Recalibration of Invasive Hemodynamic Monitoring Devices on Serial Catheterization in Adults With Fontan Circulation
- 1The Heart Center at Nationwide Children's Hospital, Columbus, Ohio, USA
- 2Division of Cardiovascular Medicine, The Ohio State University Wexner Medical Center, Columbus, Ohio, USA
Abstract
Background: There are limited data on long-term accuracy of invasive hemodynamic monitoring (IHM) in patients with nonpulsatile Fontan circulation (FC).
First-in-Human/Early Reports Summary: In a retrospective review of 20 patients with FC and CardioMEMS IHM implanted (2015-2025), simultaneous IHM and invasive mean pulmonary artery pressure measurements were available during 6 repeat catheterizations in 6 patients. IHM readings were nearly all lower than invasive measurements, with the absolute difference ranging from 1 to 10 mm Hg.
Discussion/Novelty: There are limited reports of recalibration data with IHM, and to our knowledge this is the first report of simultaneous IHM and invasive measurement in adults with FC. IHM accuracy is important with FC as small pressure differences can greatly affect physiology.
Take-Home Messages: IHM can provide outpatient pulmonary artery pressure measurements in patients with FC. Because we find discrepancy in device measurements in our cohort, we recommend recalibration with each hemodynamic catheterization in this population.
Related
- Hemodynamic effects of cycle ergometry versus dynamic unilateral arm exercise during right heart catheterization: a prospective observational study
- Clinical and Invasive Hemodynamic Correlates of the TAPSE/sPAP Ratio in Pulmonary Hypertension
- 4D-Flow MRI-Derived Wall Shear Stress as a Non-Invasive Biomarker for the Detection and Hemodynamic Characterization of Pulmonary Hypertension.
- Personalized Remote Patient Management in Heart Failure : From Implanted Hemodynamic Monitoring Toward an Artificial Intelligence-Supported Adaptive Care Cycle.
- From Catheters to Wearables: Advanced Hemodynamic Monitoring and the Emerging Role of PPG‐Based Devices
BibTeX
@article{freddo2026recalibration,
title = {{Recalibration of Invasive Hemodynamic Monitoring Devices on Serial Catheterization in Adults With Fontan Circulation}},
author = {Andrew M. Freddo and Arash Salavitabar and Benjamin Blais and Aimee K. Armstrong and Curt J. Daniels and William H. Marshall V},
journal = {JACC Case Reports},
year = {2026},
month = nov,
doi = {10.1016/j.jaccas.2026.110190},
url = {https://doi.org/10.1016/j.jaccas.2026.110190},
note = {CC BY-NC-ND 4.0},
}
RIS
TY - JOUR
TI - Recalibration of Invasive Hemodynamic Monitoring Devices on Serial Catheterization in Adults With Fontan Circulation
AU - Andrew M. Freddo
AU - Arash Salavitabar
AU - Benjamin Blais
AU - Aimee K. Armstrong
AU - Curt J. Daniels
AU - William H. Marshall V
PY - 2026
DA - 2026/11/11
JO - JACC Case Reports
DO - 10.1016/j.jaccas.2026.110190
UR - https://doi.org/10.1016/j.jaccas.2026.110190
AB - Background: There are limited data on long-term accuracy of invasive hemodynamic monitoring (IHM) in patients with nonpulsatile Fontan circulation (FC). First-in-Human/Early Reports Summary: In a retrospective review of 20 patients with FC and CardioMEMS IHM implanted (2015-2025), simultaneous IHM and invasive mean pulmonary artery pressure measurements were available during 6 repeat catheterizations in 6 patients. IHM readings were nearly all lower than invasive measurements, with the absolute difference ranging from 1 to 10 mm Hg. Discussion/Novelty: There are limited reports of recalibration data with IHM, and to our knowledge this is the first report of simultaneous IHM and invasive measurement in adults with FC. IHM accuracy is important with FC as small pressure differences can greatly affect physiology. Take-Home Messages: IHM can provide outpatient pulmonary artery pressure measurements in patients with FC. Because we find discrepancy in device measurements in our cohort, we recommend recalibration with each hemodynamic catheterization in this population.
SN - 2666-0849
ER -