Optical coherence tomography imaging device comparison study in papilloedema secondary to idiopathic intracranial hypertension
9 affiliations
- 1Birmingham Neuro-Ophthalmology, University Hospitals Birmingham NHS Foundation Trust, Queen Elizabeth Hospital, Birmingham, UK
- 2Metabolism and Systems Science, School of Medical Sciences, University of Birmingham, Birmingham, UK
- 3Department of Neurology, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK
- 4Department of Neurology, Sandwell and West Birmingham NHS Trust, Birmingham, UK
- 5Department of Neurosurgery, University Hospitals Birmingham NHS Foundation Trust, Birmingham, UK
- 6NIHR Birmingham Biomedical Research Centre, University Hospitals Birmingham NHS Foundation Trust and University of Birmingham, Birmingham, UK
- 7Neuroscience and Ophthalmology, School of Infection, Inflammation and Immunology, University of Birmingham, Birmingham, UK
- 8Academic Department of Military Surgery and Trauma, Royal Centre for Defence Medicine, Birmingham, UK
- 9Ophthalmology Department, Queen’s University, Kingston, ON Canada
Abstract
Background: Optic nerve imaging assessing papilloedema is an increasingly important clinical measure to guide treatment. The aim of this study was to understand the agreement between two commonly used OCT devices and whether pooling of papilloedema OCT imaging data could be easily facilitated.
Methods: People living with idiopathic intracranial hypertension (IIH) underwent non-dilated same-day OCT imaging performed on Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany) and Cirrus 6000 OCT (Carl Zeiss Meditec, Dublin, CA). Comparable measurements were collected (retinal nerve fibre layer (RNFL), maximum height of the optic nerve head and macular ganglion cell layer assessment). Bland-Altman plots were constructed, intraclass correlation coefficient (ICC) and Pearson correlation coefficient were calculated.
Results: Ninety-seven people with a definite diagnosis of IIH were included: 95% were women with a mean age of 31 years. Mean global RNFL thickness was 117.4μm (SD 41.6) on the Spectralis platform and 122.5μm (SD 53.8) on the Cirrus platform. There was excellent agreement between the devices for RNFL thickness measurements (ICC = 0.922), although the Bland-Altman plot revealed increasing discrepancy at higher RNFL thicknesses. Maximum disc height displayed good agreement (ICC = 0.882) between devices. There was a very strong positive correlation between Spectralis GCL volume and Cirrus GC-IPL thickness(r = 0.911).
Conclusions: This OCT imaging dataset spanning a range of papilloedema severity in people with a definite diagnosis of IIH showed good relative agreement of RNFL measurements between Spectralis and Cirrus devices at RNFL thicknesses close to normal, but greater variance with more severe papilloedema. RNFL measurements from the two devices cannot be used interchangeably. Macular GCL assessments showed a strong positive linear correlation.
Related
- Longitudinal Optical Coherence Tomography Indices in Pediatric Idiopathic Intracranial Hypertension
- Regional Perivascular Space Burden in Pediatric Idiopathic Intracranial Hypertension: A Case–Control MRI Study
- An Endovascular Approach for Idiopathic Intracranial Hypertension: Assessing Morphometric Candidacy after Venous Sinus Stenting
- Recurrent Idiopathic Intracranial Hypertension With Bilateral Abducens Nerve Palsy in a Prepubertal Child: A Case Report and Literature Review.
- Combined Idiopathic Intracranial Hypertension score, an integrated severity and treatment-oriented tool for idiopathic intracranial hypertension: a retrospective cohort study
BibTeX
@article{berman2026optical,
title = {{Optical coherence tomography imaging device comparison study in papilloedema secondary to idiopathic intracranial hypertension}},
author = {Gabriele Berman and Andreas Yiangou and Mark Thaller and Nick Capewell and Donna Holding and Georgios Tsermoulas and Alexandra J. Sinclair and Richard J. Blanch and Susan P. Mollan},
journal = {Eye},
year = {2026},
month = sep,
doi = {10.1038/s41433-026-04823-4},
url = {https://doi.org/10.1038/s41433-026-04823-4},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Optical coherence tomography imaging device comparison study in papilloedema secondary to idiopathic intracranial hypertension
AU - Gabriele Berman
AU - Andreas Yiangou
AU - Mark Thaller
AU - Nick Capewell
AU - Donna Holding
AU - Georgios Tsermoulas
AU - Alexandra J. Sinclair
AU - Richard J. Blanch
AU - Susan P. Mollan
PY - 2026
DA - 2026/09/22
JO - Eye
DO - 10.1038/s41433-026-04823-4
UR - https://doi.org/10.1038/s41433-026-04823-4
AB - Background: Optic nerve imaging assessing papilloedema is an increasingly important clinical measure to guide treatment. The aim of this study was to understand the agreement between two commonly used OCT devices and whether pooling of papilloedema OCT imaging data could be easily facilitated. Methods: People living with idiopathic intracranial hypertension (IIH) underwent non-dilated same-day OCT imaging performed on Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany) and Cirrus 6000 OCT (Carl Zeiss Meditec, Dublin, CA). Comparable measurements were collected (retinal nerve fibre layer (RNFL), maximum height of the optic nerve head and macular ganglion cell layer assessment). Bland-Altman plots were constructed, intraclass correlation coefficient (ICC) and Pearson correlation coefficient were calculated. Results: Ninety-seven people with a definite diagnosis of IIH were included: 95% were women with a mean age of 31 years. Mean global RNFL thickness was 117.4μm (SD 41.6) on the Spectralis platform and 122.5μm (SD 53.8) on the Cirrus platform. There was excellent agreement between the devices for RNFL thickness measurements (ICC = 0.922), although the Bland-Altman plot revealed increasing discrepancy at higher RNFL thicknesses. Maximum disc height displayed good agreement (ICC = 0.882) between devices. There was a very strong positive correlation between Spectralis GCL volume and Cirrus GC-IPL thickness(r = 0.911). Conclusions: This OCT imaging dataset spanning a range of papilloedema severity in people with a definite diagnosis of IIH showed good relative agreement of RNFL measurements between Spectralis and Cirrus devices at RNFL thicknesses close to normal, but greater variance with more severe papilloedema. RNFL measurements from the two devices cannot be used interchangeably. Macular GCL assessments showed a strong positive linear correlation.