Real-World External Evaluation of an OCT Image Analysis System (AI-CO) in Relation to Humphrey Visual Field Mean Deviation

  1. 1Kora Eye Clinic

Abstract

PurposeTo externally evaluate the unitless optical coherence tomography-derived AI-CO index for association with Humphrey mean deviation (MD) and discrimination of MD-defined visual field severity.

MethodsThis retrospective study included 819 eyes of 424 patients receiving glaucoma care, with AI-CO analysis and Humphrey 24-2 SITA Faster or 30-2 SITA Fast testing within 365 days. The unitless output labeled "peripheral" was termed AI-CO all. Primary analyses assessed Spearman rank correlation and discrimination at MD thresholds of -2, -6, and -12 dB. Additional analyses assessed adjusted associations and one-eye-per-patient and program-specific robustness. Exploratory analyses assessed nonlinearity and cutoffs with patient-level cross-validation.

ResultsAI-CO all correlated with Humphrey MD (Spearman {rho}=0.785; patient-cluster 95% confidence interval [CI], 0.750-0.818). Median AI-CO all values decreased from -2.10 in the normal range to -4.40, -10.15, and -17.35 in mild, moderate, and severe categories, respectively (ordered trend P<0.001). Areas under the receiver operating characteristic curve were 0.870 (95% CI, 0.842-0.895), 0.919 (0.894-0.939), and 0.943 (0.920-0.963) at the three thresholds. The adjusted spline detected modest nonlinearity with little gain in explained variance over a linear model (R2, 0.735 to 0.738). Associations remained after covariate adjustment, and findings were similar in the one-eye-per-patient analysis and across visual field programs.

ConclusionsAI-CO all showed a rank correlation with Humphrey MD and differentiated eyes across specified MD thresholds. It should be interpreted as a severity-related index that complements, rather than replaces, measured perimetry. Exploratory AI-CO all cutoffs require independent validation.

Translational RelevanceIn routine glaucoma care, AI-CO all showed a rank correlation with Humphrey MD and discriminated MD-defined severity.

Related

  1. Multi‐system evaluation of baseline‐referenced gray scale CNR profile deviation for phantom‐based ultrasound quality assurance
    Ji‐Hye Kim, Yeong‐Cheol Heo
    Journal of Applied Clinical Medical PhysicsPeer-reviewedCC BY 4.0Full textMedical imaging & radiologyPMC
  2. Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer.
    Xiaoyuan Shen, Zidong Chen, Kaixin Tan +6
    Clinical ophthalmology (Auckland, N.Z.)Peer-reviewedCC BY-NCFull textGeneral medicineEurope PMCPMC
  3. Magnetic Resonance Imaging Evaluation of Azygos Vein Deviation and Risk of Pedicle Screw Penetration in Thoracic Adolescent Idiopathic Scoliosis
    Goutham B Chand, Binoy Sidhardhan, Ashok Ramakrishnan +3
    CureusPeer-reviewedCC BYFull textSurgery & anaesthesiaEurope PMCEurope PMCPMC
  4. Perioperative Intravenous Antibiotic Regimens and Surgical Site Infection After Colorectal Surgery: Clinical Associations and External Evaluation of Risk Prediction Models
    Junqi Cui, Enoch Chi Ngai Lim, Zizhu Zhao +2
    Infectious disease reportsPeer-reviewedCC BYFull textGeneral medicineEurope PMC
  5. Development and quality evaluation of lyophilized plasma external quality assessment samples for coagulation testing in resource-limited settings
    Ngoc Nguyen Vo, Huu Tam Tran, Thi Hong Phuong Nguyen +5
    Practical laboratory medicinePeer-reviewedCC BY-NC-NDFull textGeneral medicineEurope PMC
BibTeX
@misc{kora2026realworld,
  title = {{Real-World External Evaluation of an OCT Image Analysis System (AI-CO) in Relation to Humphrey Visual Field Mean Deviation}},
  author = {Y. KORA and H. Kora and T. Kora},
  howpublished = {medRxiv},
  year = {2026},
  month = sep,
  doi = {10.64898/2026.09.08.26362248},
  url = {https://www.medrxiv.org/content/10.64898/2026.09.08.26362248v1},
  note = {CC BY-NC-ND},
}
RIS
TY  - UNPB
TI  - Real-World External Evaluation of an OCT Image Analysis System (AI-CO) in Relation to Humphrey Visual Field Mean Deviation
AU  - Y. KORA
AU  - H. Kora
AU  - T. Kora
PY  - 2026
DA  - 2026/09/08
PB  - medRxiv
DO  - 10.64898/2026.09.08.26362248
UR  - https://www.medrxiv.org/content/10.64898/2026.09.08.26362248v1
AB  - PurposeTo externally evaluate the unitless optical coherence tomography-derived AI-CO index for association with Humphrey mean deviation (MD) and discrimination of MD-defined visual field severity.  MethodsThis retrospective study included 819 eyes of 424 patients receiving glaucoma care, with AI-CO analysis and Humphrey 24-2 SITA Faster or 30-2 SITA Fast testing within 365 days. The unitless output labeled "peripheral" was termed AI-CO all. Primary analyses assessed Spearman rank correlation and discrimination at MD thresholds of -2, -6, and -12 dB. Additional analyses assessed adjusted associations and one-eye-per-patient and program-specific robustness. Exploratory analyses assessed nonlinearity and cutoffs with patient-level cross-validation.  ResultsAI-CO all correlated with Humphrey MD (Spearman {rho}=0.785; patient-cluster 95% confidence interval [CI], 0.750-0.818). Median AI-CO all values decreased from -2.10 in the normal range to -4.40, -10.15, and -17.35 in mild, moderate, and severe categories, respectively (ordered trend P<0.001). Areas under the receiver operating characteristic curve were 0.870 (95% CI, 0.842-0.895), 0.919 (0.894-0.939), and 0.943 (0.920-0.963) at the three thresholds. The adjusted spline detected modest nonlinearity with little gain in explained variance over a linear model (R2, 0.735 to 0.738). Associations remained after covariate adjustment, and findings were similar in the one-eye-per-patient analysis and across visual field programs.  ConclusionsAI-CO all showed a rank correlation with Humphrey MD and differentiated eyes across specified MD thresholds. It should be interpreted as a severity-related index that complements, rather than replaces, measured perimetry. Exploratory AI-CO all cutoffs require independent validation.  Translational RelevanceIn routine glaucoma care, AI-CO all showed a rank correlation with Humphrey MD and discriminated MD-defined severity.
ER  -