Fast Reconstruction of Motion‐Corrupted Data With Mobile‐GRAPPA: Motion and δB0 Inhomogeneity Correction Leveraging Efficient GRAPPA
- 1Department of Electrical Engineering Stanford University Stanford California USA
- 2Department of Radiology Stanford University Stanford California USA
- 3Siemens Medical Solutions USA Inc. Malvern Pennsylvania USA
Abstract
Purpose: To develop an accurate and computationally efficient motion‐corrected MRI reconstruction framework that incorporates hundreds to thousands of motion and δB0 estimates from high‐temporal‐resolution tracking.
Methods: We propose Mobile‐GRAPPA, a k‐space preprocessing approach that uses MLP‐parameterized local GRAPPA operators to jointly correct trajectory perturbations, coil reweighting, and δB0‐induced phase changes before standard downstream reconstruction. Reconstruction accuracy, noise propagation, spatial resolution, and runtime were evaluated using 3D MPRAGE, multi‐echo 3D GRE, and 3D EPTI.
Results: Experiments with discrete motion states demonstrated that Mobile‐GRAPPA followed by standard SENSE achieved image quality comparable to Aligned‐SENSE. In 3D GRE with 1620 tracked states and 3D EPTI with 544 tracked states, Mobile‐GRAPPA incorporated all state estimates with minimal motion‐correction overhead. Total reconstruction times were approximately 15 s for GRE and 20 min for EPTI, whereas full‐state Aligned‐SENSE was computationally prohibitive (approximately 10 h for GRE and multiple days for EPTI). Pseudo‐replica and PSF analyses showed limited additional noise amplification and negligible spatial‐resolution loss.
Conclusion: Mobile‐GRAPPA enables dense motion and δB0 information to be incorporated with minimal motion‐correction overhead while preserving standard SENSE, subspace, and other downstream reconstruction pipelines.
Related
- Iterative Navigator-Based Retrospective Motion Correction for Multi-Shot Brain MR Elastography and Evaluation of Single-Shot Post-Reconstruction Correction
- Jacobian-Free Unrolling: memory efficient unrolled network for 3D SPECT Reconstruction
- Staged reconstruction of a neglected distal femoral pseudoarthrosis in a skeletally immature child using Ilizarov-based deformity correction and limb lengthening: a case report
- Ventilation inhomogeneity in Duchenne muscular dystrophy before the initiation of ventilatory support: Insights from electrical impedance tomography
- Leveraging ecological momentary assessment and dynamic structural equation modeling to identify idiographic antecedents of child irritability and defiance
BibTeX
@article{lin2026fast,
title = {{Fast Reconstruction of Motion‐Corrupted Data With Mobile‐GRAPPA: Motion and δB0 Inhomogeneity Correction Leveraging Efficient GRAPPA}},
author = {Yimeng Lin and Nan Wang and Daniel Abraham and Daniel Polak and Xiaozhi Cao and Aizada Nurdinova and Stephen Cauley and Kawin Setsompop},
journal = {Magnetic Resonance in Medicine},
year = {2026},
month = sep,
doi = {10.1002/mrm.70575},
url = {https://doi.org/10.1002/mrm.70575},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Fast Reconstruction of Motion‐Corrupted Data With Mobile‐GRAPPA: Motion and δB0 Inhomogeneity Correction Leveraging Efficient GRAPPA
AU - Yimeng Lin
AU - Nan Wang
AU - Daniel Abraham
AU - Daniel Polak
AU - Xiaozhi Cao
AU - Aizada Nurdinova
AU - Stephen Cauley
AU - Kawin Setsompop
PY - 2026
DA - 2026/09/04
JO - Magnetic Resonance in Medicine
DO - 10.1002/mrm.70575
UR - https://doi.org/10.1002/mrm.70575
AB - Purpose: To develop an accurate and computationally efficient motion‐corrected MRI reconstruction framework that incorporates hundreds to thousands of motion and δB0 estimates from high‐temporal‐resolution tracking. Methods: We propose Mobile‐GRAPPA, a k‐space preprocessing approach that uses MLP‐parameterized local GRAPPA operators to jointly correct trajectory perturbations, coil reweighting, and δB0‐induced phase changes before standard downstream reconstruction. Reconstruction accuracy, noise propagation, spatial resolution, and runtime were evaluated using 3D MPRAGE, multi‐echo 3D GRE, and 3D EPTI. Results: Experiments with discrete motion states demonstrated that Mobile‐GRAPPA followed by standard SENSE achieved image quality comparable to Aligned‐SENSE. In 3D GRE with 1620 tracked states and 3D EPTI with 544 tracked states, Mobile‐GRAPPA incorporated all state estimates with minimal motion‐correction overhead. Total reconstruction times were approximately 15 s for GRE and 20 min for EPTI, whereas full‐state Aligned‐SENSE was computationally prohibitive (approximately 10 h for GRE and multiple days for EPTI). Pseudo‐replica and PSF analyses showed limited additional noise amplification and negligible spatial‐resolution loss. Conclusion: Mobile‐GRAPPA enables dense motion and δB0 information to be incorporated with minimal motion‐correction overhead while preserving standard SENSE, subspace, and other downstream reconstruction pipelines.
SN - 0740-3194
SN - 1522-2594
ER -