Skip to content
DraconiticResearch
Search Sources API
194,226 studies31 sourcesUpdated 6 min ago
PreprintCC BYChemRxiv

Hydrophobic Interaction Chromatography Framework for Drug-to-Antibody Ratio Assessment of a High DAR Cysteine-Linked ADC

1 more authors
4 affiliations
  1. 1Pharmaceutical Analysis & Digital Technologies, Merck & Co., Inc., 126 E. Lincoln Avenue, Rahway, NJ, United States of America, 07065
  2. 2Large Molecule Analytical, MRL, Merck & Co., Inc., 126 E. Lincoln Avenue, Rahway, NJ, United States of America, 07065
  3. 3School of Pharmaceutical Sciences, University of Geneva, CMU – Rue Michel-Servet 1, Geneva, Switzerland, 1211
  4. 4Institute of Pharmaceutical Sciences of Western Switzerland, University of Geneva, CMU – 11 Rue Michel-Servet 1, Geneva, Switzerland, 1211

Abstract

Accurate quantitation of the drug-to-antibody ratio (DAR) using Quality Control (QC)-friendly analytical methods is critical for characterizing antibody–drug conjugates (ADCs) and ensuring product quality. Hydrophobic interaction chromatography (HIC) is increasingly recognized as the industry standard for DAR measurement of cysteine- and engineered-cysteine–linked ADCs in the biopharmaceutical industry. While low-DAR species (DAR ≈ 2–4) are readily resolved using standard HIC conditions, high-DAR species (DAR ≈ 6–8) present significant challenges due to coelution, peak broadening, and adsorption driven by increased hydrophobicity and sample heterogeneity. Systematic strategies for developing robust HIC methods tailored to high-DAR ADCs are limited, underscoring the need to understand the interplay of separation parameters that govern selectivity and resolution. Herein, a practical workflow for HIC method development specifically designed for high-DAR ADCs using readily available LC technology is introduced. The method development strategy begins with the screening of stationary phases and salt systems, to balance retention and resolution across the DAR distribution, with a particular emphasis on selectivity for highly hydrophobic conjugates. Secondary parameters including column type, temperature and organic modifier content are then optimized to modulate analyte–stationary phase interactions and improve peak shapes without compromising ADC integrity. Building on these findings, we developed a simplified two-segment linear gradient: an initial gentle segment for effective elution of low- to mid-DAR species with adequate dispersion, followed by a shallower segment optimized for high-DAR components that require stronger desorption conditions. Additionally, in silico software-based capabilities were successfully employed to map the separation landscape of ADC multicomponent mixtures using polynomial regression models, to predict optimal operating windows and assess robustness prior to QC deployment. When applying the multifactorial modeling of retention times for high-DAR ADCs with diverse linker–payload chemistries, this workflow delivered reproducible resolution of adjacent species and critical pairs, accurate DAR quantitation, and enhanced assay transferability. This analytical framework provides a clear pathway towards streamlined HIC assay development to reliably characterize high-DAR ADCs, while also introducing in silico robustness for reliable deployment across late-stage product development.

