Ventilator-circuit configurations for noninvasive ventilation in infants with acute respiratory failure: a randomized crossover trial

4 affiliations
  1. 1Pediatric Intensive Care Unit, Fondazione IRCCS Cà Granda Ospedale Maggiore Policlinico, Milan, Italy
  2. 2Department of Anesthesiology and Intensive Care Medicine, Catholic University of the Sacred Heart, Rome, Italy
  3. 3Department of Anesthesia and Intensive Care Medicine, Niguarda Cà Granda, Milan, Italy
  4. 4Unicamillus International Medical School, Rome, Italy

Abstract

Background: Patient–ventilator synchrony may differ across ventilator and circuit configurations during noninvasive ventilation (NIV) in infants. We compared an intentional-leak single-limb turbine configuration (SLC-turbine), a dual-limb turbine configuration with an active expiratory valve (DLC-turbine), and a dual-limb compressed-gas pediatric ICU configuration (DLC-PICU) in infants receiving NIV for acute respiratory failure

Methods: In this single-center randomized crossover physiological study, infants underwent three 30-minute NIV phases in randomized order with unchanged pressure support, PEEP, and FiO₂. Esophageal pressure was used to assess the asynchrony index (AI), individual asynchronies, and inspiratory effort. Primary and secondary endpoints were AI differences between DLC-PICU and SLC-turbine and between DLC-PICU and DLC-turbine, respectively

Results: Twenty infants entered stabilization, 13 were randomized, and 12 completed the protocol (age 10 [7–21] months; SpO₂/FiO₂ 185 [160–220]). AI differed across configurations: 8.6% [6.7–16.5] with SLC-turbine, 25.4% [16.3–35.0] with DLC-turbine, and 45.7% [30.5–63.8] with DLC-PICU (p=0.0023). AI was higher with DLC-PICU than with SLC-turbine (Hodges–Lehmann difference +32.0%, 95% CI +13.9 to +50.1; p=0.0015) and DLC-turbine (+18.4%, 95% CI +5.9 to +36.3; p=0.021). Auto-triggering showed the clearest configuration-dependent difference and was more frequent with DLC-PICU than with SLC-turbine (7.5 [0.6–11.4] vs 0 [0–2.5] events/min; difference +5.30 events/min; p=0.0039), without a significant difference between DLC-PICU and DLC-turbine. Ineffective efforts, double triggering, and premature cycling did not differ across configurations. All NIV configurations reduced inspiratory effort versus unassisted spontaneous breathing, with no significant differences among NIV configurations

Conclusions: AI was lower with SLC-turbine and DLC-turbine than with DLC-PICU and lowest with SLC-turbine. Auto-triggering showed the clearest configuration-dependent difference. All configurations reduced inspiratory effort versus unassisted spontaneous breathing, without significant between-configuration differences. Because ventilator platform, circuit and expiratory architecture, and device-specific control functions differed, effects of individual components cannot be determined. These findings are specific to the ventilators and configurations evaluated. Further studies should determine whether these physiological differences translate into improved NIV tolerance or short- and long-term clinical outcomes

