Short Expandable-Wing Suture Anchors for Tibial Eminence Avulsion Fracture Fixation: Effects of Solid Versus Additively Manufactured Lattice Wings on Biomechanical Performance and Cytocompatibility.
- 1Department of Biomedical Engineering, National Yang Ming Chiao Tung University, Taipei 112, Taiwan.
- 2Department of Orthopedics, Tri-Service General Hospital Songshan Branch, National Defense Medical University, Taipei 105, Taiwan.
- 3Medical Device Innovation & Translation Center, National Yang Ming Chiao Tung University, Taipei 112, Taiwan.
Abstract
Fixation of tibial eminence avulsion fractures (TEAFs) in skeletally immature patients requires adequate mechanical stability within a limited insertion depth to minimize the risk of physeal injury. Conventional transosseous cortical-button fixation (TCBF) requires the creation of tibial bone tunnels, whereas conventional suture anchors may have excessive insertion depths. In our previous study, a short expandable-wing (SEW) suture anchor (9.5 mm in length and 5.5 mm in diameter) with CNC-machined solid wings was developed to enhance fixation strength within a limited bone volume. In the present study, a second-generation SEW anchor was developed by replacing the solid wings with 3D-printed Ti6Al4V lattice wings to provide a potentially more favorable bone-implant interface. The mechanical performance of CNC-SEW and metal additively manufactured SEW (MAM-SEW) anchors was evaluated using polyurethane-block pull-out testing and a porcine TEAF model with single- and double-suture TCBF controls. Cytocompatibility was assessed using an MG-63 direct-contact MTT assay. Pull-out loads were 574.8 ± 10.6 N for CNC-SEW and 501.2 ± 39.2 N for MAM-SEW, compared with 429.0 ± 29.0 N for a historical TWINFIX comparator. After 200 loading cycles in the porcine model, cumulative displacements were 0.9 and 0.7 mm for CNC-SEW and MAM-SEW, respectively, compared with 2.2 and 1.3 mm for single- and double-suture TCBF. The corresponding ultimate failure loads were 278.88 and 292.56 N for the SEW groups, compared with 109.86 and 223.56 N for the TCBF groups. The MAM-SEW group also exhibited higher MG-63 metabolic activity than the CNC-SEW group. These findings demonstrate favorable initial fixation performance and preliminary cytocompatibility of the lattice-wing SEW anchor, supporting its potential application in TEAF fixation. Further studies are required to evaluate long-term osseointegration and physeal safety.
Related
- Effect of Surface Roughness and Polyelectrolyte Multilayer Coatings on the In Vitro Cytocompatibility of Additively Manufactured Titanium Implants
- Additively manufactured stimulus-responsive shape memory polymers for multi-functional biomedical applications
- Surface Roughness as a Determinant of the Immunological Tissue Response to Additively Manufactured Titanium Implants
- Over-the-Counter Whitening Effects on Optical Properties of Additively-Manufactured Permanent Resins
- Effect of localized thermal history on martensitic microstructure and microhardness variation in wire-arc additively manufactured 17–4 PH stainless steel beads
BibTeX
@article{tsai2026short,
title = {{Short Expandable-Wing Suture Anchors for Tibial Eminence Avulsion Fracture Fixation: Effects of Solid Versus Additively Manufactured Lattice Wings on Biomechanical Performance and Cytocompatibility.}},
author = {Chia-Hung Tsai and Hsu-Wei Lin and Hsuan-Wen Wang and Yen Cheng and Chiao-Min Chang and Rong-Chen Lin and Chun-Li Lin},
journal = {Journal of functional biomaterials},
year = {2026},
month = nov,
doi = {10.3390/jfb17110584},
url = {https://europepmc.org/article/MED/43083029},
note = {CC BY},
}
RIS
TY - JOUR
TI - Short Expandable-Wing Suture Anchors for Tibial Eminence Avulsion Fracture Fixation: Effects of Solid Versus Additively Manufactured Lattice Wings on Biomechanical Performance and Cytocompatibility.
AU - Chia-Hung Tsai
AU - Hsu-Wei Lin
AU - Hsuan-Wen Wang
AU - Yen Cheng
AU - Chiao-Min Chang
AU - Rong-Chen Lin
AU - Chun-Li Lin
PY - 2026
DA - 2026/11/18
JO - Journal of functional biomaterials
DO - 10.3390/jfb17110584
UR - https://europepmc.org/article/MED/43083029
AB - Fixation of tibial eminence avulsion fractures (TEAFs) in skeletally immature patients requires adequate mechanical stability within a limited insertion depth to minimize the risk of physeal injury. Conventional transosseous cortical-button fixation (TCBF) requires the creation of tibial bone tunnels, whereas conventional suture anchors may have excessive insertion depths. In our previous study, a short expandable-wing (SEW) suture anchor (9.5 mm in length and 5.5 mm in diameter) with CNC-machined solid wings was developed to enhance fixation strength within a limited bone volume. In the present study, a second-generation SEW anchor was developed by replacing the solid wings with 3D-printed Ti6Al4V lattice wings to provide a potentially more favorable bone-implant interface. The mechanical performance of CNC-SEW and metal additively manufactured SEW (MAM-SEW) anchors was evaluated using polyurethane-block pull-out testing and a porcine TEAF model with single- and double-suture TCBF controls. Cytocompatibility was assessed using an MG-63 direct-contact MTT assay. Pull-out loads were 574.8 ± 10.6 N for CNC-SEW and 501.2 ± 39.2 N for MAM-SEW, compared with 429.0 ± 29.0 N for a historical TWINFIX comparator. After 200 loading cycles in the porcine model, cumulative displacements were 0.9 and 0.7 mm for CNC-SEW and MAM-SEW, respectively, compared with 2.2 and 1.3 mm for single- and double-suture TCBF. The corresponding ultimate failure loads were 278.88 and 292.56 N for the SEW groups, compared with 109.86 and 223.56 N for the TCBF groups. The MAM-SEW group also exhibited higher MG-63 metabolic activity than the CNC-SEW group. These findings demonstrate favorable initial fixation performance and preliminary cytocompatibility of the lattice-wing SEW anchor, supporting its potential application in TEAF fixation. Further studies are required to evaluate long-term osseointegration and physeal safety.