Related

  1. Hydrophobic Interaction as the Result of Interfacial Frustration
    Prithwish Mukherjee
    ChemRxivPreprintCC BYFull textMolecular biology & gene regulationChemRxiv
  2. Hydrophilic interaction chromatography and tandem mass spectrometry for the characterization of methylated oligoribonucleotides
    Khaoula Adouairi, Carole Farre, Carole Chaix +2
    Journal of Pharmaceutical and Biomedical AnalysisPeer-reviewedCC BY-NC 4.0Full text1 citedProteomics & metabolomicsGeneral biologyHAL
  3. Chroma-Agent-Alpha: An Open-Source Framework for Automated Chromatography Preprocessing, GNN-Based Peak Deconvolution, and FAIR Data Lineage
    Devendra Kataria
    ChemRxivPreprintCC BYFull textMolecular biology & gene regulationChemRxiv
  4. A unified and composable framework for modeling packed-bed liquid chromatography
    Jan Michael Breuer, Samuel Leweke, Johannes Schmölder +3
    arXivPreprintCC BY 4.0Full textMolecular biology & gene regulationarXiv
  5. Influence of antibody-antigen interaction on fluorescence of Rhodamine 6G conjugated to the antibodies measured in human serum
    D. Thiyagaraj, A. Del Re, Q. D. Pham +3
    bioRxivPreprintCC BY-NCFull textBiophysicsbioRxiv
BibTeX
@misc{barrientos2026hydrophobic,
  title = {{Hydrophobic Interaction Chromatography Framework for Drug-to-Antibody Ratio Assessment of a High DAR Cysteine-Linked ADC}},
  author = {Rodell C. Barrientos and Wei Lin and Meghan O’Leary DaSilva and Zhengqi Zhang and Eli Larson and Devin M. Makey and Heather Wang and Andrew Singh and Pankaj Aggrawal and Xiaodong Bu and Davy Guillarme and Emmanuel Appiah-Amponsah and Erik L. Regalado},
  howpublished = {ChemRxiv},
  year = {2026},
  month = oct,
  doi = {10.26434/chemrxiv.15011093/v1},
  url = {https://chemrxiv.org/doi/full/10.26434/chemrxiv.15011093/v1},
  note = {CC BY},
}
RIS
TY  - UNPB
TI  - Hydrophobic Interaction Chromatography Framework for Drug-to-Antibody Ratio Assessment of a High DAR Cysteine-Linked ADC
AU  - Rodell C. Barrientos
AU  - Wei Lin
AU  - Meghan O’Leary DaSilva
AU  - Zhengqi Zhang
AU  - Eli Larson
AU  - Devin M. Makey
AU  - Heather Wang
AU  - Andrew Singh
AU  - Pankaj Aggrawal
AU  - Xiaodong Bu
AU  - Davy Guillarme
AU  - Emmanuel Appiah-Amponsah
AU  - Erik L. Regalado
PY  - 2026
DA  - 2026/10/27
PB  - ChemRxiv
DO  - 10.26434/chemrxiv.15011093/v1
UR  - https://chemrxiv.org/doi/full/10.26434/chemrxiv.15011093/v1
AB  - Accurate quantitation of the drug-to-antibody ratio (DAR) using Quality Control (QC)-friendly analytical methods is critical for characterizing antibody–drug conjugates (ADCs) and ensuring product quality. Hydrophobic interaction chromatography (HIC) is increasingly recognized as the industry standard for DAR measurement of cysteine- and engineered-cysteine–linked ADCs in the biopharmaceutical industry. While low-DAR species (DAR ≈ 2–4) are readily resolved using standard HIC conditions, high-DAR species (DAR ≈ 6–8) present significant challenges due to coelution, peak broadening, and adsorption driven by increased hydrophobicity and sample heterogeneity. Systematic strategies for developing robust HIC methods tailored to high-DAR ADCs are limited, underscoring the need to understand the interplay of separation parameters that govern selectivity and resolution. Herein, a practical workflow for HIC method development specifically designed for high-DAR ADCs using readily available LC technology is introduced. The method development strategy begins with the screening of stationary phases and salt systems, to balance retention and resolution across the DAR distribution, with a particular emphasis on selectivity for highly hydrophobic conjugates. Secondary parameters including column type, temperature and organic modifier content are then optimized to modulate analyte–stationary phase interactions and improve peak shapes without compromising ADC integrity. Building on these findings, we developed a simplified two-segment linear gradient: an initial gentle segment for effective elution of low- to mid-DAR species with adequate dispersion, followed by a shallower segment optimized for high-DAR components that require stronger desorption conditions. Additionally, in silico software-based capabilities were successfully employed to map the separation landscape of ADC multicomponent mixtures using polynomial regression models, to predict optimal operating windows and assess robustness prior to QC deployment. When applying the multifactorial modeling of retention times for high-DAR ADCs with diverse linker–payload chemistries, this workflow delivered reproducible resolution of adjacent species and critical pairs, accurate DAR quantitation, and enhanced assay transferability. This analytical framework provides a clear pathway towards streamlined HIC assay development to reliably characterize high-DAR ADCs, while also introducing in silico robustness for reliable deployment across late-stage product development.
ER  -