Trial registration: ClinicalTrials.gov, NCT04017780. Registered on 12 July 2019

Related

  1. Monitoring of regional ventilation and ventilator-induced lung injury in pediatric acute respiratory distress syndrome (MoVe lupARDS): a study protocol for a randomized controlled trial
    Andreas Hansmann, Johanna Moersdorf, Tamene Melaku +10
    TrialsPeer-reviewedCC BY 4.0Full textClinical trials & evidence synthesisCrossrefEurope PMC
  2. Global Protective Ventilator Settings Are Not Regional Protection: A Compartment-Resolved Framework for Personalizing Mechanical Ventilation in Acute Respiratory Distress Syndrome
    Journal of clinical medicinePeer-reviewedCC BYFull textGeneral medicineEurope PMC
  3. Airway occlusions to measure inspiratory effort, respiratory drive, and lung mechanics during noninvasive ventilation
    Francesco Murgolo, Domenico Luca Grieco, Rossana Soloperto +16
    American Journal of Respiratory and Critical Care MedicinePeer-reviewedCC BY 4.0Full text1 citedGeneral medicinePMC
  4. Comparative analysis of comfort of noninvasive ventilation (NIV) with a conventional oronasal mask and a bridge-free oronasal mask: A randomized controlled trial.
    Bhavani Prasad Guravana, Manoj Kamal, Pradeep K Bhatia +3
    Journal of anaesthesiology, clinical pharmacologyPeer-reviewedCC BY-NC-SAFull textSurgery & anaesthesiaEurope PMCPMC
  5. Semi‐recumbent position for the prevention of ventilator‐associated pneumonia in adults requiring mechanical ventilation
    Wenjun Jiang, Yamin Chen, Xiao Li +6
    The Cochrane Database of Systematic ReviewsPeer-reviewedPublisher termsFull textInfectious diseasePMC
BibTeX
@article{chidini2026ventilatorcircuit,
  title = {{Ventilator-circuit configurations for noninvasive ventilation in infants with acute respiratory failure: a randomized crossover trial}},
  author = {Giovanna Chidini and Domenico Luca Grieco and Thomas Langer and Cristina Villa and Stefano Scalia Catenacci and Tiziana Marchesi and Maria Francesca Patria and Giacomo Grasselli and Edoardo Calderini and Cesare Gregoretti},
  journal = {Critical Care},
  year = {2026},
  month = nov,
  doi = {10.1186/s13054-026-06447-2},
  url = {https://doi.org/10.1186/s13054-026-06447-2},
  note = {CC BY-NC-ND 4.0},
}
RIS
TY  - JOUR
TI  - Ventilator-circuit configurations for noninvasive ventilation in infants with acute respiratory failure: a randomized crossover trial
AU  - Giovanna Chidini
AU  - Domenico Luca Grieco
AU  - Thomas Langer
AU  - Cristina Villa
AU  - Stefano Scalia Catenacci
AU  - Tiziana Marchesi
AU  - Maria Francesca Patria
AU  - Giacomo Grasselli
AU  - Edoardo Calderini
AU  - Cesare Gregoretti
PY  - 2026
DA  - 2026/11/10
JO  - Critical Care
DO  - 10.1186/s13054-026-06447-2
UR  - https://doi.org/10.1186/s13054-026-06447-2
AB  - Background: Patient–ventilator synchrony may differ across ventilator and circuit configurations during noninvasive ventilation (NIV) in infants. We compared an intentional-leak single-limb turbine configuration (SLC-turbine), a dual-limb turbine configuration with an active expiratory valve (DLC-turbine), and a dual-limb compressed-gas pediatric ICU configuration (DLC-PICU) in infants receiving NIV for acute respiratory failure  Methods: In this single-center randomized crossover physiological study, infants underwent three 30-minute NIV phases in randomized order with unchanged pressure support, PEEP, and FiO₂. Esophageal pressure was used to assess the asynchrony index (AI), individual asynchronies, and inspiratory effort. Primary and secondary endpoints were AI differences between DLC-PICU and SLC-turbine and between DLC-PICU and DLC-turbine, respectively  Results: Twenty infants entered stabilization, 13 were randomized, and 12 completed the protocol (age 10 [7–21] months; SpO₂/FiO₂ 185 [160–220]). AI differed across configurations: 8.6% [6.7–16.5] with SLC-turbine, 25.4% [16.3–35.0] with DLC-turbine, and 45.7% [30.5–63.8] with DLC-PICU (p=0.0023). AI was higher with DLC-PICU than with SLC-turbine (Hodges–Lehmann difference +32.0%, 95% CI +13.9 to +50.1; p=0.0015) and DLC-turbine (+18.4%, 95% CI +5.9 to +36.3; p=0.021). Auto-triggering showed the clearest configuration-dependent difference and was more frequent with DLC-PICU than with SLC-turbine (7.5 [0.6–11.4] vs 0 [0–2.5] events/min; difference +5.30 events/min; p=0.0039), without a significant difference between DLC-PICU and DLC-turbine. Ineffective efforts, double triggering, and premature cycling did not differ across configurations. All NIV configurations reduced inspiratory effort versus unassisted spontaneous breathing, with no significant differences among NIV configurations  Conclusions: AI was lower with SLC-turbine and DLC-turbine than with DLC-PICU and lowest with SLC-turbine. Auto-triggering showed the clearest configuration-dependent difference. All configurations reduced inspiratory effort versus unassisted spontaneous breathing, without significant between-configuration differences. Because ventilator platform, circuit and expiratory architecture, and device-specific control functions differed, effects of individual components cannot be determined. These findings are specific to the ventilators and configurations evaluated. Further studies should determine whether these physiological differences translate into improved NIV tolerance or short- and long-term clinical outcomes  Trial registration: ClinicalTrials.gov, NCT04017780. Registered on 12 July 2019
SN  - 1364-8535
SN  - 1466-609X
ER  -