Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.
- 1College of Bioresource Sciences, Nihon University, 1866 Kameino, Fujisawa, Kanagawa, Japan. [email protected].
- 2College of Bioresource Sciences, Nihon University, 1866 Kameino, Fujisawa, Kanagawa, Japan.
- 3Hitachi High-Tech Analysis Corporation, 1-17-1, Toranomon, Minato-ku, Tokyo, Japan.
Abstract
Background
Therapeutic drug monitoring (TDM) is essential for drugs with narrow therapeutic ranges and high pharmacokinetic variability. However, significant barriers hinder TDM implementation. Outsourcing of testing can delay timely clinical decision-making, with possible delays of up to 1 week. Additionally, management of analyte-specific certified reference material (CRM) presents substantial operational challenges, particularly for unstable compounds. The present study therefore developed and validated a high-performance liquid chromatography (HPLC) method based on relative molar sensitivity (RMS) of stable non-analyte reference materials (carbamazepine and caffeine) to quantify five clinically important analytes-acetaminophen, two beta-lactam antibiotics (cefmetazole and cefazolin), and amiodarone and its active metabolite, N-desethylamiodarone.Methods
RMS values were calculated as the ratio of the analyte and non-analyte reference material calibration curve slopes. CRM solutions were used at 6 to 10 concentrations per analyte. HPLC analysis utilized analyte-specific detection wavelengths in the range 240-275 nm. For validation, control serum samples were spiked at 2 or 3 concentrations per analyte and processed via spin-column solid-phase extraction. Quantitative accuracy was assessed by comparing the RMS-based results with conventional absolute calibration methods.Results
All calibration curves demonstrated excellent linearity, with correlation coefficients in the range 0.9993-1.0000. Calibration curve slope reproducibility was high, with relative standard deviations of 0.3%-1.5% (n = 3). Quantitative analyses of spiked serum samples showed close agreement between the RMS and conventional methods, with relative differences not exceeding ± 0.50% for either calibrant for any analyte. Both methods achieved comparable precision, with relative standard deviations of < 4.6%.Conclusions
The validated RMS-based HPLC method provides accurate and precise quantification of structurally diverse drugs using stable, the International System of Units (SI)-traceable, non-analyte reference materials. This approach eliminates the need for analyte-specific authentic standards for routine analyses, thus substantially reducing procurement costs and addressing the instability of some analytes. The rapid and accurate analysis enables timely reporting, potentially leading to improved clinical outcomes through expedited in-house TDM implementation in hospital laboratories with limited resources.
Related
- In Vitro Release Performance of Naproxen Sodium/Acetaminophen Pediatric Suspensions Using a Mini-Paddle Apparatus and USP Paddle Apparatus
- Enhancing acetaminophen solubility and stability using an aqueous betaine chloride–acetic acid natural deep eutectic solvent: experimental and mechanistic insights
- The Role of Titanium Species in Ti-SBA-15 Materials for the UV–Photocatalytic Degradation of Acetaminophen in Water
- Metabolite profiling of a Bolboschoenus planiculmis extract and its antioxidant, anti-inflammatory, and enzyme inhibitory properties with potential for skin health applications
- Organic electrochemical transistors for metabolite sensing across the transition from in vitro to in vivo
BibTeX
@article{ohtsuki2026development,
title = {{Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.}},
author = {Takashi Ohtsuki and Yume Enomoto and Akane Yasuda and Risa Imai and Nonoka Seto and Makoto Hashimoto and Miyuki Matsushita and Satoru Morikawa and Hiroshi Matsufuji},
journal = {Journal of pharmaceutical health care and sciences},
year = {2026},
month = aug,
doi = {10.1186/s40780-026-00615-w},
url = {https://europepmc.org/article/MED/42972490},
note = {CC BY},
}
RIS
TY - JOUR
TI - Development of a relative molar sensitivity-based HPLC method for quantifying acetaminophen, two beta-lactams, amiodarone, and its metabolite.
AU - Takashi Ohtsuki
AU - Yume Enomoto
AU - Akane Yasuda
AU - Risa Imai
AU - Nonoka Seto
AU - Makoto Hashimoto
AU - Miyuki Matsushita
AU - Satoru Morikawa
AU - Hiroshi Matsufuji
PY - 2026
DA - 2026/08/03
JO - Journal of pharmaceutical health care and sciences
DO - 10.1186/s40780-026-00615-w
UR - https://europepmc.org/article/MED/42972490
AB - Background Therapeutic drug monitoring (TDM) is essential for drugs with narrow therapeutic ranges and high pharmacokinetic variability. However, significant barriers hinder TDM implementation. Outsourcing of testing can delay timely clinical decision-making, with possible delays of up to 1 week. Additionally, management of analyte-specific certified reference material (CRM) presents substantial operational challenges, particularly for unstable compounds. The present study therefore developed and validated a high-performance liquid chromatography (HPLC) method based on relative molar sensitivity (RMS) of stable non-analyte reference materials (carbamazepine and caffeine) to quantify five clinically important analytes-acetaminophen, two beta-lactam antibiotics (cefmetazole and cefazolin), and amiodarone and its active metabolite, N-desethylamiodarone.Methods RMS values were calculated as the ratio of the analyte and non-analyte reference material calibration curve slopes. CRM solutions were used at 6 to 10 concentrations per analyte. HPLC analysis utilized analyte-specific detection wavelengths in the range 240-275 nm. For validation, control serum samples were spiked at 2 or 3 concentrations per analyte and processed via spin-column solid-phase extraction. Quantitative accuracy was assessed by comparing the RMS-based results with conventional absolute calibration methods.Results All calibration curves demonstrated excellent linearity, with correlation coefficients in the range 0.9993-1.0000. Calibration curve slope reproducibility was high, with relative standard deviations of 0.3%-1.5% (n = 3). Quantitative analyses of spiked serum samples showed close agreement between the RMS and conventional methods, with relative differences not exceeding ± 0.50% for either calibrant for any analyte. Both methods achieved comparable precision, with relative standard deviations of < 4.6%.Conclusions The validated RMS-based HPLC method provides accurate and precise quantification of structurally diverse drugs using stable, the International System of Units (SI)-traceable, non-analyte reference materials. This approach eliminates the need for analyte-specific authentic standards for routine analyses, thus substantially reducing procurement costs and addressing the instability of some analytes. The rapid and accurate analysis enables timely reporting, potentially leading to improved clinical outcomes through expedited in-house TDM implementation in hospital laboratories with limited resources.
SN - 2055-0294
ER -