Performance of a Gamma Detection Unit with Likelihood-Based Localization Method
4 affiliations
- 1Institute of Medical Device Technology, University of Stuttgart, Pfaffenwaldring 9, 70569,Stuttgart, Germany
- 2Institute of Industrial Automation and Software Engineering, University of Stuttgart,Stuttgart, Germany
- 3Medsun labs sp. z o. o., Warszawa, Poland / Faculty of Computer Science AGH University of Krakow,Krakow, Poland
- 4Institute of Medical Device Technology,Stuttgart, Germany
Abstract
In this paper a new gamma detector for combination with ultrasound imaging to localize radioactively marked lymph nodes is presented. The detector consists of six silicon photomultiplier diodes arranged around a rectangle, to fit an ultrasound device in the center. The localization is calculated depending on a probability field. The system was evaluated with a Co-57 gamma source, which was positioned in a 3 times 3 matrix for three different distances. The position estimation achieves an euclidean error of 2 - 47 mm dependent on the position and distance.
Related
- Quantum Annealing for Gamma-ray Source Detection and Localization with AGILE Data
- Machine learning for advancing gamma source localization in a human phantom using NaI (Tl) detector spectra in radiobioassay
- Optimal state detection without likelihood estimation
- Likelihood-Based Unsupervised Anomaly Detection in CMS Dijet Events
- A method for enhanced gamma-proton discrimination with imaging atmospheric Cherenkov telescopes
BibTeX
@article{robin2026performance,
title = {{Performance of a Gamma Detection Unit with Likelihood-Based Localization Method}},
author = {Meffle Robin J. and Baum Sebastian and Heryan Katarzyna and Pott Peter P.},
journal = {Current Directions in Biomedical Engineering},
year = {2026},
month = oct,
doi = {10.1515/cdbme-2026-0186},
url = {https://doi.org/10.1515/cdbme-2026-0186},
note = {CC BY-NC-ND},
}
RIS
TY - JOUR
TI - Performance of a Gamma Detection Unit with Likelihood-Based Localization Method
AU - Meffle Robin J.
AU - Baum Sebastian
AU - Heryan Katarzyna
AU - Pott Peter P.
PY - 2026
DA - 2026/10/26
JO - Current Directions in Biomedical Engineering
DO - 10.1515/cdbme-2026-0186
UR - https://doi.org/10.1515/cdbme-2026-0186
AB - In this paper a new gamma detector for combination with ultrasound imaging to localize radioactively marked lymph nodes is presented. The detector consists of six silicon photomultiplier diodes arranged around a rectangle, to fit an ultrasound device in the center. The localization is calculated depending on a probability field. The system was evaluated with a Co-57 gamma source, which was positioned in a 3 times 3 matrix for three different distances. The position estimation achieves an euclidean error of 2 - 47 mm dependent on the position and distance.
SN - 2364-5504
ER -