P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography

Abstract

Background: Pulse oximetry is widely used to assess sleep disordered breathing and nocturnal hypoxia. Although variability between devices can influence clinical decision-making, agreement during polysomnography (PSG) remains poorly characterised. This study aims to describe agreement and clinical interchangeability between two common pulse oximeters.

Methods: A retrospective cohort study of adults undergoing overnight PSG to determine the agreement between oximeters regarding oxygen desaturation index (ODI) and percentage of total sleep time with oxygen saturation (SpO2) ≤88% (%TST≤88%). Secondary outcomes were agreement of SpO2 (baseline awake, mean sleep and nadir) and the proportion of participants meeting criteria for nocturnal supplemental oxygen with each device. Bland-Altman analyses were conducted with predefined clinically acceptable differences of ±5 events/h for ODI and ±3% for SpO2 determined a priori.

Results: Data from 335 polysomnographic studies (316 participants) were included. Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h). Agreement for %TST ≤88% also fell outside the maximum acceptable difference (MD 6.1 ± 12.5%, LOA -18.5 to 30.5%). Individuals with baseline SpO2 ≤92% measured with either oximeter, were more likely to meet criteria for nocturnal supplemental oxygen therapy with the Nonin WristOx2 compared with the Masimo Radical-7 oximeter (baseline SpO2 ≤92% using Nonin: 37.3% vs 11.9%, p<.001; baseline SpO2 ≤92% using Masimo: 60% vs 24%, p<.004, respectively).

Conclusion: Agreement between oximeters was poor. These oximeters should not be used interchangeably as measurement differences may alter classification of disease severity, impacting clinical management.

Related

  1. Clinical and Polysomnographic Correlates of Subsequent Police-Recorded Motor Vehicle Accidents in Obstructive Sleep Apnea Patients: One- and Two-Year Follow-Up After Diagnostic Laboratory Polysomnography
    Yu-Ching Hsu, Yu-Wen Chien, Cheng-Yu Lin
    Medicina (Kaunas, Lithuania)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  2. P145. Agreement between continuous positive airway pressure (CPAP) machine-derived apnoea-hyponoea index (AHI) and polysomnography measured AHI in children
    D O'Neill, A Tai, T Goddard +5
    Sleep advances : a journal of the Sleep Research SocietyPeer-reviewedCC BY-NC-NDFull textPhysiologyEurope PMCPMC
  3. In-laboratory polysomnography as the reference standard in sleep research: a bibliometric overview of human studies.
    Panpan Liu, Mingzhu He, Wenlong Xing +2
    Clinics (Sao Paulo, Brazil)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  4. Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison
    Rodrigo Torres-Castro, Matías Otto-Yáñez, Claudia Saldaña-Villegas +5
    Diagnostics (Basel, Switzerland)Peer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. P015. Medical escalation criteria for heart rate monitoring during polysomnography
    J Stonehouse, A Perkins, A Matthews +2
    Sleep advances : a journal of the Sleep Research SocietyPeer-reviewedCC BY-NC-NDFull textSurgery & anaesthesiaEurope PMCPMC
BibTeX
@article{matthews2026p017,
  title = {{P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography}},
  author = {J Matthews and O McGuiness and M Milross and A Piper and K Melehan and F Lowrie and A Myint and C Menadue},
  journal = {Sleep advances : a journal of the Sleep Research Society},
  year = {2026},
  month = nov,
  doi = {10.1093/sleepadvances/zpag103.082},
  url = {https://europepmc.org/article/PMC/PMC13783417},
  note = {CC BY-NC-ND},
}
RIS
TY  - JOUR
TI  - P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography
AU  - J Matthews
AU  - O McGuiness
AU  - M Milross
AU  - A Piper
AU  - K Melehan
AU  - F Lowrie
AU  - A Myint
AU  - C Menadue
PY  - 2026
DA  - 2026/11/01
JO  - Sleep advances : a journal of the Sleep Research Society
DO  - 10.1093/sleepadvances/zpag103.082
UR  - https://europepmc.org/article/PMC/PMC13783417
AB  - Background: Pulse oximetry is widely used to assess sleep disordered breathing and nocturnal hypoxia. Although variability between devices can influence clinical decision-making, agreement during polysomnography (PSG) remains poorly characterised. This study aims to describe agreement and clinical interchangeability between two common pulse oximeters.  Methods: A retrospective cohort study of adults undergoing overnight PSG to determine the agreement between oximeters regarding oxygen desaturation index (ODI) and percentage of total sleep time with oxygen saturation (SpO2) ≤88% (%TST≤88%). Secondary outcomes were agreement of SpO2 (baseline awake, mean sleep and nadir) and the proportion of participants meeting criteria for nocturnal supplemental oxygen with each device. Bland-Altman analyses were conducted with predefined clinically acceptable differences of ±5 events/h for ODI and ±3% for SpO2 determined a priori.  Results: Data from 335 polysomnographic studies (316 participants) were included. Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h). Agreement for %TST ≤88% also fell outside the maximum acceptable difference (MD 6.1 ± 12.5%, LOA -18.5 to 30.5%). Individuals with baseline SpO2 ≤92% measured with either oximeter, were more likely to meet criteria for nocturnal supplemental oxygen therapy with the Nonin WristOx2 compared with the Masimo Radical-7 oximeter (baseline SpO2 ≤92% using Nonin: 37.3% vs 11.9%, p<.001; baseline SpO2 ≤92% using Masimo: 60% vs 24%, p<.004, respectively).  Conclusion: Agreement between oximeters was poor. These oximeters should not be used interchangeably as measurement differences may alter classification of disease severity, impacting clinical management.
SN  - 2632-5012
ER  -