaneous bailout}},
author = {Anh Do Nguyen and Tien Hoa Phan and Son Truong Do Chau},
journal = {European heart journal. Case reports},
year = {2026},
month = oct,
doi = {10.1093/ehjcr/ytag804},
url = {https://europepmc.org/article/PMC/PMC13795251},
note = {CC BY-NC},
}
RIS
TY - JOUR
TI - Balloon shaft fracture, deflation failure, and intracoronary entrapment: a stepwise percutaneous bailout
AU - Anh Do Nguyen
AU - Tien Hoa Phan
AU - Son Truong Do Chau
PY - 2026
DA - 2026/10/27
JO - European heart journal. Case reports
DO - 10.1093/ehjcr/ytag804
UR - https://europepmc.org/article/PMC/PMC13795251
AB - Background Balloon catheter fracture during percutaneous coronary intervention (PCI) is rare but potentially catastrophic. The literature describes three distinct forms: shaft fracture, undeflatable balloon, and entrapped balloon. Published bailout strategies typically address each problem in isolation; however, no specific guidance exists when multiple forms occur simultaneously.Case summary A 62-year-old man with multivessel coronary disease underwent staged PCI of a left circumflex-obtuse marginal bifurcation. During lesion preparation, a reused semi-compliant balloon (2.5 × 14 mm) fractured after multiple inflations, leaving an inflated, entrapped balloon fragment at the bifurcation with impaired antegrade flow. A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the tip of the guiding catheter. The procedure was completed successfully with stenting, and the patient recovered uneventfully.Discussion Sequential management converted a high-risk situation into a tractable single problem. Distal displacement of the inflated fragment relieved ischaemia before any retrieval attempt was made. Trapping with a second balloon straddling the guide catheter tip may improve grip when the fragment partly protrudes from the guiding catheter.
SN - 2514-2119
ER -