Iterative Navigator-Based Retrospective Motion Correction for Multi-Shot Brain MR Elastography and Evaluation of Single-Shot Post-Reconstruction Correction
- 1Department of Biomedical Engineering, Fu Foundation School of Engineering and Applied Science, Columbia University, New York, New York, USA
- 2Department of Neurology, Massachusetts General Hospital, Boston, Massachusetts, USA
- 3Department of Radiology, Vagelos College of Physicians and Surgeons, Columbia University, New York, New York, USA
Abstract
Purpose:: Brain MR elastography (MRE) is uniquely sensitive to subject motion, as even small phase deviations can produce large errors in the mechanical properties estimated. Existing motion correction strategies have not been validated for MRE and are often incompatible with complex multi-shot or joint temporal reconstructions.
Methods:: We propose COMET (Correction Of Motion in Elastography via Transformation), a framework that estimates rigid-body motion from navigator images and embeds transformation matrices into an iterative reconstruction and displacement-vector reorientation for multi-shot MRE. We additionally validate post-reconstruction correction and displacement-vector reorientation for single-shot MRE. The framework was evaluated on MRE from 9 pairs of EPI single-shot and 9 pairs of spiral multi-shot acquisitions, each containing a with-motion scan and a reference. Additionally, four representative motion patterns with increasing amplitude were simulated to indicate how different motions affect MRE quality and results, and a brain tumor MRE dataset with simulated motion was used to assess clinical effect.
Results:: Motion correction improved stiffness accuracy across all tested datasets, reducing average stiffness errors by up to 240.86% in multi-shot spiral MRE, and by up to 44.28% in single-shot EPI, across in vivo motion ranging from 4.81 to 67.58 mm translation and 1.85° – 23.63° rotation. In a pediatric glioma dataset, simulated motion completely obscured tumor stiffness contrast, while correction restored tumor stiffness visibility. Robustness of estimated stiffness maps depended on motion amplitude, temporal motion patterns, and encoding strategies.
Conclusion:: This approach enables robust motion correction for single- and multi-shot brain MRE, and expands reliable MRE to clinical and challenging populations.
Related
- Effectiveness of Contemporary Invisalign® Clear Aligners for Derotation and Deep Bite Correction of Different Degrees of Severity in Adolescents and Adults: A Retrospective Clinical Study
- Fast Reconstruction of Motion‐Corrupted Data With Mobile‐GRAPPA: Motion and δB0 Inhomogeneity Correction Leveraging Efficient GRAPPA
- Staged reconstruction of a neglected distal femoral pseudoarthrosis in a skeletally immature child using Ilizarov-based deformity correction and limb lengthening: a case report
- Aberration correction in fetal ultrasound: Clinical evaluation on image quality improvement
- Clinical decision support tool evaluation: Correction of direct oral anticoagulant dosing errors by clinical pharmacists and physicians
BibTeX
@article{shi2026iterative,
title = {{Iterative Navigator-Based Retrospective Motion Correction for Multi-Shot Brain MR Elastography and Evaluation of Single-Shot Post-Reconstruction Correction}},
author = {Zhuoyu Shi and Alex Michael Cerjanic and Grace McIlvain},
journal = {Magnetic resonance in medicine},
year = {2026},
month = oct,
doi = {10.1002/mrm.70625},
url = {https://doi.org/10.1002/mrm.70625},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Iterative Navigator-Based Retrospective Motion Correction for Multi-Shot Brain MR Elastography and Evaluation of Single-Shot Post-Reconstruction Correction
AU - Zhuoyu Shi
AU - Alex Michael Cerjanic
AU - Grace McIlvain
PY - 2026
DA - 2026/10/09
JO - Magnetic resonance in medicine
DO - 10.1002/mrm.70625
UR - https://doi.org/10.1002/mrm.70625
AB - Purpose:: Brain MR elastography (MRE) is uniquely sensitive to subject motion, as even small phase deviations can produce large errors in the mechanical properties estimated. Existing motion correction strategies have not been validated for MRE and are often incompatible with complex multi-shot or joint temporal reconstructions. Methods:: We propose COMET (Correction Of Motion in Elastography via Transformation), a framework that estimates rigid-body motion from navigator images and embeds transformation matrices into an iterative reconstruction and displacement-vector reorientation for multi-shot MRE. We additionally validate post-reconstruction correction and displacement-vector reorientation for single-shot MRE. The framework was evaluated on MRE from 9 pairs of EPI single-shot and 9 pairs of spiral multi-shot acquisitions, each containing a with-motion scan and a reference. Additionally, four representative motion patterns with increasing amplitude were simulated to indicate how different motions affect MRE quality and results, and a brain tumor MRE dataset with simulated motion was used to assess clinical effect. Results:: Motion correction improved stiffness accuracy across all tested datasets, reducing average stiffness errors by up to 240.86% in multi-shot spiral MRE, and by up to 44.28% in single-shot EPI, across in vivo motion ranging from 4.81 to 67.58 mm translation and 1.85° – 23.63° rotation. In a pediatric glioma dataset, simulated motion completely obscured tumor stiffness contrast, while correction restored tumor stiffness visibility. Robustness of estimated stiffness maps depended on motion amplitude, temporal motion patterns, and encoding strategies. Conclusion:: This approach enables robust motion correction for single- and multi-shot brain MRE, and expands reliable MRE to clinical and challenging populations.
SN - 0740-3194
SN - 1522-2594
ER -