Impact of the position effects in the energy calibration of the BULLKID detector

Abstract

BULLKID is a cryogenic solid-state detector based on an array of silicon dice equipped with Kinetic Inductance Detectors (KIDs) for low-mass dark matter searches and coherent elastic neutrino-nucleus scattering. Energy calibration can be performed using optical LED photon bursts, which are absorbed within the first few hundred nanometers of the substrate. An energy calibration using a 241Am gamma-ray source revealed a <10% discrepancy with the LED calibration. A recent study by Matava and Williams suggested that phonon statistics and phonon collection varying with the interaction position could explain such a discrepancy. This work evaluates the impact of these effects on the calibration of the BULLKID detector, demonstrating they both have a negligible effect. We clarify the results obtained during the 241Am calibration regarding the origin of the observed discrepancy.

Related

  1. Cryogenic Optoelectronic Characterization of InGaAsP LEDs for Low-Energy Detector Calibration
    Muad Ghaith, Wolfgang Rau, Philippe Di Stefano +4
    arXivPreprintarXiv licenceFull textApplied physics & instrumentationHigh-energy & particle physicsarXiv
  2. Electron and photon energy calibration with the ATLAS detector using LHC Run 2 data
    Georges Aad, Braden Keim Abbott, Kira Abeling +2,944
    JINSTPeer-reviewedCC BY 4.0Full textHigh-energy & particle physicsHAL
  3. CNN-based Position Reconstruction for the T0 Detector in the HIRFL-CSR External-target Experiment
    Shuai-Ke Lv, D Hu, M Shao +2
    arXivPreprintCC BY-NC-ND 4.0Full textApplied physics & instrumentationarXiv
  4. Detectability-Aware Screening of Post-Merger Gravitational-Wave Damping Anomalies in Real Detector Noise: A Graph-Spectral Approach with Empirical Null Calibration
    Ruslan Alaskarov
    arXivPreprintCC BY 4.0Full textInstrumentation & methodsGravitation & general relativityarXiv
  5. Demonstration of a cryogenic, switchable electron source for low-temperature detector calibration
    D. A. Bennett, M. Borghesi, P. Campana +17
    arXivPreprintarXiv licenceFull textApplied physics & instrumentationarXiv
BibTeX
@misc{folcarelli2026impact,
  title = {{Impact of the position effects in the energy calibration of the BULLKID detector}},
  author = {Matteo Folcarelli and Matteo Cappelli and Giorgio del Castello and Camilla Bonomo and Angelo Cruciani and Daniele Delicato and Leonardo Pesce and Davide Quaranta and Marco Vignati},
  howpublished = {HAL},
  year = {2026},
  month = sep,
  doi = {10.48550/arxiv.2609.30004},
  eprint = {2609.30004},
  archivePrefix = {arXiv},
  url = {https://hal.science/hal-05781777v1},
  note = {CC BY 4.0},
}
RIS
TY  - UNPB
TI  - Impact of the position effects in the energy calibration of the BULLKID detector
AU  - Matteo Folcarelli
AU  - Matteo Cappelli
AU  - Giorgio del Castello
AU  - Camilla Bonomo
AU  - Angelo Cruciani
AU  - Daniele Delicato
AU  - Leonardo Pesce
AU  - Davide Quaranta
AU  - Marco Vignati
PY  - 2026
DA  - 2026/09/24
PB  - HAL
DO  - 10.48550/arxiv.2609.30004
UR  - https://hal.science/hal-05781777v1
AB  - BULLKID is a cryogenic solid-state detector based on an array of silicon dice equipped with Kinetic Inductance Detectors (KIDs) for low-mass dark matter searches and coherent elastic neutrino-nucleus scattering. Energy calibration can be performed using optical LED photon bursts, which are absorbed within the first few hundred nanometers of the substrate. An energy calibration using a ^241Am gamma-ray source revealed a <10% discrepancy with the LED calibration. A recent study by Matava and Williams suggested that phonon statistics and phonon collection varying with the interaction position could explain such a discrepancy. This work evaluates the impact of these effects on the calibration of the BULLKID detector, demonstrating they both have a negligible effect. We clarify the results obtained during the ^241Am calibration regarding the origin of the observed discrepancy.
ER  -