Cardio-respiratory-gated proton pencil beam scanning for cardiac ablation: workflow and end-to-end validation
6 affiliations
- 1Mass General Brigham, Harvard Medical School
- 2Yale University
- 3EBAMed SA
- 4Massachusetts Institute of Technology
- 5Pyramid Technical Consultants
- 6University of Utah
Abstract
Background: Stereotactic Arrhythmia Radiotherapy (STAR) offers a non-invasive alternative for ventricular tachycardia ablation, yet motion management remains a critical challenge. While respiratory gating is established, cardiac gating requires a dedicated workflow and gating device.
Methods: This study establishes and validates an end-to-end workflow for cardiorespiratory gated proton STAR in six healthy swine. Key innovations include a novel pencil-beam scanning (PBS) controller enabling mid-spot beam pausing and resumption, and the integration of CardioKit, an ultrasound-based image-guidance system for real-time soft-tissue tracking. The workflow was evaluated by comparing single-gated (respiratory only) and dual-gated (cardiorespiratory) delivery strategies, assessing system latency, dosimetric accuracy on dynamic phantoms, and motion tracking fidelity against 4D-CT benchmarks.
Results: The upgraded PBS system demonstrated a gate response latency of <550 µs, effectively mitigating interplay effects in dynamic phantom measurements through gated beam deliveries. CardioKit provided respiratory motion traces highly correlated with 4D-CT data (Pearson r > 0.97) and responded to arrhythmic events within 5 ms. In vivo , dual-gating restricted delivery to the diastolic window (70–100% R–R), which resulted in a reduction in the Internal Target Volume (ITV) expansion by a mean of 45% compared to single-gating (19% vs. 64%), thereby substantially sparing healthy myocardium. Although dual-gating increased total beam-on time, the mid-spot interrupt capability ensured robust dose accumulation without treatment aborts.
Conclusion: The workflow for imaging, treatment planning, and cardiorespiratory-gated beam delivery was viable for treating all six subjects. The PBS system responded to gating commands < 550 µs and CardioKit responded to irregularities in heartbeat in <5 ms. Cardiorespiratory gating enabled an average 45% reduction in target expansion due to cardiac motion. Comparison of CardioKit respiratory motion traces with respiratory 4D-CT yielded differences of 1.1-1.4 mm in total motion amplitude and 0.97-0.99 Pearson correlation factor. These findings are valuable for establishing a workflow for cardiorespiratory-gated STAR treatment with CardioKit.
Related
- High-power pulsed-laser ablation of aluminum alloy: Simulation model and experimental validation
- Validation of a Computational Respiratory System Model for Mechanical Ventilation
- Attosecond current control and timing in a scanning tunnelling microscope
- Enhanced nanometer-scale linewidth metrology using dark-field through-focus scanning optical microscopy
- Scanning through space and time: past, present, and future of time-resolved scanning transmission soft X-ray microscopy
BibTeX
@article{depauw2026cardiorespiratorygated,
title = {{Cardio-respiratory-gated proton pencil beam scanning for cardiac ablation: workflow and end-to-end validation}},
author = {Nicolas Depauw and Felicitas Bijari and Ethan Cascio and Evan Driscoll and Adriano Garonna and Jagdip Kang and Helen Deng and Hyeongseok Kim and Marina Macdonald-Soccorso and Willian Nett and Marc Normandin and Matthieu Pellissier-Tanon and Rebecca Woodcock and Koushiar Yaghoubian and Georges El Fakhri and Robert Levine and H.Immo Lehmann and Yen-Lin Chen and Kyung-Wook Jee},
journal = {Frontiers in Oncology},
year = {2026},
month = oct,
doi = {10.3389/fonc.2026.1923998},
url = {https://www.frontiersin.org/articles/10.3389/fonc.2026.1923998/full},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Cardio-respiratory-gated proton pencil beam scanning for cardiac ablation: workflow and end-to-end validation
AU - Nicolas Depauw
AU - Felicitas Bijari
AU - Ethan Cascio
AU - Evan Driscoll
AU - Adriano Garonna
AU - Jagdip Kang
AU - Helen Deng
AU - Hyeongseok Kim
AU - Marina Macdonald-Soccorso
AU - Willian Nett
AU - Marc Normandin
AU - Matthieu Pellissier-Tanon
AU - Rebecca Woodcock
AU - Koushiar Yaghoubian
AU - Georges El Fakhri
AU - Robert Levine
AU - H.Immo Lehmann
AU - Yen-Lin Chen
AU - Kyung-Wook Jee
PY - 2026
DA - 2026/10/23
JO - Frontiers in Oncology
DO - 10.3389/fonc.2026.1923998
UR - https://www.frontiersin.org/articles/10.3389/fonc.2026.1923998/full
AB - Background: Stereotactic Arrhythmia Radiotherapy (STAR) offers a non-invasive alternative for ventricular tachycardia ablation, yet motion management remains a critical challenge. While respiratory gating is established, cardiac gating requires a dedicated workflow and gating device. Methods: This study establishes and validates an end-to-end workflow for cardiorespiratory gated proton STAR in six healthy swine. Key innovations include a novel pencil-beam scanning (PBS) controller enabling mid-spot beam pausing and resumption, and the integration of CardioKit, an ultrasound-based image-guidance system for real-time soft-tissue tracking. The workflow was evaluated by comparing single-gated (respiratory only) and dual-gated (cardiorespiratory) delivery strategies, assessing system latency, dosimetric accuracy on dynamic phantoms, and motion tracking fidelity against 4D-CT benchmarks. Results: The upgraded PBS system demonstrated a gate response latency of <550 µs, effectively mitigating interplay effects in dynamic phantom measurements through gated beam deliveries. CardioKit provided respiratory motion traces highly correlated with 4D-CT data (Pearson r > 0.97) and responded to arrhythmic events within 5 ms. In vivo , dual-gating restricted delivery to the diastolic window (70–100% R–R), which resulted in a reduction in the Internal Target Volume (ITV) expansion by a mean of 45% compared to single-gating (19% vs. 64%), thereby substantially sparing healthy myocardium. Although dual-gating increased total beam-on time, the mid-spot interrupt capability ensured robust dose accumulation without treatment aborts. Conclusion: The workflow for imaging, treatment planning, and cardiorespiratory-gated beam delivery was viable for treating all six subjects. The PBS system responded to gating commands < 550 µs and CardioKit responded to irregularities in heartbeat in <5 ms. Cardiorespiratory gating enabled an average 45% reduction in target expansion due to cardiac motion. Comparison of CardioKit respiratory motion traces with respiratory 4D-CT yielded differences of 1.1-1.4 mm in total motion amplitude and 0.97-0.99 Pearson correlation factor. These findings are valuable for establishing a workflow for cardiorespiratory-gated STAR treatment with CardioKit.
SN - 2234-943X
ER -