P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography
Abstract
Background: Pulse oximetry is widely used to assess sleep disordered breathing and nocturnal hypoxia. Although variability between devices can influence clinical decision-making, agreement during polysomnography (PSG) remains poorly characterised. This study aims to describe agreement and clinical interchangeability between two common pulse oximeters.
Methods: A retrospective cohort study of adults undergoing overnight PSG to determine the agreement between oximeters regarding oxygen desaturation index (ODI) and percentage of total sleep time with oxygen saturation (SpO2) ≤88% (%TST≤88%). Secondary outcomes were agreement of SpO2 (baseline awake, mean sleep and nadir) and the proportion of participants meeting criteria for nocturnal supplemental oxygen with each device. Bland-Altman analyses were conducted with predefined clinically acceptable differences of ±5 events/h for ODI and ±3% for SpO2 determined a priori.
Results: Data from 335 polysomnographic studies (316 participants) were included. Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h). Agreement for %TST ≤88% also fell outside the maximum acceptable difference (MD 6.1 ± 12.5%, LOA -18.5 to 30.5%). Individuals with baseline SpO2 ≤92% measured with either oximeter, were more likely to meet criteria for nocturnal supplemental oxygen therapy with the Nonin WristOx2 compared with the Masimo Radical-7 oximeter (baseline SpO2 ≤92% using Nonin: 37.3% vs 11.9%, p<.001; baseline SpO2 ≤92% using Masimo: 60% vs 24%, p<.004, respectively).
Conclusion: Agreement between oximeters was poor. These oximeters should not be used interchangeably as measurement differences may alter classification of disease severity, impacting clinical management.
Related
- Clinical and Polysomnographic Correlates of Subsequent Police-Recorded Motor Vehicle Accidents in Obstructive Sleep Apnea Patients: One- and Two-Year Follow-Up After Diagnostic Laboratory Polysomnography
- P145. Agreement between continuous positive airway pressure (CPAP) machine-derived apnoea-hyponoea index (AHI) and polysomnography measured AHI in children
- In-laboratory polysomnography as the reference standard in sleep research: a bibliometric overview of human studies.
- Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison
- P015. Medical escalation criteria for heart rate monitoring during polysomnography
BibTeX
@article{matthews2026p017,
title = {{P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography}},
author = {J Matthews and O McGuiness and M Milross and A Piper and K Melehan and F Lowrie and A Myint and C Menadue},
journal = {Sleep advances : a journal of the Sleep Research Society},
year = {2026},
month = nov,
doi = {10.1093/sleepadvances/zpag103.082},
url = {https://europepmc.org/article/PMC/PMC13783417},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnography
AU - J Matthews
AU - O McGuiness
AU - M Milross
AU - A Piper
AU - K Melehan
AU - F Lowrie
AU - A Myint
AU - C Menadue
PY - 2026
DA - 2026/11/01
JO - Sleep advances : a journal of the Sleep Research Society
DO - 10.1093/sleepadvances/zpag103.082
UR - https://europepmc.org/article/PMC/PMC13783417
AB - Background: Pulse oximetry is widely used to assess sleep disordered breathing and nocturnal hypoxia. Although variability between devices can influence clinical decision-making, agreement during polysomnography (PSG) remains poorly characterised. This study aims to describe agreement and clinical interchangeability between two common pulse oximeters. Methods: A retrospective cohort study of adults undergoing overnight PSG to determine the agreement between oximeters regarding oxygen desaturation index (ODI) and percentage of total sleep time with oxygen saturation (SpO2) ≤88% (%TST≤88%). Secondary outcomes were agreement of SpO2 (baseline awake, mean sleep and nadir) and the proportion of participants meeting criteria for nocturnal supplemental oxygen with each device. Bland-Altman analyses were conducted with predefined clinically acceptable differences of ±5 events/h for ODI and ±3% for SpO2 determined a priori. Results: Data from 335 polysomnographic studies (316 participants) were included. Agreement between oximeters for ODI exceeded the maximum acceptable difference (mean difference (MD) 4.8 ±6.7/h, limits of agreement (LOA) -8.4 to 18 events/h). Agreement for %TST ≤88% also fell outside the maximum acceptable difference (MD 6.1 ± 12.5%, LOA -18.5 to 30.5%). Individuals with baseline SpO2 ≤92% measured with either oximeter, were more likely to meet criteria for nocturnal supplemental oxygen therapy with the Nonin WristOx2 compared with the Masimo Radical-7 oximeter (baseline SpO2 ≤92% using Nonin: 37.3% vs 11.9%, p<.001; baseline SpO2 ≤92% using Masimo: 60% vs 24%, p<.004, respectively). Conclusion: Agreement between oximeters was poor. These oximeters should not be used interchangeably as measurement differences may alter classification of disease severity, impacting clinical management.
SN - 2632-5012
ER -