Compensation of Time-of-Flight Drift Using an Internal Reference Ion during Long-Duration LIMAS Depth Profiling

  1. 1Department of Earth and Planetary Sciences, Faculty of Science, Hokkaido University, Sapporo 060–0810, Japan

Abstract

In this study, the temporal behavior and compensation of time-of-flight (TOF) drift during long-duration depth profiling using Laser Ionization Mass Nanoscope (LIMAS) were investigated. Noble gas depth profiling by LIMAS requires detection of weak isotope signals with nanometer-scale depth resolution. However, such measurements often take several hours to a couple of days, during which TOF peak-position drift and statistical uncertainty may affect analytical precision. In this study, a 4He-implanted olivine substrate was analyzed as a test sample. The TOF spectra were processed by depth-layer stacking, peak extraction, and peak-position analysis. The 24Mg2+ peak was used as a matrix-derived internal reference to monitor and compensate for TOF drift. The peak-position shifts of measured ions showed clear linear correlations with the 24Mg2+ peak shift, indicating that the observed TOF drift is largely systematic. A 24Mg2+-based compensation reduced the peak-position fluctuations of the other ion species to less than 1 ns and improved the mass resolving power. These results demonstrate that the 24Mg2+ peak can be used as an internal reference for TOF drift compensation during long-duration measurements, providing a robust approach for micro- to nanoscale noble gas depth profiling using LIMAS.

Related

  1. Collision Cross Section Measurement within a Time-of-Flight Mass Analyzer
    Hamish Stewart, Dmitry Grinfeld, Arne Stefes +2
    ChemRxivPreprintLicence not statedGeneral chemistryChemRxiv
  2. Phase model for analysis and compensation of metalens fabrication errors in a refractive–metalens hybrid imaging system
    Junsong Wang, Mingxu Piao, Xian Zhang +3
    Optics ExpressPeer-reviewedLicence not statedMaterials & polymersCrossref
  3. Activation of a water molecule using a mononuclear Mn complex: from Mn-aquo, to Mn-hydroxo, to Mn-oxyl via charge compensation
    Benedikt Lassalle-Kaiser, Christelle Hureau, Dimitrios Pantazis +6
    Energy & Environmental SciencePeer-reviewedLicence not stated57 citedInorganic & organometallic chemistryHAL
  4. Beyond Swelling: Electrochemical Doping Tunes the Balance between Ion Compensation and Exchange in a Hydrated Organic Mixed Ionic–Electronic Conductor
    Zixuan Wei, Ying Xiang, Hesam Makki +12
    ChemRxivPreprintCC BYFull textMaterials & polymersChemRxiv
  5. Design and implementation of a Wireless Power Transfer with double-sided LCC compensation for a farming robot
    Aryadharma Sudhartio, Marojahan Tampubolon, Gervasius Geovan +2
    PLOS OnePeer-reviewedCC BY 4.0Full textMaterials & polymersPLOS
BibTeX
@article{bajo2026compensation,
  title = {{Compensation of Time-of-Flight Drift Using an Internal Reference Ion during Long-Duration LIMAS Depth Profiling}},
  author = {Ken-ichi Bajo},
  journal = {Mass Spectrometry},
  year = {2026},
  month = oct,
  doi = {10.5702/massspectrometry.a0201},
  url = {https://doi.org/10.5702/massspectrometry.a0201},
  note = {CC BY-NC 4.0},
}
RIS
TY  - JOUR
TI  - Compensation of Time-of-Flight Drift Using an Internal Reference Ion during Long-Duration LIMAS Depth Profiling
AU  - Ken-ichi Bajo
PY  - 2026
DA  - 2026/10/30
JO  - Mass Spectrometry
DO  - 10.5702/massspectrometry.a0201
UR  - https://doi.org/10.5702/massspectrometry.a0201
AB  - In this study, the temporal behavior and compensation of time-of-flight (TOF) drift during long-duration depth profiling using Laser Ionization Mass Nanoscope (LIMAS) were investigated. Noble gas depth profiling by LIMAS requires detection of weak isotope signals with nanometer-scale depth resolution. However, such measurements often take several hours to a couple of days, during which TOF peak-position drift and statistical uncertainty may affect analytical precision. In this study, a 4He-implanted olivine substrate was analyzed as a test sample. The TOF spectra were processed by depth-layer stacking, peak extraction, and peak-position analysis. The 24Mg2+ peak was used as a matrix-derived internal reference to monitor and compensate for TOF drift. The peak-position shifts of measured ions showed clear linear correlations with the 24Mg2+ peak shift, indicating that the observed TOF drift is largely systematic. A 24Mg2+-based compensation reduced the peak-position fluctuations of the other ion species to less than 1 ns and improved the mass resolving power. These results demonstrate that the 24Mg2+ peak can be used as an internal reference for TOF drift compensation during long-duration measurements, providing a robust approach for micro- to nanoscale noble gas depth profiling using LIMAS.
SN  - 2187-137X
SN  - 2186-5116
ER  -