ECAP‐Guided Closed‐Loop Spinal Cord Stimulation for Axial Thoracic Pain: A Case Report
5 affiliations
- 1The Ohio Pain Clinic Dayton Ohio USA
- 2Northeast Ohio Medical University Rootstown Ohio USA
- 3Wright State University Dayton Ohio USA
- 4Elite Pain Doctors Mason Ohio USA
- 5Ohio University College of Medicine Athens Ohio USA
Abstract
Thoracic spine pain is among the least commonly reported pain syndromes. Conventional therapeutic strategies are frequently ineffective and may be associated with adverse events. Closed‐loop spinal cord stimulation (SCS) assisted with evoked compound action potentials (ECAPs) has emerged as a promising, safe and effective alternative. In this case report, we describe the successful use of ECAP‐controlled closed‐loop SCS in a patient with refractory thoracic spine pain. A female patient in her 40s presented with several axial thoracic pain refractory to all conservative and interventional therapies. Upon admission, the numeric rating scale (NRS) for pain score was 8/10. Functional limitations included restricted sleep duration (3 h) and limited standing and walking tolerance (20 min). The patient underwent a one‐week trial of ECAP‐controlled closed‐loop SCS, during which 24 million smart adjustments were automatically performed to maintain consistent neural activation. The dose accuracy was ±5 μV, with a target ECAP of 19.4 μV, and the measure was 16.4 μV, and the patient used the device 100% of the time. Following treatment, the patient reported improvement in pain and functional outcomes, with NRS scores decreasing to 2/10, and substantial improvements in sleep (6 h), standing, and walking duration. This case suggests that ECAP‐assisted closed‐loop SCS may provide meaningful pain relief and functional improvements in patients with refractory spine pain.
Related
- Spinal Cord Compression Syndrome Revealing Thoracic Ewing Sarcoma in a Young Adult: A Case Report
- Non-Contiguous Cervical and Thoracic Spinal Cord Injuries After Severe Polytrauma: A Case Report.
- Perioperative Spinal Cord Stimulation Combined With Targeted Muscle Reinnervation for Chronic Neuropathic Pain: A Case Report.
- Case Report: Beyond chronicity: sustained multidimensional response to spinal cord stimulation three decades after traumatic spinal cord injury.
- Preoperative Thoracic Compensation is Associated With a Lower Risk of Proximal Junctional Kyphosis After Lower Thoracic Fusion for Adult Spinal Deformity
BibTeX
@article{soin2026ecapguided,
title = {{ECAP‐Guided Closed‐Loop Spinal Cord Stimulation for Axial Thoracic Pain: A Case Report}},
author = {Aviraj Soin and Sarah Kassis and Richard Joseph Witte and Ben Sloop and Cameron Brittain},
journal = {Clinical case reports},
year = {2026},
month = nov,
doi = {10.1002/ccr3.73785},
url = {https://europepmc.org/article/PMC/PMC13725216},
note = {CC BY},
}
RIS
TY - JOUR
TI - ECAP‐Guided Closed‐Loop Spinal Cord Stimulation for Axial Thoracic Pain: A Case Report
AU - Aviraj Soin
AU - Sarah Kassis
AU - Richard Joseph Witte
AU - Ben Sloop
AU - Cameron Brittain
PY - 2026
DA - 2026/11/01
JO - Clinical case reports
DO - 10.1002/ccr3.73785
UR - https://europepmc.org/article/PMC/PMC13725216
AB - Thoracic spine pain is among the least commonly reported pain syndromes. Conventional therapeutic strategies are frequently ineffective and may be associated with adverse events. Closed‐loop spinal cord stimulation (SCS) assisted with evoked compound action potentials (ECAPs) has emerged as a promising, safe and effective alternative. In this case report, we describe the successful use of ECAP‐controlled closed‐loop SCS in a patient with refractory thoracic spine pain. A female patient in her 40s presented with several axial thoracic pain refractory to all conservative and interventional therapies. Upon admission, the numeric rating scale (NRS) for pain score was 8/10. Functional limitations included restricted sleep duration (3 h) and limited standing and walking tolerance (20 min). The patient underwent a one‐week trial of ECAP‐controlled closed‐loop SCS, during which 24 million smart adjustments were automatically performed to maintain consistent neural activation. The dose accuracy was ±5 μV, with a target ECAP of 19.4 μV, and the measure was 16.4 μV, and the patient used the device 100% of the time. Following treatment, the patient reported improvement in pain and functional outcomes, with NRS scores decreasing to 2/10, and substantial improvements in sleep (6 h), standing, and walking duration. This case suggests that ECAP‐assisted closed‐loop SCS may provide meaningful pain relief and functional improvements in patients with refractory spine pain.