Towards in vivo photon dosimetry using radiation-sensitive mixed-core ultrasound phase-change contrast agents

Abstract

Objective. Advanced radiotherapy applies complex treatment plans that require verification of the beam delivery. Therefore, in vivo dosimetry, measuring delivered radiation doses directly on the patient, has become increasingly important. Radiation-sensitive phase-change nanodroplets were recently proposed for in vivo radiation sensing. The radiation-induced conversion of superheated nanodroplets into microbubbles generates ultrasound contrast from which radiation deposition can be derived. While successful for in vivo proton range verification, previously used nanodroplet formulations were not suitable for in vivo photon dosimetry due to stringent requirements on the degree of superheat of the nanodroplet core. Here, we used a mixed-core strategy, combining different perfluorocarbons in the same nanodroplet core, to modulate the photon response and droplet stability at body temperature. Approach. Lipid-shelled nanodroplets with different cores (pure octafluoropropane, pure decafluorobutane and a 17:4 (v/v) octafluoropropane:decafluorobutane mixture) were dispersed in deionized water in phantom containers and irradiated with a photons at various temperatures. Simultaneous high-frame rate ultrasound acquisition allowed to link ultrasound contrast generation to the radiation delivery. Nanodroplets were characterized using 19F NMR spectroscopy. Main results. Photons induced nanodroplet-mediated ultrasound contrast strong enough for clinical applications at body temperature. A clear trade-off was observed between a stronger photon response and maintaining sufficient droplet stability. Neither octafluoropropane (lacking stability) nor decafluorobutane (lacking photon response) cores proved suitable. On the contrary, the mixed-core formulation exhibited a strong photon response with minimal background already at small radiation doses (0.1 Gy). This paves the way for future in vivo dosimetry.

Related

  1. EP09.28 Quality assurance in a national randomized trial of radiation target volume reduction in glioblastoma: results from the contouring and dosimetry dummy run
    D Di Perri, C De Laet, M Lambrecht +28
    Neuro-OncologyPeer-reviewedPublisher termsFull textApplied physics & instrumentationPMC
  2. Single-Pulse Dosimetry for Ultra-High-Dose-Rate X-Ray Pulses Based on Radiation-Induced Acoustic Signals
    Qihang Han, Ke Chen, Chenghao Hua +7
    arXivPreprintarXiv licenceFull textApplied physics & instrumentationarXiv
  3. A single radiation hard pixel detector for proton dosimetry, imaging, energy resolution and beam monitoring
    Nico Frederic Brosda, Stephane Higueret, Nicolas Arbor +7
    HALPreprintarXiv licenceFull textApplied physics & instrumentationBiological & medical physicsHALarXiv
  4. Optical contrast from speckle patterns with focused modern synchrotron radiation beams
    Federico Zontone, Yuriy Chushkin, Marco Cammarata +1
    Optics ExpressPeer-reviewedLicence not statedOptics & photonicsCrossref
  5. EP09.13 Incidence, characteristics, and risk factors of radiation-induced contrast enhancements (RICE) after re-irradiation for central nervous system and base of skull tumors
    P Mara, C Lütgendorf-Caucig, J Gora +6
    Neuro-OncologyPeer-reviewedPublisher termsFull textApplied physics & instrumentationPMC
BibTeX
@misc{carlier2026towards,
  title = {{Towards in vivo photon dosimetry using radiation-sensitive mixed-core ultrasound phase-change contrast agents}},
  author = {Bram Carlier and Reza Bahrami Gorji and Qingyuan Tan and Meritxell Gambús-Marfà and Uwe Himmelreich and Meng-Xing Tang and Koen Van Den Abeele and Sophie V. Heymans},
  howpublished = {arXiv},
  year = {2026},
  month = nov,
  doi = {10.48550/arxiv.2611.10142},
  eprint = {2611.10142},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2611.10142},
}
RIS
TY  - UNPB
TI  - Towards in vivo photon dosimetry using radiation-sensitive mixed-core ultrasound phase-change contrast agents
AU  - Bram Carlier
AU  - Reza Bahrami Gorji
AU  - Qingyuan Tan
AU  - Meritxell Gambús-Marfà
AU  - Uwe Himmelreich
AU  - Meng-Xing Tang
AU  - Koen Van Den Abeele
AU  - Sophie V. Heymans
PY  - 2026
DA  - 2026/11/06
PB  - arXiv
DO  - 10.48550/arxiv.2611.10142
UR  - https://arxiv.org/abs/2611.10142
AB  - Objective. Advanced radiotherapy applies complex treatment plans that require verification of the beam delivery. Therefore, in vivo dosimetry, measuring delivered radiation doses directly on the patient, has become increasingly important. Radiation-sensitive phase-change nanodroplets were recently proposed for in vivo radiation sensing. The radiation-induced conversion of superheated nanodroplets into microbubbles generates ultrasound contrast from which radiation deposition can be derived. While successful for in vivo proton range verification, previously used nanodroplet formulations were not suitable for in vivo photon dosimetry due to stringent requirements on the degree of superheat of the nanodroplet core. Here, we used a mixed-core strategy, combining different perfluorocarbons in the same nanodroplet core, to modulate the photon response and droplet stability at body temperature. Approach. Lipid-shelled nanodroplets with different cores (pure octafluoropropane, pure decafluorobutane and a 17:4 (v/v) octafluoropropane:decafluorobutane mixture) were dispersed in deionized water in phantom containers and irradiated with a photons at various temperatures. Simultaneous high-frame rate ultrasound acquisition allowed to link ultrasound contrast generation to the radiation delivery. Nanodroplets were characterized using 19F NMR spectroscopy. Main results. Photons induced nanodroplet-mediated ultrasound contrast strong enough for clinical applications at body temperature. A clear trade-off was observed between a stronger photon response and maintaining sufficient droplet stability. Neither octafluoropropane (lacking stability) nor decafluorobutane (lacking photon response) cores proved suitable. On the contrary, the mixed-core formulation exhibited a strong photon response with minimal background already at small radiation doses (0.1 Gy). This paves the way for future in vivo dosimetry.
ER  -