Stress-informed parametric shape adaptive patient-specific implant design in computer-assisted jaw reconstruction
- 1School of Advanced Engineering, University of Petroleum and Energy Studies (UPES)
- 2Division of Oral and Maxillofacial Surgery, Faculty of Dentistry, The University of Hong Kong
Abstract
Introduction: Reconstructing the jaw after tumor removal or trauma increasingly relies on custom 3D-printed titanium plates. This customization applies to the implant's shape. It usually does not apply to the internal structure. The width and thickness of the plate are still determined by the designer, based on experience and clinical observation. The critical design parameters of the PSI govern its mechanical performance. Hence, if the design could analyze these parameters and adjust them accordingly, it could help avoid plate fracture. This study introduces a parametric-design approach that closes this gap.
Methods: The proposed algorithm addresses these challenges by using stress data derived from finite element analysis (FEA) to selectively reinforce only the high-stress regions of a reconstruction plate, rather than uniformly reinforcing the entire plate. The reinforcement is smoothly integrated into the plate’s geometry to prevent abrupt changes in cross-section that could create stress concentrations. The approach was applied to a baseline plate at two regions of elevated stress.
Results and discussion: Plate A reinforced only the primary zone; this reduced peak stress by 36.0% and strain by 43.8% in that zone, the largest reduction observed across both design variants, and also lowered stress and strain in the second, unmodified zone by 12.7% and 14.3%, respectively. Plate B reinforced both zones; this further reduced stress and strain at the second zone by approximately 30.6% and 33.3%, respectively, and reduced strain at the primary zone by approximately 40.6% relative to baseline. Relative to Plate A, Plate B showed a slightly higher stress at the first zone; this stress remained well below the original baseline value. This work shows that patient-specific implant geometry can be shaped directly by biomechanical feedback. This algorithm offers an efficient computational workflow, enabling implants that are not only anatomically tailored but also mechanically optimized as per given loading conditions.
Related
- Time-Varying Coactivation Pattern and Knee Joint Kinematics and Kinetics Following ACL Reconstruction: A Statistical Parametric Mapping Approach
- Limits of Trial-Adaptive Neural Language Fusion Across Large Language Models in P300 Brain Computer Interfaces
- Teaching methodology for the course “Computer Methods in Physics” for prospective physics teachers based on adaptive technologies
- High-Resolution Parallel-Translation CT Reconstruction Based on a Patch-Based Adaptive-Regularization-Guided Diffusion Model
- Statistical Parametric Mapping reveals sex-specific, phase-dependent differences in bilateral three-dimensional ground reaction force during Wushu Changbing PiJi in high-level athletes
BibTeX
@article{nayak2026stressinformed,
title = {{Stress-informed parametric shape adaptive patient-specific implant design in computer-assisted jaw reconstruction}},
author = {Ankit Nayak and Yu-Xiong Su and Jane Jingya Pu},
journal = {Frontiers in Oncology},
year = {2026},
month = oct,
doi = {10.3389/fonc.2026.1966850},
url = {https://www.frontiersin.org/articles/10.3389/fonc.2026.1966850/full},
note = {CC BY 4.0},
}
RIS
TY - JOUR
TI - Stress-informed parametric shape adaptive patient-specific implant design in computer-assisted jaw reconstruction
AU - Ankit Nayak
AU - Yu-Xiong Su
AU - Jane Jingya Pu
PY - 2026
DA - 2026/10/16
JO - Frontiers in Oncology
DO - 10.3389/fonc.2026.1966850
UR - https://www.frontiersin.org/articles/10.3389/fonc.2026.1966850/full
AB - Introduction: Reconstructing the jaw after tumor removal or trauma increasingly relies on custom 3D-printed titanium plates. This customization applies to the implant's shape. It usually does not apply to the internal structure. The width and thickness of the plate are still determined by the designer, based on experience and clinical observation. The critical design parameters of the PSI govern its mechanical performance. Hence, if the design could analyze these parameters and adjust them accordingly, it could help avoid plate fracture. This study introduces a parametric-design approach that closes this gap. Methods: The proposed algorithm addresses these challenges by using stress data derived from finite element analysis (FEA) to selectively reinforce only the high-stress regions of a reconstruction plate, rather than uniformly reinforcing the entire plate. The reinforcement is smoothly integrated into the plate’s geometry to prevent abrupt changes in cross-section that could create stress concentrations. The approach was applied to a baseline plate at two regions of elevated stress. Results and discussion: Plate A reinforced only the primary zone; this reduced peak stress by 36.0% and strain by 43.8% in that zone, the largest reduction observed across both design variants, and also lowered stress and strain in the second, unmodified zone by 12.7% and 14.3%, respectively. Plate B reinforced both zones; this further reduced stress and strain at the second zone by approximately 30.6% and 33.3%, respectively, and reduced strain at the primary zone by approximately 40.6% relative to baseline. Relative to Plate A, Plate B showed a slightly higher stress at the first zone; this stress remained well below the original baseline value. This work shows that patient-specific implant geometry can be shaped directly by biomechanical feedback. This algorithm offers an efficient computational workflow, enabling implants that are not only anatomically tailored but also mechanically optimized as per given loading conditions.
SN - 2234-943X
ER -